Page 1
WAGNER & GUDER MEDICAL GMBH Direction for use GA05-2021_cubeLITE_ENG issue May 2021...
Page 2
Direction for use, cubeLITE To this manual The instructions for use are part of the scope of delivery. Please read manual before use refraction unit Make yourself familiar with the safety regulations Store at the place of use of the appliance ...
Direction for use, cubeLITE Table of contents SAFETY INSTRUCTIONS ....................4 Description of symbols ...................... 4 Application area ........................ 5 Informations for the operator ..................... 6 REQUIREMENTS FOR OPERATION ................8 Before the first operation ....................8 Before the operation ......................8 During the operation ......................
Direction for use, cubeLITE Safety instructions Description of symbols sign “Follow instructions for use“ especially read the safety instructions and follow the enclosed documents Equipment labelling „nameplate“ - MD medical device I - Main fuses - Unit name - Operating cycle patient chair...
Any use other than that specified is not permitted. Intended use The cubeLITE examination unit is a versatile device for positioning a sitting patient and/or customer in front of ophthalmic examination equipment. Both manual and electromotive adjustment options are available for positioning.
Direction for use, cubeLITE Informations for the operator Before using the device, make yourself thoroughly familiar with the contents of the operating instructions. Also observe the operating instructions for accessories and other system components. The examination unit may only be installed and commissioned by qualified personnel who are familiar with the installation, commis- sioning and operation of the product.
Page 7
Direction for use, cubeLITE The lifting column of the patient chair is not designed for permanent operation. After an operating time of more than 1 minute, the lifting column of the patient chair requires a cooling time of 9 minutes.
Direction for use, cubeLITE Requirements for operation Before the first operation Check the stability of the examination unit Check the fixing points of the installed equipment Pay attention to the maximum permissible weights of the de- vices ...
Direction for use, cubeLITE Positioning and installation instructions Correct adjustment of the refraction unit Tough floor Wagner & Guder Medical GmbH issue May 2021...
Direction for use, cubeLITE General construction cubeLITE projector- mounting phoropter- mounting d= 22mm reading spot pillar patient seat control panel base frame (phoropter control unit) electric pillar adjustable base plate Figure: 1: general structure cubeLITE Wagner & Guder Medical GmbH issue May 2021...
Direction for use, cubeLITE Phoropter arm Type I The Phoropterarm type I allows a smooth and safe moving in front of the patient. The arm reaches its final position by pulling the handle into the manual end position. The maximum permissible load of the arm is 15 kg. The correct working high can adjust at the pillar.
Direction for use, cubeLITE Patient chair Patient chair is solid connected to the base plate. The electric movement has a stroke of 200 mm. The basic seat „ONE“ is equipped with fixed arm rest and a standard letherette qualitiy. Different seat shells are available.
Direction for use, cubeLITE Operation Electrical connection The electrical connection of the unit is on a properly installed safety socket or the included wall mounted box. Please use only the cable of mains connection, 2,5m cable length, wire size 1.0 mm2 or the cable of the wall connection box.
Direction for use, cubeLITE Commissioning Connect power cord to refraction place Connect the power cord to a properly installed power outlet. Putt the operating button „I“ green lighting The unit is ready for operation when the green switch is shining.
Direction for use, cubeLITE Cleaning For cleaning the varnished parts and the seat upholstery of cubeLITE please use a wet cloth. Into the water you can add a mild cleaning agent (if necessary „scouring milk“). Obstinate dirt at varnished surfaces can be removed with the help of cleaning solvent or rectified spirits.
Direction for use, cubeLITE Maintenance of the device Safety check The examination unit is maintenance-free and designed for a service life of 6 years. To ensure safe and proper operation and a long service life at all times, regular preventive maintenance must be carried out.
Direction for use, cubeLITE Environmental conditions 1. For operations Feature Permited value range +5°C … +40°C Temperature 30% … 75% no condensation Relative humidity Operating height till 2000m a.n.z 2. For Transport and storage Feature Permited value range -20°C … +70°C Temperature 10% …...
Direction for use, cubeLITE Patients surrounding D=3m Figure: 8: Patients surrounding As patients surrounding the above areas is highlighted. In this area will ensure for the patient, the maximum possible protection. In the area surrounding the patient may experience dangerous areas. Here is a special duty of care of the doctor or optician is precisely in connection with the installed equipment.
Direction for use, cubeLITE Technical data Anschlussspannung 100- 230VAC / nominal input voltage 50-60Hz Anschlussleistung connected load 6A (10A) Beleuchtung electric lighting 5VA (LED) Gerätetyp class of unit Schutzklasse class of protection Gesamthöhe maximum height 1765mm Maximalgewicht Einheit maximum weight of unit...
Direction for use, cubeLITE Warranty and waste disposal If there are defects from installation, assembly or materials within 24 months after purchase, we guarantee the fastest and free repair of the refraction unit, or after our decision, a free exchange. For electronic components such as power supply or motherboard defect within 12 months will be set free repair.
Electromagnetic Emissions (Tab. 201 according to DIN EN 60601-1-2) The refraction place cubeLITE is intended for use in an electromagnetic environment as described below. The customer or user of the refraction place cubeLITE should ensure that the device is used in such an environment.
Page 22
Direction for use, cubeLITE Guidance and manufacturer’s declaration – Electromagnetic immunity This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should assure that it is used in such an environment.
Page 23
Direction for use, cubeLITE Guidance and manufacturer’s declaration – Electromagnetic immunity This product is intended for use in the electromagnetic environment specified below. The customer or the user of this product should assure that it is used in such an environment.
Page 24
Direction for use, cubeLITE Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broad-cast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
Direction for use, cubeLITE Eliminating electromagnetic interference Service Band (MHz) Test frequency Max. power Distance (MHz) P(W) d (m) Different services (TETRA 400) 380-390 FRS-460, GMRS 460 430-470 LTE Band 13/17 704-707 710/745/780 GSM800/900 LTE Band 5 Mobile phone CT1+, CT2, CT3...
Page 26
Direction for use, cubeLITE cubeLITE, electric scheme Wagner & Guder Medical GmbH issue May 2021...
Direction for use, cubeLITE Purchase order details Spare parts Description Order number Power cord 2,5m MC-B1D12500 Fuse T 6,3 A / H - Littlefuse 021506.3TXP Pillar patient chair – TU401166 E100 Illuminants LED – GU5.3 E200 Seat shell ONE S1500...