Power frequency (50/60
Hz) magnetic field
environment IEC
61000-4-8
Radiated RF
electromagnetic fields IEC
61000-4-3
Conducted disturbances
induced by RF
electromagnetic fields IEC
61000-4-6
NOTE 1: Ut stands for the AC main voltage prior to application of the test level.
NOTE 2: The BiliSoft 2.0 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the BiliSoft 2.0.
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
<5 % Ut (>95 % dip in Ut)
for 1.0 cycle AND
70% Ut (30% dip in Ut) for
0.5 sec
At a phase angle of 0°
<5% Ut (>95% dip in Ut) for
5 sec.
30 A/m
10 V/m over
80 MHz – 2.7 GHz
Modulation: 80% AM at 1
kHz
3 V rms over
150 kHz – 80 GHz
6 V rms in ISM and
amateur radio bands
between 150 kHz – 80 GHz
Modulation: 80% AM at 1
kHz
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
<5 % Ut (>95 % dip in Ut)
for 1.0 cycle AND
70% Ut (30% dip in Ut) for
0.5 sec
At a phase angle of 0°
<5% Ut (>95% dip in Ut) for
5 sec.
30 A/m
10 V/m over
80 MHz – 2.7 GHz
Modulation: 80% AM at 1
kHz
3 V rms over
150 kHz – 80 GHz
6 V rms in ISM and
amateur radio bands
between 150 kHz – 80 GHz
Modulation: 80% AM at 1
kHz
If the BiliSoft 2.0
requires continued
operation during a
power outage, it is
recommended to use
an uninterruptible
power supply.
Magnetic fields should
be at levels common
to a hospital or home
setting.
Radio devices such as
cell phones and
walkie-talkies should
not be used within 30
cm (12 inches) of the
BiliSoft 2.0 Light Box or
power cord
.
2
9-3