3. Prepare a quick connect Air Hose by inserting a
metal hose clamp (P/N 1211-0036, standard on
MBT 201A systems) & a male quick connect hose
mount Fitting (P/N 1259-0087) into each end of a
54 inch (137cm) length of Air Hose. Push the
ridged ends of the Fittings into the hose; slide &
twist the metal hose clamps over the connections
to secure. You may already have this piece if you
have other quick disconnect handpieces
configured.
4. Prepare a VisiFilter in the following manner.
a) Connect a 1 inch (2.5cm) length of clear pvc
Air Hose to each side of the VisiFilter; push and
turn hose onto VisiFilter nipple to seat.
b) To the free end of the Air Hose connected to the
FLOW IN side of the VisiFilter, insert the ridged
end of a female quick connect hose mount Fitting
(P/N 1259-0086).
c) Insert the ridged end of a male quick connect
hose mount Fitting (P/N 1259-0087) in the free
end of the Air Hose connected to the FLOW
OUT side of the VisiFilter.
d) Connect the VisiFilter Air Hose (with attached
male quick connect hose mount Fitting) to the
power source AUTO SNAP-V AC Port.
5. For vacuum, insert the male quick connect hose mount Fitting connected to the long Air Hose into the
female Fitting on the 1 inch (2.5cm) Air Hose (connected to VisiFilter). For pressure, insert the male
quick connect hose mount Fitting directly into the Controllable PRESSURE Port.
When removing any Air Hose, turn and pull. DO NOT attempt to pull Air
Hose directly off. Damage to or breakage of Vacuum Fitting or VisiFilter
may occur.