Calibration Light Source
Installation and Operation Instructions
Description
The XE-1 Xenon Calibration Source is a wavelength calibration source for UV-VIS-Shortwave NIR
spectrophotometric systems. The XE-1 produces Xenon lines from 916-1984 nm, and is an ideal lamp
to use when performing fast, accurate spectrometer wavelength calibrations.
The XE-1 features an SMA 905 Connector for interfacing with our optical fibers. It operates with a 12
VDC power supply (included with the unit) or 9V battery (not included).
The following sections detail the features of the XE-1 Xenon Calibration Light Sources.
The XE-1 is NOT designed to operate as an excitation source in your experiments.
Spectral lines above 922 nm are not easily detectable by Ocean Optics spectrometers.
008-00000-XE1-01-1209
XE-1 Xenon
XE-1 Xenon Calibration Light Source
Note
1

Advertisement

loading

Summary of Contents for Ocean Optics XE-1 Xenon

  • Page 1 Installation and Operation Instructions Description The XE-1 Xenon Calibration Source is a wavelength calibration source for UV-VIS-Shortwave NIR spectrophotometric systems. The XE-1 produces Xenon lines from 916-1984 nm, and is an ideal lamp to use when performing fast, accurate spectrometer wavelength calibrations.
  • Page 2: Parts Included

    Dangerous voltages are present, and there are no user-serviceable parts inside. Never open the XE-1. Additional Accessories The following are additional accessories available from Ocean Optics that you may need, depending on your system set-up:  Spectrometer ...
  • Page 3 You will be calculating the value for I and the three Cs. Calibration Requirements for Ocean Optics Spectrometers To re-calibrate the wavelength of your Ocean Optics spectrometer using the XE-1, you will need the following items:  XE-1 Xenon Calibration Light Source ...
  • Page 4: Calibration Procedure

    If you are using Microsoft Excel, choose Tools | Add-Ins and check AnalysisToolPak and AnalysisTookPak-VBA. Calibration Procedure Please see the appropriate manual for your model Ocean Optics spectrometer for instructions on how to calibrate it. Ocean Optics manuals are located at http://www.oceanoptics.com/technical/operatinginstructions.asp.
  • Page 5 XE-1-CAL Installation and Operation Instructions Strong XE Emission Lines by Wavelength (nm) There are more emission lines shown below than printed on the label on the XE-1 housing. The label is intended as a quick, convenient reference and does not list every XE emission line that exists. Below is a list of wavelengths for Xenon (as well as Krypton and Neon).
  • Page 6 XE-1-CAL Installation and Operation Instructions Specifications Table Output Low-pressure gas discharge lines of Xenon Dimensions (in mm): 125.7 x 70 x 25.8 Power consumption: 250 mA at 12 VDC Power requirements: 12 VDC wall transformer (included) or 9 VDC battery (optional) Bulb life: Approx.