With mobile mechanical oscillation therapy (15 pages)
Summary of Contents for AffloVest RP8300
Page 2
COPE OF ELIVERY ARTS Rechargeable Battery Hand-held Controller RP8003 RP8004 Adapter Cable for D/C Operation Power Supply Unit/Charger RP8002 AffloVest RP8001 L - RP8300 M - RP8200 S - RP8100 AffloVest Instructions for Use Carrying Case RP8005...
XPLANATION OF YMBOLS Test and Identification Symbols - AffloVest TÜV Symbol - A Tested Medical Product CE Mark - A European Conformity Symbol for Medical Products Body Floating (Not Grounded/Earthed) Direct Current (DC Power) Alternating Current (AC Power) Test and Identification Symbols - RP8001 Power Supply Unit/Charger Not for Outdoor Use TÜV Rheinland Approval - In Accordance with EN60601-1...
Communané Européenne (European Community) - Directives governing electrical and electronic devices Attention! Read Operating Instructions Before Use! Do Not Short-Circuit! Do Not Place Near Or Into Fire! Do Not Submerge In Water! ECHNICAL Nominal Voltage: 12V DC Nominal Power Consumption: Approx.
• Make sure that the AffloVest fits tightly, with even body contact. • There is no need to worry if you are disturbed while using your AffloVest. Your vest will interrupt all set functions after a programmed time and assume a safe standby state.
• When the battery has reached its maximum charge, disconnect it from the charger and connect it to the AffloVest. • It is not recommended to start a new cycle if more than half of the LED lights on the battery are not illuminated, as it cannot be guaranteed that there will be sufficient power to complete the cycle.
▬ The vest operates on a timer of roughly 8 minutes ▬ The vest can be operated from either a battery or wall plug • A microprocessor control, with which input commands are assigned to the respective function via the membrane keypad, is at the core of the AffloVest.
UTTONS AND UNCTIONS NTENSITY EVELS Press the P3 button to start the PERCUSSION program. All motors operate in a pulsating/tapping mode. Percussion of the upper body with a regular rhythm. Press the P2 button to start the VIBRATION program. All motors operate uniformly at the set intensity level. Deep oscillation of the thoracic cavity.
The LED lights P1, P2, or P3 on the hand-held controller light up to indicate the currently active program. The OFF LED light indicates: Constant flashing = AffloVest ready for operation Continuous light = Selected oscillation program active.
EGULATIONS FOR OMMERCIAL LINICAL 12 M ECHNICAL NSPECTION VERY ONTHS Check power consumption at the following setting: VIBRATION program Set intensity level to INTENSE: Target value: 30-35 watts Maximum permissible measured or actual value: 35 watts AFETY NFORMATION The device is certified with the standard EN 60601 and corresponds to the Medical Devices Directive 93/42EEC: HE MANUFACTURER HOWEVER SHALL ACCEPT NO LIABILITY WHATSOEVER FOR ANY DAMAGE AND...
UARANTEE The AffloVest has a limited lifetime warranty. The battery is guaranteed for 12 months. Every battery undergoes several tests and inspections during the manufacturing process. We are committed to ensuring the highest quality standards. The guarantee stipulates that we will repair, free of charge, proven material defects during the guarantee period.
NVIRONMENTAL ROTECTION This symbol, on products and/or accompanying documents, signifies that electrical and electronic products must be disposed of separately from any household waste at the end of their service. To ensure that these products are recycled effectively, deposit them at municipal collection points for recyclable materials where they will be accepted free of charge.
ROUBLESHOOTING A problem may not necessarily be caused by the product. You can save time and money if you rectify some simple problems yourself. The following section is designed to help you troubleshoot your vest. YMPTOM AUSE EMEDY "OFF" LED light does not flash when Plug connection of the connection cable and - Plug in correctly vest is ready for operation.
Voltage fluctuations/flicker emissions as per IEC Compliant purposes. 61000-3-2 The AffloVest is intended for use in the electromagnetic environment specified below. The customer or user of the AffloVest should ensure that it is used in such an environment. IEC 60601-T NTERFERENCE...
Page 16
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AffloVest is used exceeds the applicable compliance level specified above, the AffloVest should be observed to verify normal operation. If abnormal performance features are observed, additional measures may be necessary, such as re-orienting or relocating the AffloVest.
Page 17
Recommended distances between portable and mobile RF telecommunications equipment and the AffloVest The AffloVest is intended for use in an electromagnetic environemnt, in which RF interference is controlled. The customer or user of the AffloVest can help prevent electromagnetic interference by...