Page 3
The information contained in this document is subject to change without notice. Reproduction or republication of this document in any form or format is prohibited without the written consent of QIAGEN Lake Constance GmbH. Trademarks Trademarks and labels used are the property of their respective owners.
Environment ................. 8 Maintenance and Care ..........22 Safety Instructions............... 9 Warranty and Customer Services ........ 22 Contents of the FLUO SENS integrated Package ..10 15.1 Liability for Defects ............23 Start up ................10 15.2 Warranty ................23 Mounting ................
Consequently the detector module provides direct results, not just raw data. Measurement results can be accessed via a simple serial interface. External devices such as PCs, embedded controllers or PLCs can readily access measurement results via the integrated serial interface. FLUO SENS integrated Manual January 2013...
265nm .. 980nm wavelengths MOLEX 52610-1071 Available detection 365nm .. 980nm wavelengths Specifications given in the list above are for reference only. Exact specifications for the individual detectors are given in the corresponding data sheets. FLUO SENS integrated Manual January 2013...
Page 6 of 24 3 Connections The FLUO SENS integrated features a 10way flex cable in the rear part of the housing (see 4.1 and 4.2). This flex cable is used for communication and power supply. Dimensions are shown left.
The housing of the detector is connected to ground return. To achieve precise alignment between the detector and the sample, the front lens is designed with a Ø8f8 fit (shown bottom left). ESMO40-MB-xxxx Detector Dimensions. ESMO30-MB-xxxx Detector Dimensions. FLUO SENS integrated Manual January 2013...
Do not expose the device to direct sunlight Protect the device from high humidity and contact with liquids Do not expose the device to excessive heat ESMO35-MB-xxxx Detector Dimensions. Do not expose the device to strong electromagnetic radiation FLUO SENS integrated Manual January 2013...
Warning! Do not look directly into the optics! Detectors emit The device should only be connected and strong light radiation that can cause permanent disconnected in power-off mode. Warning! damage to the eye. Danger! FLUO SENS integrated Manual January 2013...
Page 10 of 24 6 Contents of the FLUO SENS integrated Package Carefully unpack the contents of the package and check that all components (shown left) are contained therein. For claims, please contact customer service. The contains the following components: ...
1 and 254 before being connected. Each attached detector can be simultaneously prompted to carry out a fluorescence measurement by using the Start command in tandem with the MODBUS address „255“ or the external trigger input. FLUO SENS integrated Manual January 2013...
(EMVG), the radiation can be suppressed by routing the connecting cable through a suitable ferrite core, for example Würth Elektronik 7427221. Additionally, the FLUO SENS Integrated housing can be earthed to ground using one of the mounting threads. FLUO SENS integrated Manual...
MODBUS is based on a master/slave system. The master is a superordinate controller, for example a PC or a programmable controller. The MODBUS Controllers in the FLUO SENS integrated are slave devices. There are no bus conflicts, since only a single node sends at a time.
Detectors always respond to Write/Read Access Function Code 06: Provides write access to 16bit-registers with MODBUS address 0. This is useful, for example, if the address in register 165 has been altered. Tip! FLUO SENS integrated Manual January 2013...
Value Lo (Register 235) 00 Value Lo 02 Number of Registers Lo 02 Value Hi (Register 236) 00 LRC 06 LRC 04 LRC 07 LRC F1 End Frame CRLF End Frame CRLF End Frame CRLF End Frame CRLF FLUO SENS integrated Manual January 2013...
LEDx Power Min Minimum allowed set value for “LEDx Power” Auto trigger measurement: Measurement is started at power on, all values are taken after one h/w trigger FLUO SENS integrated Manual January 2013...
Page 17 of 24 10.6 FLUO SENS integrated Memory Allocation FLUO SENS integrated Manual January 2013...
Page 19
Set Modbus Address to 2 you have to write 2 * 256 = 512 to the detector word address 164. Registers 400 to 511 are general purpose registers for storage of any user data (available in firmware version 1.46, dated 19.12.2011 and higher). FLUO SENS integrated Manual January 2013...
The measurement values must be read cyclically from the respective registers if the ticket counter has been altered. With cyclical readouts, the current measurement values will each be overwritten by the latest valid measurement values. FLUO SENS integrated Manual January 2013...
The FL Digital program allows the user to collect and display data from the FLUO SENS integrated. Instrument sensitivity parameters can be setup and saved to the FLUO SENS integrated, data can be stored on disk. A detailed description of...
Try communication with MODBUS address „0“. Test example in section 10.7. Cause: Reset-wire (see section 3) is grounded. Correction: Examine the level of the reset wire and, if necessary, connect via a resistor to +5V. FLUO SENS integrated Manual January 2013...
In case of a defective device, please do not attempt to repair the device yourself. There are no self-maintenance parts within the device. In case of any possible defects, please contact QIAGEN Lake Constance GmbH (address given on page 2).
Lake Constance GmbH. Transportation of the defective product repair or replace the product and return it to you as fast as is to be insured by the business partner of QIAGEN Lake possible. Please be aware that it may take several days to Constance GmbH.
Page 25
Products are subject to the quality standards in compliance / accordance with DIN ISO 9001. QIAGEN Lake Constance GmbH Jacques-Schiesser-Strasse 3 D-78333 Stockach Germany Website www.qiagen.com Email: info@qiagen.com...