%UDHERQ 0HGLFDO &RUSRUDWLRQ 3URGXFW Per una traduzione italiana di questa Pour une traduction française de ce guide guida dell'utente, vada a d'utilisateur, allez à www.braebon.com www.braebon.com Für eine deutsche Übersetzung dieses Para una traducción española de esta Benutzerführers, gehen Sie zu guía del usuario, vaya a...
Page 3
Respiratory pressures are converted into voltage signals compatible with a variety of data acquisition systems. The differential pressure system uses either custom Braebon™ cannulas (Model 0589, 0588, 0582s, or others) or standard oxygen (O ) cannulas with the Braebon™...
Page 4
6DIHW\ ,QIRUPDWLRQ For your personal safety, please read the safety conventions and the warnings and cautions in this manual. 6DIHW\ &RQYHQWLRQV These are the safety conventions for this manual. The table below lists the safety symbol, the name for the symbol, and the meaning of the symbol. The Warning message appears in the manual before Warning procedures or tasks that must be strictly observed to avoid...
Page 5
Braebon™ safety filter (Model 0583) with each patient. The Braebon™ safety filter is required to prevent the spread of contaminants between patients and to prevent moisture damage to the pressure sensor. Failure to use the Braebon™ safety filter will void the warranty.
Page 6
Facilitates connection to a CPAP circuit Input • Oral Input Facilitates recording nasal and oral breathing on separate channels by connecting a Braebon™ Ultima Oral & Nasal Dual Lumen Cannula (Model 0588) to both the oral and nasal inputs • Gain and Baseline Adjustment •...
Page 7
'XDO $LUIORZ 3UHVVXUH 6HQVRU .LW $' &RQWHQWV The 0580A and 0580D Dual Airflow Pressure Sensor kits include the following items: • One Ultima Dual Airflow Pressure Sensor™ (Model 0585) • One adult dual lumen nasal & oral cannula with hydrophobic filter (Model 0588) •...
Page 8
$ERXW WKH 8OWLPD 'XDO $LUIORZ 3UHVVXUH 6HQVRU The Ultima Dual Airflow Pressure Sensor™ is a differential pressure transducer that detects and ± amplifies pressure swings in the cm of H O range. The voltage corresponding to the differential pressure is output for data acquisition. The Ultima Dual Airflow Pressure Sensor™...
Page 9
,QSXWV The Ultima Dual Airflow Pressure Sensor™ has three inputs: nasal input, nasal differential input (CPAP connection), and oral input. Nasal Input With the nasal input, you can either combine oral and nasal breathing into one channel using a single lumen, oral/nasal cannula connected to the nasal input or you can record nasal breathing only using any nasal cannula connected to the nasal input.
Page 10
2XWSXWV On the left side of the Ultima Dual Airflow Pressure Sensor™, there are six outputs — three nasal outputs and three oral outputs. See Table 1 Output Types. You may choose to record from one to six outputs from the Ultima Dual Airflow Pressure Sensor™. Table 1 Output Types Output Name Signal Type...
Page 11
*DLQ DQG %DVHOLQH $GMXVWPHQW 6FUHZV On the right side of the Ultima Dual Airflow Pressure Sensor™, there are four, 20-turn screws that allow you to adjust both the gain and baseline settings for both the nasal and oral outputs. The baseline is set to zero volts in our factory.
Page 12
%DWWHU\ 6WDWXV ,QGLFDWRU /(' Centred on the face of the Ultima Dual Airflow Pressure Sensor™, there is a Battery Status Indicator LED. The LED flashes either green or red to indicate the status of the batteries in the Ultima Dual Airflow Pressure Sensor™. See Table 2 Battery Status Usually, the batteries will last about 30 - 50 nights (8-hour recordings) depending on the battery type used.
Page 13
&RQQHFWLQJ WKH 8OWLPD 'XDO $LUIORZ 3UHVVXUH 6HQVRU Warning: Always use a new single-use Ultima Airflow Pressure Cannula (Model 0589, 0588, 0582s, or others) and a new safety filter (Model 0583) with each patient. The safety filter is required to prevent the spread of contaminants between patients and to prevent moisture damage to the pressure sensor.
