Stryker Secure Connect Connected Hospital 5212 Operation & Maintenance Manual

Stryker Secure Connect Connected Hospital 5212 Operation & Maintenance Manual

Advertisement

Quick Links

S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™
O O p p e e r r a a t t i i o o n n s s / / M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l
5212
Connected Hospital®
EN
5212-009-201 Rev A.0
2020/10

Advertisement

loading

Summary of Contents for Stryker Secure Connect Connected Hospital 5212

  • Page 1 S S e e c c u u r r e e ® ® C C o o n n n n e e c c t t ™ ™ O O p p e e r r a a t t i i o o n n s s / / M M a a i i n n t t e e n n a a n n c c e e M M a a n n u u a a l l 5212 Connected Hospital®...
  • Page 3 Non-ionizing radiation China RoHS with declarable substances Catalogue number Serial number For US Patents see www.stryker.com/patents Manufacturer Protection from liquid splash I I P P X X 4 4 Class II electrical equipment: equipment in which protection against electric shock does not rely...
  • Page 5 T T a a b b l l e e o o f f C C o o n n t t e e n n t t s s Warning/Caution/Note Definition ..........................2 Summary of safety precautions ..........................2 Introduction ................................3 Product description ..............................3 Expected service life ...............................3 Disposal/recycle..............................3...
  • Page 6 W W a a r r n n i i n n g g / / C C a a u u t t i i o o n n / / N N o o t t e e D D e e f f i i n n i i t t i i o o n n The words W W A A R R N N I I N N G G , C C A A U U T T I I O O N N , and N N O O T T E E carry special meanings and should be carefully reviewed.
  • Page 7 I I n n t t r r o o d d u u c c t t i i o o n n This manual assists you with the operation or maintenance of your Stryker product. Read this manual before operating or maintaining this product.
  • Page 8 S S e e c c u u r r e e C C o o n n n n e e c c t t : 18 VDC, 1.67A Uses infrared (IR) LED and Bluetooth based on Stryker proprietary communication scheme...
  • Page 9 Competent authority of the European Member State where the user and/or patient is established. To view your operations or maintenance manual online, see https://techweb.stryker.com/. Have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical Support. Include the serial number in all written communication.
  • Page 10 C C A A U U T T I I O O N N - - Always match the dip-switches on SB1 and SB2 to the connected product configuration to avoid the risk of head wall damage. N N o o t t e e - - To confirm dip-switch configuration, contact Stryker customer service or technical support ( Contact information (page 5)).
  • Page 11 4 in. - 5 in. (10.2 cm - 12.7 cm) ± 0.5 in. (± 1.3 cm) F F i i g g u u r r e e 3 3 – – M M o o u u n n t t i i n n g g s s p p e e c c i i f f i i c c a a t t i i o o n n s s 6.
  • Page 12 I I n n s s t t a a l l l l a a t t i i o o n n c c h h e e c c k k l l i i s s t t Follow this checklist for the 5212 S S e e c c u u r r e e C C o o n n n n e e c c t t : Confirm that you do not have any unused components after installation.
  • Page 13 O O p p e e r r a a t t i i o o n n S S e e c c u u r r e e C C o o n n n n e e c c t t i i n n d d i i c c a a t t o o r r s s / / f f u u n n c c t t i i o o n n s s The S S e e c c u u r r e e C C o o n n n n e e c c t t has indicator symbols and a nurse call cord out cancel button (A) that is located on top of the product.
  • Page 14 N N o o t t e e - - Stryker service tool required if you use the S S e e c c u u r r e e C C o o n n n n e e c c t t scanner.
  • Page 15 N N o o t t e e - - Stryker service tool required if you use the S S e e c c u u r r e e C C o o n n n n e e c c t t scanner.
  • Page 16 Remove product from service before you perform the preventive maintenance inspection. Check all items listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive maintenance checks more often based on your level of product usage. Service only by qualified personnel.
  • Page 17 For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327- 0770. Stryker’s preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') include the following active ingredients: • n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium chloride - 0.154% •...
  • Page 18 C C l l e e a a n n i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
  • Page 19 D D i i s s i i n n f f e e c c t t i i n n g g C C A A U U T T I I O O N N • Do not clean, disinfect, service, or perform maintenance while the product is in use. •...
