Advertisement

Quick Links

Table of content
Preamble ..................................................................................... 3
Scope of delivery ......................................................................... 4
Important hints - conventional use ................................................................ 5
Contraindications ......................................................................................... 5
Precautions .................................................................................................. 6
Configuration of the device ............................................................................ 7
Course of exercise EMG+ES ....................................................... 9
Preparation .................................................................................................. 9
Set the current ........................................................................................... 10
EMG-setting............................................................................................... 11
The biofeedback-training ............................................................................. 13
Electrical stimulation-training ..................................................................... 15
EMG-triggered electrical stimulation ........................................................... 18
operation mode pressure ........................................................... 20
preparation ................................................................................................. 20
set the pre pressure ................................................................................... 20
start the diagnosis ..................................................................................... 20
General exercises ...................................................................... 22
Maintenance .............................................................................. 25
Debugging ................................................................................. 26
Therapy settings ........................................................................................ 30
Patient administration ................................................................................ 34
Exercise data ............................................................................................. 36
System settings ......................................................................................... 39
Technical data ............................................................................ 43
Warranty..................................................................................... 48
Manual Conpreva
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Conpreva and is the answer not in the manual?

Questions and answers

Summary of Contents for Haynl-Elektronik Conpreva

  • Page 1: Table Of Contents

    General exercises ..............22 Maintenance ................25 Debugging ................. 26 Further informations for the physican or therapist ......28 Therapy settings ..................30 Patient administration ................34 Exercise data ..................... 36 System settings ..................39 Technical data ................43 Warranty..................48 Manual Conpreva...
  • Page 2: Preamble

    Haynl-Elektronik GmbH. pain reduction (Tens) The company Haynl-Elektronik GmbH reserves itself the right to make at any time and without advance notice, improvements and changes. All data in this operating instruction are regularly examined. Corrections are made in the next expenditure.
  • Page 3: Scope Of Delivery

    The therapy device is used for the treatment of functional muscle dispositions and several kinds of • therapy device Conpreva incontinence. Please read through this manual • 1 charging device (LG PTA med) completly before you start using the device.
  • Page 4: Precautions

    Please inform your top view doctor as well if skin irritations should appear during sensor 1 the training. datalink Explosion hazard if used in the presense of flammable anesthetics. sensor 2 MultiLink side view reset button Manual Conpreva...
  • Page 5: Course Of Exercise Emg+Es

    Visit for the training a place, which is not in direct equalisation with cable proximity of television sets, loudspeakers, fluorescent The DMA must lamps or the like. The Conpreva is equipped with highly only connected if sensitive measuring electronics. Strong outside you wan’t to electromagnetic fields could cause measuring errors.
  • Page 6: Set The Current

    You can always reduce or increase the current during the exercise. channel 2 channel 1 Exercise with (right channel) electrical (left channel) stimulation uses only channel 1 for EMG-scale EMG-scale measurement, so only one emg scale is shown EMG-mode EMG-mode EMG-range EMG-range Manual Conpreva...
  • Page 7: The Biofeedback-Training

    To select an other animation during the exercise press the resume button. There are two special animations for the coordination training: channel 1 activ channel 2 activ Try to balance the scale. Additionally the EMG-values and the difference between the two channels is displayed. Manual Conpreva...
  • Page 8: Electrical Stimulation-Training

    Concentrate during the entire training on the concerned musculature and hold it loose and relaxed. This area graph displays both channels in one diagram. (green - channel 1, red - channel 2). The EMG-range can be differently scaled. Manual Conpreva...
  • Page 9: The Combined Training (Biofeedback+Electrical Stimulation)

    If an animation does not please you, you have the possibility to select between different animations. Press in addition during the tense or relax phase the Resume button. Five further diagrams stand to the selection. Manual Conpreva...
  • Page 10: Emg-Triggered Electrical Stimulation

