TABLE OF CONTENTS 1. INTRODUCTION ............................2. INSTALLATION .............................. 2.1 PRELIMINARY TESTS ........................2.2 SAFETY CABLE ............................2.3 SENSOR ORIENTATION ......................... 2.4 SENSOR INSTALLATION ......................2.4.1 SUSPEND THE FIRST SENSOR ....................2.4.2 CONNECT THE SECOND SENSOR TO THE FIRST SENSOR ..........
Data is collected by connecting the IPI string to a readout device (PC, datalogger, SCADA system, etc.) via a customer-specified readout cable. MODEL 6180 VERTICAL IN-PLACE INCLINOMETER SYSTEM | INTRODUCTION | 1...
(terminal) sensor, Model 6180T. This can be used to retrieve the assembly in the event of an accident, and it also can be helpful when lowering the assembly into the casing. Model 6180-6 safety cable assemblies purchased from consist of a GEOKON...
Insert the 6180T sensor into the casing, making sure to orient the wheels FIGURE 4: A & B Directions correctly for proper axis orientation (see Section 2.3), and with the male cable connector facing up toward the top of the casing. MODEL 6180 VERTICAL IN-PLACE INCLINOMETER SYSTEM | INSTALLATION | 3...
FIGURE 5: Holes for Safety Cable Eye Bolts vs Holes for Support Holes FIGURE 6: Support Sensor with Rod 2.4.2 CONNECT THE SECOND SENSOR TO THE FIRST SENSOR Each 6180 segment is supplied with a barbed locking pin pre-installed. Ring Locking Pin...
This will ensure the pins and receptacles on the interior of the connectors align correctly. Push the male and female connectors together until they are completely mated. MODEL 6180 VERTICAL IN-PLACE INCLINOMETER SYSTEM | INSTALLATION | 5...
2.5.1 CONNECT THE CABLE ASSEMBLY TO THE BRACKET Hook the supplied thimble through the bottom-side eyelet on the suspension bracket (6180-2). Feed the bare cable end of the suspension cable assembly (6180-1) through the bottom-side eyelet on the suspension bracket. 6 | INSTALLATION | GEOKON...
2.5.2 CONNECT THE CABLE ASSEMBLY TO THE SENSOR STRING The suspension cable assembly attaches to the sensor string similar to how the sensors attach to each other. For illustrated steps refer to Section 2.4.2. MODEL 6180 VERTICAL IN-PLACE INCLINOMETER SYSTEM | INSTALLATION | 7...
If your datalogger has built-in RS-485 communications, connect the wiring using the diagram in the figure below. FIGURE 18: Wiring of Datalogger without built-in RS-485 Conversion 8 | INSTALLATION | GEOKON...
Pin 4 Pin 3 Pin 2 FIGURE 21: Female Waterproof Connector Wire Color Function Power Black Ground White RS-485+ Data High Green RS-485- Data Low TABLE 1: Four-Pin Wiring Chart MODEL 6180 VERTICAL IN-PLACE INCLINOMETER SYSTEM | INSTALLATION | 9...
MODBUS RTU PROTOCOL 3.1 INTRODUCTION TO MODBUS Model 6180 inclinometers use the industry standard Modbus Remote Terminal Unit (RTU) protocol to communicate with the chosen readout method. As the name suggests, Modbus was designed to work on what is known as a bus network, meaning that every device receives every message that passes across the network.
DATA REDUCTION 4.1 INCLINATION CALCULATION The output of the 6180 Inclinometer Sensor is a corrected angle of inclination. The standard sensor has a full range of ±90° and a calibrated range of ±15°. Register values for the Gauge Factor and Offset are factory-written to the Modbus registers for each sensor using calibration data.
TROUBLESHOOTING Maintenance and troubleshooting of Model 6180 In-Place Inclinometers is confined to periodic checks of the cable connections. The sensors are sealed and there are no user serviceable parts. Should difficulties arise, consult the list of possible solutions shown below. Refer to Appendix C for Modbus error codes.
Note: The sensor stores and transmits errors in binary code to compact the information. Though unlikely, two errors could occur in one measurement cycle. The resulting code will be the sum of the error numbers, e.g., error 4 plus error 8 appears as number 12. 20 | MODBUS ADDRESSABLE SYSTEM | GEOKON...
APPENDIX D. CRBASIC PROGRAMMING D.1 SAMPLE CR1000 PROGRAM The following sample program reads one 6180 sensor string with three biaxial sensors. The string in this example communicates with the CR1000 through the control ports C1 and C2, which are setup as COM1. A RS-485 to TTL converter required.
Page 30
'Use Modbus command to retrieve A Axis and B Axis Degree Readings ModbusMaster (ErrorCode,ComC1,115200,Count,3,A_Axis_Degrees(Count),&H101,1,1,10,0) ModbusMaster (ErrorCode,ComC1,115200,Count,3,B_Axis_Degrees(Count),&H103,1,1,10,0) 'Use Modbus command to retrieve Thermistor Celsius from string ModbusMaster (ErrorCode,ComC1,115200,Count,3,Celsius(Count),&H107,1,1,10,0) 'Delay before proceeding to next reading Delay (1,1,Sec) Next 'Call Table to store Data CallTable Test NextScan EndProg 22 | CRBASIC PROGRAMMING | GEOKON...
Page 32