Page 1
12/03 5100-8-700 Rev-D Footswitch 5100-8 Stryker Instruments European Authorized Rep: 4100 E. Milham RA/QA Manager Kalamazoo, Michigan Stryker France (USA) 49001 ZAC Satolas Green Pusignan 269-323-7700 Av. de Satolas Green 1-800-253-3210 69881 MEYZIEU Cedex France European Equivalent 5100-8-713 Rev-B Japanese Equivalent 5100-8-721 Rev-B...
Page 2
User/Patient Safety information provided in the TPS Console instruction booklet. Familiarization with the Total Performance System prior to use is important. If you have any questions, contact your Stryker Instruments representative or Stryker Customer Service at 1-800-253- 3210. WARNING: Prior to use, inspect for damage.
Page 3
Footswitch Functions PEDALS (factory default settings) Reverse Depress pedal for variable speed (up to maximum selected on console). Hand- piece operates in the counterclockwise direction. Forward Depress pedal for variable speed (up to maximum selected on console). Hand- piece operates in the clockwise direction. PADS (factory default settings) The activated handpiece (TPS or Endo) determines the operation of the pads.
Page 4
Solution Handpiece is not Make sure cable plugged into console connector is fully seated. Footswitch is damaged. Return to Stryker for repair.* *See Instrument Repair/Loaner Program. Outside the USA, contact your nearest Subsidiary. Specifications Model No.: 5100-8 Footswitch Electrical: Input = 5VDC, 18mA Output = 2.5VDC to 4.9VDC...