  • Page 20 T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g P P r r o o b b l l e e m m S S o o l l u u t t i i o o n n 1.
  • Page 21 S S e e r r v v i i c c e e R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , l l e e f f t t s s i i d d e e ) ) T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 22 F F i i g g u u r r e e 8 8 – – R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( l l e e f f t t ) ) 8.
  • Page 23 R R e e p p l l a a c c i i n n g g t t h h e e b b a a t t t t e e r r y y ( ( 3 3 7 7 - - p p i i n n c c o o n n n n e e c c t t o o r r , , r r i i g g h h t t s s i i d d e e ) ) T T o o o o l l s s r r e e q q u u i i r r e e d d : : •...
  • Page 24 S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r a a s s s s e e m m b b l l y y , , l l e e f f t t 521200380105 Rev AC (Reference only) Torque item AB to 0.34 ±...
  • Page 25 Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-201 Rev A.0...
  • Page 26 Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-201 Rev A.0...
  • Page 27 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 300900380940 Wallside infrared board 521200380111...
  • Page 28 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 521200380803 37-pin cable assembly 521200380806...
  • Page 29 S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r a a s s s s e e m m b b l l y y , , r r i i g g h h t t 521200380110 Rev AC (Reference only) Torque item AB Torque item AD...
  • Page 30 Torque item AA to 0.29 ± 0.05 ft-lb Torque item AB to 0.34 ± 0.05 ft-lb 5212-009-201 Rev A.0...
  • Page 31 Torque item AC to 0.58 ± 0.05 ft-lb 5212-009-201 Rev A.0...
  • Page 32 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 300900380940 Wallside infrared board 521200380111...
  • Page 33 I I t t e e m m N N u u m m b b e e r r N N a a m m e e Q Q u u a a n n t t i i t t y y 521200380801 Battery cable assembly 521200380802...
  • Page 34 W W i i r r e e l l e e s s s s n n o o t t i i f f i i c c a a t t i i o o n n s s For product equipped with wireless communication technology, these statements apply to the countries as indicated: C C o o u u n n t t r r y y N N o o t t i i f f i i c c a a t t i i o o n n...
  • Page 35 E E M M C C i i n n f f o o r r m m a a t t i i o o n n W W A A R R N N I I N N G G •...
  • Page 36 G G u u i i d d a a n n c c e e a a n n d d m m a a n n u u f f a a c c t t u u r r e e r r ’ ’ s s d d e e c c l l a a r r a a t t i i o o n n - - e e l l e e c c t t r r o o m m a a g g n n e e t t i i c c i i m m m m u u n n i i t t y y Main power quality should ±0.5 kV, ±1 kV lines to lines ±0.5 kV, ±1 kV lines to lines...
  • Page 37 Portable and mobile RF communications equipment should follow the guidance in the table titled “Recommended separation distances between portable and mobile RF communication equipment and the 5212 S S e e c c u u r r e e C C o o n n n n e e c c t t locator.” If the mobile service is not listed in the table, the recommended separation...
  • Page 38 R R e e c c o o m m m m e e n n d d e e d d s s e e p p a a r r a a t t i i o o n n d d i i s s t t a a n n c c e e s s b b e e t t w w e e e e n n p p o o r r t t a a b b l l e e a a n n d d m m o o b b i i l l e e R R F F c c o o m m m m u u n n i i c c a a t t i i o o n n e e q q u u i i p p m m e e n n t t a a n n d d t t h h e e 5 5 2 2 1 1 2 2 S S e e c c u u r r e e C C o o n n n n e e c c t t l l o o c c a a t t o o r r The 5212 S S e e c c u u r r e e C C o o n n n n e e c c t t locator is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
  • Page 39 R R e e c c y y c c l l i i n n g g p p a a s s s s p p o o r r t t 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 0 0 5 5 / / 5 5 2 2 1 1 2 2 0 0 0 0 3 3 8 8 0 0 1 1 1 1 0 0 Rev AC R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t...
  • Page 40 7 7 0 0 0 0 0 0 0 0 0 0 8 8 8 8 6 6 7 7 9 9 5 5 Rev AB R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t Q Q u u a a n n t t i i t t y y I I t t e e m m...
  • Page 41 7 7 0 0 0 0 0 0 0 0 0 0 9 9 1 1 5 5 3 3 9 9 0 0 Rev AB R R e e c c y y c c l l a a b b l l e e p p a a r r t t I I m m p p o o r r t t a a n n t t Q Q u u a a n n t t i i t t y y I I t t e e m m...
  • Page 42 S S e e c c u u r r e e C C o o n n n n e e c c t t B B B B I I D D N N o o t t e e - - Give this form to your Stryker representative or your IT system analyst so they can create the associations on the server.
  • Page 44 Stryker Medical 3800 E. Centre Avenue Portage, MI 49002 5212-009-201 Rev A.0 2020/10 WCR: AA.14...