Page 14
AC headbox. headbox 3. Use the Braebon™ recommended AC recorder settings AC Headbox Recording System listed in Table 3 AC Figure 2 Complete AC Recorder Connection Recorder Settings.
Page 15
7R FRQQHFW WR D '& DPSOLILHU '& ,QSXW RU 0XOWLSOH[HU FRQQHFWLRQ 1. Using the DC Interface cable Dual lumen, oral and nasal cannula (Model 0594), connected to both the nasal and oral connect a 1 mm input keyhole connector to each of the outputs you would like to record.
Page 16
7R FRQQHFW WR $OLFH 5HVSLURQLFV 1. Using the Alice Note: You Interface cable Dual lumen, oral and nasal can use (Model 0593), cannula connected to both one to six the nasal and oral input connect a 1 mm outputs on the keyhole connector to pressure each of the outputs...
Page 17
0583) attached, connect the safety filter on the cannula to the nasal input only. • Subtract a known 1. Connect the CPAP circuit to the Braebon™ Safety Filter pressure from the nasal (Model 0583). See Figures 5, 6, and 7 for illustrations of baseline different CPAP connection options.
Page 18
CPAP Figure 5 CPAP Adapter Connection and Connection to the Nasal Differential Input Tee Connection Connection CPAP Figure 6 CPAP Mask Connection and CPAP Tee Connection to the Nasal Differential Input Tee Connection CPAP Figure 7 CPAP Mask Connection 8rpvtÃurÃVyvhÃ9hyÃ6v syÃQ r rÃTr ...
Page 19
Warning: Always use a new single-use Ultima Airflow Pressure Cannula (Model 0589, 0588, 0582s, or others) and a new Braebon™ safety filter (Model 0583) to prevent the spread of contaminants between patients and to prevent damage to the pressure sensor.
Page 20
67(3 $GMXVW WKH JDLQ LI QHFHVVDU\ The Ultima Dual Airflow Pressure Sensor™ contains four 20-turn potentiometers (pots) which allow for precise adjustment of both gain and baseline settings for the nasal and oral outputs. Under most circumstances, you will NOT need to make any adjustments. $GMXVWLQJ WKH JDLQ You can enlarge or reduce the size of the waveform on the display by turning the gain screw clockwise or counterclockwise respectively.
Page 21
&DOLEUDWLQJ WKH 8OWLPD 'XDO $LUIORZ 3UHVVXUH 6HQVRU The Ultima Dual Airflow Pressure Sensor™ is a qualitative pressure sensor designed to measure nasal and/or oral pressure associated with human respiratory activity. For qualitative purposes, you do not need to calibrate the pressure sensor. However, you may obtain quantitative data when you calibrate the pressure sensor with a water manometer.
Page 22
Continued use of Braebon™ safety filters (Model 0583) is the best way to maintain your pressure sensor. The safety filter extends the life of the pressure sensor because it prevents particulates and moisture from damaging the pressure sensor.
Page 23
5HSODFLQJ WKH %DWWHU\ The batteries will last about 30 - 50 nights (8-hour recordings) depending on the battery type used. Caution: Use only 1.5V AA batteries or damage to the Ultima Dual Airflow Pressure Sensor™ will result. Do NOT mix battery types. Do NOT insert the batteries backwards or damage may occur to the pressure sensor.
Page 24
7URXEOHVKRRWLQJ If you have difficulty using this product, please verify the following: Problem Solution • Replace the batteries. LED blinks Red The batteries will last between 30 to 50 nights (8-hour recordings) every two seconds depending on the batteries used (i.e., Alkaline vs. Lithium). •...
Page 28
Nor shall BRAEBON MEDICAL CORPORATION be liable, or in any way responsible, for any damages or defects in the Sensor which were caused by abuse, misuse, tampering, neglect, incorrect battery type, or repairs or attempted repairs performed by anyone other than an authorized servicer.