Quick Links

Operation Instructions for
CALIDOSE
for use with Lantheus Xenon Xe-133 Gas
Vial holder
Dispenser
DISPENSER
®
LOADING
1.
Separate shielded vial holder from dispenser
2.
Check Huber point needles inside the dispenser
to ensure that they are not blocked (if necessary,
clean by pushing a fine wire through needles)
3.
Attach a hypodermic needle (or other connector)
securely to Luer Lock fitting on front end of
dispenser.
4.
Remove cap of the Xenon Xe-133 lead shipping
tube, being careful to point opened tube away
from body.
5.
Place the open end of the shielded vial holder
tightly against the open end of the shipping tube.
6.
Simultaneously invert both to allow one vial to
slip from shipping tube into shielded vial holder.
7.
Carefully separate the pieces shown in figure,
being certain to point the open ends away from
the body. There should be one vial in the shielded
vial holder, septum end out. Recap the lead
shipping tube.

Advertisement

loading

Summary of Contents for Lantheus Medical Imaging CALIDOSE DISPENSER

  • Page 1 Operation Instructions for CALIDOSE DISPENSER ® for use with Lantheus Xenon Xe-133 Gas LOADING Separate shielded vial holder from dispenser Check Huber point needles inside the dispenser to ensure that they are not blocked (if necessary, clean by pushing a fine wire through needles) Attach a hypodermic needle (or other connector) securely to Luer Lock fitting on front end of dispenser.
  • Page 2 To order from Lantheus, contact your Lantheus Sales Representative or call Customer Service: 1-800-299-3431 www.lantheus.com CALIDOSE, Lantheus Medical Imaging, and the corporate logo are registered trademarks of Lantheus Medical Imaging, Inc. ©2021 Lantheus Medical Imaging, Inc. All rights reserved. NUC 231 January 2021...