Advertisement

Insertion/Wet-Tap pH and ORP Assembly (No ATC)
Product Instructions

Description of Parts

S675
Insertion assembly with 12" insertion depth, 18" total length.
S676
Insertion assembly with 18" insertion depth, 24" total length.
S677
Insertion assembly with 24" insertion depth, 30" total length.

Mechanical Installation

1. Unscrew the hex nut from the adapter body and retract rod
into body until you hear a "CLICK" (FIG 1)
2. Unscrew the white or black nut from the adapter body and
pull out the insertion tube and the parts mounted on it as
shown in FIG. 2.
3. Thread body assembly into 1" full-port ball-valve (user sup-
plied) as shown in FIG. 3. Be sure and seal threads with Pipe
Thread Sealing Tape.
4. Install electrode into rod assembly by pushing electrode
into large end of rod and rotate to align connector as shown
in FIG. 4. Apply liberal amount of 0-ring grease to o-rings on
electrode. After engaging connectors, rotate 1/4 run to lock
electrode in place. DO NOT INSERT INSERTION ROD
ASSEMBLY INTO SOLUTION WITHOUT ELECTRODE AT-
TACHED!!!
5. Insert rod into body assembly until o-ring face seal seals
against body. Then slide white nut over rod from back side
and tighten onto body. (FIG. 5)
6. Open ball valve then loosen hex nut and push rod into de-
sired position. Tighten nut to hold position. (FIG. 6)
7. Push cap cable assembly into insertion rod and rotate to
align tabs of BNC receptacle on rod with grooves of BNC in cap.
CAUTION: MAKE SURE ELECTRODE IS ORIENTED AS
SHOWN IN FIG 4. OR WITH GLASS POINTED DOWN-
WARD WHEN INSTALLING OR REMOVING SO THAT
LIQUID CANNOT FLOW BACK INTO THE INTERNAL
CONNECTOR!
Parts covered by this product data sheet include:
S675, S676, S677
Page 1 of 3
PRODUCT INSTRUCTION SHEET
FIG. 1
RETRACT INTO
HOUSING UNTIL
YOU HEAR A
FIG. 2
FIG. 3
FIG. 4
1/4 TURN
CLOCKWISE
AFTER PUSHING IN
ALIGN TABS
AND PRESS IN
TO ENGAGE
Grease all o-ring seals
before installation
TURN
COUNTER-
CLOCKWISE
BEFORE
RETRACTING
ROD
HEX NUT
CLICK
Form: InstallInsert [Rev: 2012-05-16]
©2011 Sensorex Corporation

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the S675 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Sensorex S675

  • Page 1 WARD WHEN INSTALLING OR REMOVING SO THAT TO ENGAGE LIQUID CANNOT FLOW BACK INTO THE INTERNAL CONNECTOR! Grease all o-ring seals before installation Parts covered by this product data sheet include: S675, S676, S677 Page 1 of 3 Form: InstallInsert [Rev: 2012-05-16] ©2011 Sensorex Corporation...
  • Page 2 FIG. 7 1/4 TURN FIG. 8 TURN COUNTER- CLOCKWISE BEFORE RETRACTING RETRACT INTO HOUSING UNTIL YOU HEAR A CLICK Parts covered by this product data sheet include: S675, S676, S677 Page 2 of 3 Form: InstallInsert [Rev: 2012-05-16] ©2011 Sensorex Corporation...
  • Page 3 SHOWN IN FIG 4. OR WITH GLASS POINTED DOWNWARD WHEN INSTALLING OR RE- MOVING SO THAT LIQUID CANNOT FLOW BACK INTO THE INTERNAL CONNECTOR! Parts covered by this product data sheet include: S675, S676, S677 Page 3 of 3 Form: InstallInsert [Rev: 2012-05-16] ©2011 Sensorex Corporation...

This manual is also suitable for:

S676S677

Table of Contents