4. CLEANING AND MAINTENANCE

The device should be cleaned as follows:
• Ensure the wall power outlet is turned off before unplugging the device from the power supply where
appropriate.
• Wipe the exterior and basin of the device (when the Water Chamber has been removed) with a clean,
damp (not wet) cloth and mild dishwashing detergent. Do not use harsh abrasives, solvents or alcohol
as these may damage the device.
Daily
Clean the Water Chamber, Chamber Lid and the Breathing Tube:
• Unlock the Chamber Lid by twisting in the anti-clockwise direction and lift it away from the device.
• Wash the Chamber Lid in a mild dishwashing detergent; rinse thoroughly and dry.
• Disconnect the Breathing Tube from the mask and the device Elbow.
• Clean the Breathing Tube with warm, soapy water; rinse thoroughly and hang with both ends pointing
to the floor to dry.
• Remove the Water Chamber and discard any remaining water. Do not remove the rubber seal on the
chamber.
• The one-piece Water Chamber should not be disassembled.
• The Water Chamber can be washed in a domestic dishwashing machine, or cleaned and washed with
mild dishwashing detergent, then rinsed and dried.
Weekly
Thoroughly clean the Water Chamber, Outlet Seal and the Elbow:
• Soak the inside of the Water Chamber for 10 minutes in a solution of one part white vinegar to two
parts water.
• Empty the vinegar solution and rinse the Water Chamber thoroughly with water.
• Disconnect the Elbow from the device and the Outlet Seal can be easily removed by lightly pushing it
from the back of the device into the Water Chamber housing.
• The Elbow and the Outlet Seal can be cleaned and washed with mild dishwashing detergent, then
rinsed and dried.
• Please place the Outlet Seal before the Elbow in seal cavity when refitting.
Air Filter
The Air Filter is located on the rear of the device. The Air Filter ensures the air is filtered from particles
and dust. To access, first pull out the Filter Grill, then pinch the gauze to remove. The Filter Grill must be
pushed back into the cavity to keep the Air Filter in place.
Replace the Air Filter when it becomes significantly discolored, at least once every three months or after
1,000 hours' worth of running time.
NOTE: When replacing the Air Filter the black stripe should be on the inside.
Breathing Tube and Water Chamber
It is recommended that you inspect the device and all parts on a regular basis after cleaning. The
equipment should be replaced immediately if there is any sign of cracking, discoloration or leaking.
NOTE: The device is not serviceable and should not be disassembled. Refer all therapy and equipment
queries to your healthcare provider.
The device is intended for single or multiple use. Refer to the high-level disinfection instructions in the
Healthcare Provider Manual for cleaning between users.