TROUBLESHOOTING (cont.)
Base Alignment Instructions
1. Check the space between the platform
and the base. The space should be even
on all four sides of the base. To test if the
space is even gently press down on one
of the toe-end cornersand note the
depressed space. Press down on the
other toe-end corner and compare the
space. Do the same for all four corners to
detech which corner is higher or lower
than others.
2. To align the space between the platform
and base the hex screws on the
underside of the base need to be
adjusted. To access the underside of the
base obtain a sturdy chair and place the
top of the pillar with the height rod side
facing down on the seat of the chair, as
shown.
3. On the underside of the scale base
remove the two black caps from the "heel
end".
4. With the caps removed you can access
the alignment screws that allows the "heel
end" of the base to move up or down. Use
a 7/64" or 1/8" allen wrench to adjust the
screws. As you adjust the screws, watch
the space between the base and platform
move to ensure the space is even. When
the space looks even, test the platform by
pressing each corner one time. The space
on all four corners should be the same as
the "toe end". If all are even replace the
black caps and return the scale to its
upright position.