    If you cross the green line with the amplitude of the EMG scale, the current is switched on. If not the green threshold comes down. Notice the stimulated musculature. Try to tense up these muscles in the next motion conception phase. Manual Conpreva...
  • Page 11: Operation Mode Pressure

    All diagnostic programs only record data, i.e. resume key in while the exercise. the Conpreva does not give any instructions like tense up or an easing. Manual Conpreva...
  • Page 12: General Exercises

    The therapy device has an actual working time of Additionally to indication-referred training, the approx. 4 hours. In the right top of the display you find Conpreva includes exercises for the improvement of the momentary charging state of the integrated the perseverance, strength and co-ordination of the rechargable batteries.
  • Page 13: Maintenance

    Clean the therapy device never under flowing water. Storage in the suit-case. Safety regulations Attention, keep the plugs dry! For the Conpreva are safety-relevant controls in accordance with MedBetreibV, §6, prescribed. (product according with annex 1 of the regulation) Manual Conpreva...
  • Page 14: Debugging

    2 hours on/off button rechargeable batter- charge the batteies doesn’t work ies are fully dis- charged Manual Conpreva...
  • Page 15: Further Informations For The Physican Or Therapist

    Therapy settings Here you can adjust all parameters, which are The Conpreva has extensive options, which permit important for the therapy, e.g. all times or the an optimal and individual therapy organization. frequencies, with which the patient has to train. Stan- Principal item of this functionality forms the main dard disease patterns are pre-defined.
  • Page 16: Therapy Settings

    The The pressure and incontinence programs are available pulse width is adjustable between 40μs and 500μs under the menu-entry torso. Manual Conpreva...
  • Page 17 Example Select electrical stimulation as mode of operation. The assistant queries only the current time and the break time. If you want to set a frequency modulation, set the frequency to zero. On the screen appears frequency modulation. Manual Conpreva...
  • Page 18: Patient Administration

    Enter successively first name, address, date of birth etc. Press the resume button, if the fields should remain empty. For selective data input select the interesting point with the turn switch. Above in the screen you see the category, in the left illustration e.g. surname. Manual Conpreva...
  • Page 19: Exercise Data

    Exercise data training continuity 70%: average time, which was • the patient over 70% of its maximum, without The Conpreva can show the developing of the training fallen below 70% meetings in wave shape. average: average of the tensephase in μV •...
  • Page 20: System Settings

    • button in the exercise in mBar (absolute) Proceed similarly, in order to change the volume of average pressure: absolute in mBar • the Speech and the EMG. pre pressure: in mBar • Manual Conpreva...
  • Page 21 EMG setting, not however with the automatic remains with the patient. The patient cannot change EMG setting. the therapy parameters. The Trigger options are divided into Trigger automatic, with the appendant Update rate and Trigger limit with appedant time limit. Manual Conpreva...
  • Page 22: Technical Data

    4 pole socket for the connection to the PC over a USB cable connection (5000V examined/optically decouples). 3.5 mm of socket for headphones. Multilink for connection with DMA or charging device LG PTA med Power supply NiMH recharable batteries size AA 2500 mAh, 1,2 V charging device LG PTA (230V, 50Hz) Manual Conpreva...
  • Page 23 Prevention of high regulating of the current outside the stimulation phase 40 to • soft increase/ decrease of the current 500µs • Electrode alarm automatic limitation and adjustment of the current during contact reduction • Protection class: II; IP 20 Manual Conpreva...
  • Page 24 Classification after Medicine product law Medicine product class IIa This device applies to the medicine product law MPG and the medicine product operator regulation (MPG BetreibV), attachment 1. Manual Conpreva...
  • Page 25: Warranty

    COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR INCIDENTAL OR CONSEQUENTAL DAMAGES! If problems with the equipment should arise, please contact the manufacturer: Haynl-Elektronik GmbH Magdeburger Str. 117a 39218 Schönebeck Germany Tel.: +49(3928) 69414 Fax: +49(3928) 76222 www.haynl.com...

Table of Contents