COMMON CAUSES OF NO SOUNDS
COMMON CAUSES OF NO SOUNDS
• Batteries need to be replaced. After installing new batteries, make
sure to turn the remote off and back on. This allows the device to
reset.
NOTE: SmartScope™ does not turn off.
• Battery saver mode. The remote will go into battery saver mode if
left on the same setting for 8 minutes.
• Using multiple remotes. One remote will operate all Smart-
Scopes™ within 100 ft. simultaneously as long as the serial numbers
are compatible. Using multiple remotes can cause interference.
• Interference. Wireless communication from another transmitter
using 433 MHz may cause sound selection issues or existing sounds
to stop or restart. Reset the remote by turning it off briefly, then on
again and select a different Heart or Lung sound.
• Using an external speaker. When using an amplifier/speaker, no
sounds will be heard through the SmartScope™.
• Check volume. Make sure the volume on the amplifier/speaker is
turned up.