Henry Schein PowerSpin 112-6907 Operation Manual

Lx centrifuge with variable speed and timer

Advertisement

Quick Links

PowerSpin LX Centrifuge
with Variable Speed and Timer
Item # 112-6907 with 6 place rotor
Item # 112-6909 with 6 place rotor
Operation MANUAL
Revision Date 2014/03
Distributed by (in US only):
HENRY SCHEIN INC.
135 Duryea Road
(manufacturer)
Melville, NY 11747 USA
Henry Schein U.K. Holdings Ltd.
Medcare House, Centurion Close
Gillingham Business Park
Gillingham ME8 0SB U.K.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PowerSpin 112-6907 and is the answer not in the manual?

Questions and answers

Summary of Contents for Henry Schein PowerSpin 112-6907

  • Page 1 Item # 112-6909 with 6 place rotor OPERATION MANUAL Revision Date 2014/03 Distributed by (in US only): HENRY SCHEIN INC. 135 Duryea Road (manufacturer) Melville, NY 11747 USA Henry Schein U.K. Holdings Ltd. Medcare House, Centurion Close Gillingham Business Park Gillingham ME8 0SB U.K.
  • Page 2 Read this manual before operating Power-Spin™ LX centrifuge Introduction Henry Schein, Inc. has been a long standing provider of quality laboratory equipment for the medical, veterinary, and educational market. With a range of bench top centrifuges designed to meet today’s stringent lab needs, as well as a full range of microscopes, Henry Schein is constantly searching for newer and better technology.
  • Page 3 The warranty excludes any damage caused by shipping. Inspect for any possible shipping damage. If you note any damage, please contact HENRY SCHEIN. Save the original packaging. If it becomes necessary to return the instrument to us, we ask that it be returned in the original packaging.
  • Page 4 Contact support at HENRY SCHEIN immediately at 800-588-9776 or 732-274-1155. Please be prepared with the serial # (found on the rear of the machine near the electrical...
  • Page 5 Calibration According to Code of Federal Regulations Title 21, centrifuges require verification or calibration as follows: 1) Before initial use; 2) after repair or adjustments; and 3) 6 month after use. The centrifuge timer, however, should be checked for accuracy at least every 3 month. 2) The speed should fall within ±100 rpm of the speed display.
  • Page 6 Use a Philip type screwdriver to take the snap-in type holder cover off. Remove the blown fuse and replace it with the same type and rating fuse which is: 250V e) Put the snap-in fuse holder back. 5. Intermittent motor operation; or no operation. Call support at HENRY SCHEIN: 800-588-9776 or 732-274-1155.
  • Page 7 Specifications Speed________________________________500 – 4,000 rpm Speed accuracy ± 100 rpm @ 3000 RPM Width________________________________10.2 inches Depth________________________________11.8 inches Height________________________________8 inches Weight________________________________18 lbs...