Practical Instrument Electronics
1
EZ-Check™ Switch
Slide the switch to select from three user stored values
for the desired calibration points. The User can select HI,
DIAL and LO positions. These values can easily be
changed to suit the calibration requirements.
.
Instructions for Enabling and Disabling the Configuration Options
1. Turn the Model 520/521 on with the ON/OFF Switch
2. Press the EZ-Dial Knob
CONFIGURATION" message is displayed.
3. Select options by turning the EZ-Dial Knob until the arrow points
to the desired option.
4. The option can be enabled or disabled by tapping the EZ-Dial
Knob.
The Model 520/521 configuration menu will exit automatically after 5
seconds of inactivity and go to normal operation with the options selected.
These options are recalled at turn on until they are changed again.
841 Holt Road #1 • Webster, NY 14580
Practical Instrument Electronics, Inc. Copyright  2003. All rights reserved.
Basic Keypad Operations
Model 520/521 Configuration
3
while the "PRESS EZ-DIAL KNOB FOR
Tel: 585-872-9350 • Fax: 585-872-2638
Model 520/521
TC Simulator(s)
Operating Instructions
2
ON/OFF Switch
Slide the ON/OFF Switch to turn the Model
520/521 on or off.
3
EZ-Dial™ Knob
The EZ-Dial™ Knob has two adjustment
speeds. Simply turning the EZ-Dial™ Knob will
select fine adjustments. While pressing down
and turning the EZ-Dial™ Knob will make
course adjustments.
Note: When the EZ-Check™ Switch is in the
HI or LO position, pressing and holding the EZ-
Dial™ Knob without turning will store new HI
or LO values.
2
.
sales@piecal.com • http://www.piecal.com/
520/521-9002 2/27/03
1-3

Advertisement

loading

Summary of Contents for PIE 520

  • Page 1 Dial™ Knob without turning will store new HI or LO values. Model 520/521 Configuration Instructions for Enabling and Disabling the Configuration Options 1. Turn the Model 520/521 on with the ON/OFF Switch 2. Press the EZ-Dial Knob while the “PRESS EZ-DIAL KNOB FOR CONFIGURATION” message is displayed.
  • Page 2 Hint: For faster calibrations, the product has been designed so the position of the switch can be felt. This tactile feature allows continuous monitoring of the device being calibrated without looking back at the Model 520/521 display. This is also useful in poor lighting or under difficult operating conditions. 841 Holt Road #1 • Webster, NY 14580 Tel: 585-872-9350 •...
  • Page 3 The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our guarantee. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment.

This manual is also suitable for:

521