MRI Room Oxygen Monitor - SafeAir®
OPERATORS MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for WARDRAY PREMISE SafeAir Room Oxygen Monitor

  • Page 1 April 2020 GEN.S.20/296 MR105D MRI Room Oxygen Monitor - SafeAir® OPERATORS MANUAL...
  • Page 2 Index Page MR ASTM Safety Definitions Page Introduction Page General Information Page Contents included Page System Description Page Oxygen Monitor Parts Reference Page System Installation Page Diagram for installation Page Power Up Guide Page Power Down Guide Page 12 –13 Maintenance Page Servicing &...
  • Page 3 If you have any problems in the meantime or would like any advice about this or any other MR products from the Wardray Premise range, please contact us at our Head Office: Wardray Premise Limited, Hampton Court Estate, Summer Road...
  • Page 4: General Information

    General Information The SafeAir® Oxygen Monitor has been designed to give a warning of depleted oxygen levels in the scan room. The unit is simple to fit either during the installation of the scanner, or retro fitted into existing MRI rooms afterwards, and does not require the scanner to be ramped down.
  • Page 5: Contents Included

    Contents included SafeAir® Oxygen Monitor unit 230V AC - 18 V DC Transformer - Part No: MR40-LR 2 pin European / continental socket plug 3 pin UK socket plug 30m Flexible PVC hose Vent Relay / Emergency Extract Connector Envitec - Instructions for use and PC software CD Envitec - Short manual leaflet...
  • Page 6: System Description

    System Description The SafeAir® unit comprises an Envitec O2 oxygen monitor system supported by an air sampling pump, HEPA air filter, pressure switch, alarm buzzer and associated airlines. The unit has an internal ‘dryfit’ battery which powers the external alarm and vent relay circuits in the event of mains power loss.
  • Page 7 MR105D SafeAir® Room Oxygen Monitor parts reference Air inlet ‘Envitec’ O2 Oxygen Monitor Alarm buzzer Fuse Reset button ‘Envitec’ Power On/Off button Airflow Fault light Serial No. Emergency Vent Active light Extract Connector 18V DC Power supply in Battery off Vent Reset...
  • Page 8: System Installation

    System Installation Please note: This unit is specifically designed to monitor oxygen levels within the MRI Scanner room safely from within the control room. Unpack the items and check against the packing list. Position the Oxygen monitor in a suitable location in the MRI Control Room to allow the end user to be able to see the LCD screen and, under fault conditions, hear the audible alarm.
  • Page 9 Drawing 1420C...
  • Page 10 Power Up Guide For the first time, or after an extended period of being disconnected from the main supply, the unit should be left powered on whilst the internal ‘dryfit’ battery and Envitec batteries are charged up - ideally overnight. However, we recommend that the unit is left on permanently. The Envitec screen will display an animated battery charging symbol (See Fig 1).
  • Page 11 Power Down Guide The system can be left running 24/7. If the user requires the system to be powered down, either due to an alarm condition or power saving over a long weekend when O2 monitoring is not required, then a shutdown procedure must be followed to avoid flattening the internal batteries.
  • Page 12: Maintenance

    Important It is recommended that the filter and fuel cell sensor are replaced every 6 months. The replacement kit (Ref: MR106) may be obtained directly from Wardray Premise Ltd or in overseas countries from our local agents. When switched on the fuel cell cable plug has a voltage on its terminals (to power the fuel cell), which must not be allowed to touch the chassis if left disconnected.
  • Page 13 Medical Oxygen Sensor Fuel Cell Fuel Cell cable connector Ease out gently here HEPA Vent Filter O2 Sensor ‘T’ Piece Housing Fig 1 Fig 2...
  • Page 14: Warranty

    If the unit is damaged please do not attempt to repair as this could invalidate your warranty and could compromise its safety, please call or email Wardray Premise Limited: T: +44 (0) 20 8398 9911 F: +44 (0) 20 8398 8032 E-MAIL: sales@wardray-premise.com...