INTRODUCTION We thank you for purchasing a LMS Cooled Incubator, which we trust, will give you many years of satisfactory service. Manufactured in the United Kingdom using the finest materials and modern production techniques your Cooled Incubator has already undergone extensive testing in our factory.
Page 4
LMS COOLED INCUBATORS Quick Start Instructions Commissioning • Remove all packing materials (CARE: door keys wrapped in shelf packaging). • Check for signs of external damage (this must be reported to the supplier immediately). • Check supply voltage and cycles to ensure it is in accordance with your supply.
All cooled Incubators within the LMS Series 1 and 3 range incorporate both a door lock and magnetic gasket with all control features being set into a control panel and positioned so as to minimise the risk of accidental alteration.
INITIAL EXAMINATION AND INSTALLATION COMMISSIONING Remove all packing materials. (ii) Check for signs of external damage (this MUST be reported immediately to the Carrier and Supplier). (iii) Check that the electrical supply details on the information Rating Plate, which is fitted to the rear panel, are in accordance with the available electrical supply.
This device is fitted within the air mixing chamber to safeguard the cooled incubator should a fault condition of overheating develop. Operation of this device will be indicated by the Red ‘safety’ neon light on the front panel being illuminated. In the event of this happening the cooled incubator must be switched off until the reason for the fault condition is established and the fault itself is rectified.
OPTIONAL EXTRAS WHERE FITTED ALL NON PROGRAMMABLE MODELS AUTOMATIC TEMPERATURE CYCLING On these models, the digital controllers are mounted either side of a time clock and will be illuminated in turn according to the time clock settings. (NB – On changeover the controller will briefly indicate 18.8.8 for few seconds –...
Operation (see figure below) Load chart paper as previously described. If required, the meter pointer/stylus (8) can be set to mechanical zero, with power off, by carefully turning the zero adjustment screw (11). LMS Chart Recorder...
Page 11
Note that the stylus may not move freely if the recorder is in the imprint section of its operating cycle with the stylus pressed against the chart paper. In this case the recorder chart motor should be operated for a few seconds until the imprint pressure is released, or the imprint drive wheel (12) can be manually rotated.
Section 4 MAINTENANCE & REPAIR EQUIPMENT RATINGS 220/240V AC, 50 Hz. Up to 11 amps max. (see rating plate for details). The cooled incubator is designed to operate in an environment from +10ºC to 30ºC with a relative humidity from 5 - 85% RH non-condensing. Correct functionality of the cooled incubator or its safety features cannot be guaranteed if it is used outside these ratings.
LMS including any carriage charges. If it is necessary to return equipment to LMS, the equipment must be packaged so as to prevent any damage. Guarantee is for 2 years and cannot be passed on to a subsequent equipment owner.
Page 14
FOR SPARES AND SERVICING Please contact: LMS Ltd. The Modern Forge Amherst Hill Riverhead Sevenoaks, Kent TN13 2EL Tel. 01732 451866 Fax. 01732 450127 Email: sales@lms.ltd.uk www.lms.ltd.uk...