Strip
1. Test time
Test time of Blood Glucometer in each case is 10 seconds.
2. Test range
Test range of Blood Glucometer is 1.1mmol/L~33.3mmol/L
3. Measurement repeatability of Blood Glucometer and Blood Glucose Test
Strip
Precision of the repeated test results of Blood Glucometer and Blood
Glucose Test Strip should comply with the requirements in Table 1.
Test Range
<5.5mmol/L(<100mg/dL)
≥5.5 mmol/L(≥100 mg/dL)
4. System accuracy of Blood Glucometer and Blood Glucose Test Strip
For Blood Glucometer and Blood Glucose Test Strip, 95% of the deviation of
the test results should comply with the requirements in Table 2:
Test Range
≤4.2mmol/L(<75mg/dL)
>4.2mmol/L(>75mg/dL)
Blood
Blood
Glucometer
Glucometer
Blood Glucometer
Glucometer and
Table 1 Repeatability
Standard deviation SD<0.42 mmol/L( <7.7 mg/dL)
Coefficient of variation CV<7.5%
Table 2 Accuracy
Allowable deviation
Not more than ±0.83mmol/L(±15mg/dL)
Not more than ±20%
8
and
and
Blood
Blood
Glucose
Glucose
and Blood
Blood Glucose
Glucose
Precision