CONTENTS Important Safeguards ............................... 1 The Purpose of this Manual ........................... 2 Product Description – Intended Use......................2 General Safety ..............................2 Contraindications for Use ..........................2 Intended User Profile ............................2 Installation Guide .............................. 3 Operation Instruction ............................5 Maintenance &...
IMPORTANT SAFEGUARDS When using electrical products, especially when children are present, basic safety precautions should always be followed, including the following: This device can be used in home healthcare and professional healthcare environment. DANGER - READ ALL INSTRUCTIONS BEFORE USING THE APPLIANCE DANGER –...
3. GENERAL SAFETY It is important to read the information in this user manual before you use your HealthCo Royal system. Please follow the guidelines below for your added safety and maintaining system performance.
1. Unpack the system and place the control unit at the foot end of the bed. 2. Remove the existing mattress from the bed. 3. Place the HealthCo Royal Alternating Full Replacement System directly onto the bed frame and position the air hose at the foot end of bed (mattress top cover indicates foot end).
Page 7
14. If an emergency mattress deflation is required, turn the CPR valve to ‘’OPEN’’ position (Fig. 1). The mattress should deflate within 30 seconds. To close CPR valve simply turn the valve to the ‘’CLOSE’’ position (Fig. 2). Fig. 1 Turn the dial to OPEN position Fig.
COMFORT LOW PRESSURE AUTO HealthCo MODE LOCK Royal STATIC ALTERNATE Full Automatic System Description Item 1. Power Switch – Power On (Green LED), Standby (Amber LED), 2. Pressure Setting Button - allows for manual adjustment of pressures by pressing “Comfort”...
8. MAINTENANCE & TROUBLESHOOTING No daily maintenance is required. Or to maintain the condition of the alternating mattress system, service the system regularly according to the schedule recommended by your distributor. This equipment should only be serviced by a qualified and authorised technicians. For common trouble shooting tips please refer to the table below.
Page 10
10. Alternatively remove the cover and launder, 70º C (160º F), using normal detergents. It is essential that articles be thoroughly dried after all cleaning procedures and before storage 11. Perform the following steps to clean the power unit and hose fittings: - Wipe all controls, chassis and hose fittings with a damp cloth and a mild detergent.
10. SPECIFICATION Master Control Unit Model No. P-16AD Pump (M16-12) Size (mm) 250(L)x110(W)x210(H) Weight (Kg) Cycle Time (min) 12 min Min/Max Pressure 20 ~ 60 mmHg +/- 6 mmHg Max Flow-rate >4.5 L/min Rated Voltage AC 100-240V, 50/60Hz Max Current 0.2-0.1A Fuse Rating T2AL 250V...
Page 12
The HealthCo Royal is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. The customer or the user of the HealthCo Royal should assure that it is used in such an environment. Electromagnetic Environment-Guidance...
Page 13
Manufacturer’s declaration-electromagnetic immunity The HealthCo Royal is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. The customer or the user of the HealthCo Royal should assure that it is used in such an environment. Electromagnetic...
Page 14
The HealthCo Royal is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. The customer or the user of the HealthCo Royal should assure that is used in such and environment. Electromagnetic Environment- Immunity...
Page 15
The HealthCo Royal is intended for use in the electromagnetic environment (for home healthcare and professional healthcare) specified below. The customer or the user of the HealthCo Royal should assure that it is used in such an environment. Compliance Test...
Alternating Current 15. EXPECTED SERVICE LIFE The HealthCo Royal pump has an expected service life of 5 years. To maintain the condition of the pump have it serviced regularly according to the recommended schedule. Do NOT use unapproved accessories or attempt to modify, disassemble or otherwise misuse the system..
• In no event shall HealthCo be liable for any direct, indirect or consequential damage or loss resulting from the use of equipment. 17. DISCLAIMER WARNING: The use of side rails and other restraints can result in injury or death - through potential entrapment and potential patient falls.
Page 18
Internal and external research is and will always be ongoing. These risk factors should not be taken as prescriptive criteria. C. HealthCo support surfaces should be seen as an aid to care and DO NOT replace the need for good nursing care and intervention. All HealthCo products must be used as part of an individualised care plan which include proper nursing practices i.e.