Specifications - Lifestyle LM5A Manual

Stationary oxygen concentrator
Table of Contents

Advertisement

2. FEATURES

2.3 Specifications

1. Power supply: AC120V, 60Hz; Current:3.5A; Power: 450VA
2. Sound level: ≤ 50dB (A) Acoustic power level: 58.5dB (A)
3. Maximum recommended flow: 5L/min
4. Flow Range at Outlet Pressure of 0 kPa: 0.5~5L/min
Flow Range at Outlet Pressure of 7 kPa: 0.5~5L/min
Change in maximum recommended flow when back pressure of 7 kPa is
applied: < 0.5 L/min
5. Oxygen Concentration: When 0.5~5L/min ,93%±3% (after turning on 30
minutes)
6. Output Pressure: 38kPa±5kPa
7. Release Pressure during machine operation: 250kPa±50kPa
8. Weight: 35.5lbs
9. Dimension: 13"×10.3"×21.3" (330×260×540mm)
10. Altitude: The oxygen concentration will not reduce at altitudes up to 6,000ft /
1828 meters. At altitudes between 6,000ft / 1828 meters and 13,100ft / 4,000
meters, the efficiency will decrease to less than 90%.
11. Safety System:
① Current overload or line surge auto-shutdown.
② High temperature compressor auto-shutdown.
③ High pressure alarm and auto-shutdown.
④ Low pressure alarm and auto-shutdown.
⑤ Low oxygen concentration alarm.
12. Minimum Operating Time: 30 minutes
13. Electric Classification: Class Ⅱ equipment, Type BF applied part (Nasal
oxygen cannula)
14. Mode of Operation: Continuous duty
9