Do you have a question about the Imago and is the answer not in the manual?
Questions and answers
Summary of Contents for ECM Imago
Page 1
User Manual Version 1.06.1 Imago reference: 90-1945 User Manual reference: 70-2607 Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 1 sur 20...
......................18 LEANING AND DISINFECTION OF THE DEVICE 7.2 C ......................18 LEANING AND DISINFECTION OF PROBES 8. ENVIRONMENTAL CONDITIONS OF USE...................... 19 Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 2 sur 20...
You may then click together one part of the body strap(elastic). Then, put the left or right hand inside the neck strap and place the Imago in front of you Finally, you can click together the other part of the body strap.
3. Safety instructions 3.1 Probe information 3.1.1 Probe types and corresponding applications The probes provided with the Imago device are the following : - LB760P Linear Rectal probe, 7,5 MHz (this rectal probe is used for gynaecology and reproduction diagnosis).
Then wait until the system starts and displays the user interface on the screen. The Imago equipment starts in B mode with the probe connected on the probe connector. If no probe is connected to the probe connector, the system will display a label “No probe”. In this case, connect a probe to the connector and press “Freeze”, or restart the equipment.
In Freeze mode of the ultrasound image, the right page displays and the functions below are available. To return to the homepage, just press Freeze again Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 6 sur 20...
Simply slip your finger right to left (or left to right), in order to reach the good page (in live or frozen mode). Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 7 sur 20...
Store I : allows to register an image. Store C : allows to register the last few seconds on the scanned session (clip/video). Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 8 sur 20...
Page 9
USB drive directly if connected. If no USB is connected the review mode will access the internal memory of the device. Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 9 sur 20...
Page 10
Image Formats : Choose the format for picture saving (bmp, jpg, imag, dcm). Dicom Local Store : Digital Imaging and Communications in Medicine (DICOM). Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 10 sur 20...
Page 11
If Restore the last annotation is activated, the last annotation will be automatically proposed. System System Datas about your version of IMAGO (Date & time format, Firmware, Factory). Also the user can select: Auto Freeze, Force store clip with USB, Keypad, Right-Handed / Left-Handed. • System : change the date or time format •...
Page 12
FULL SCREEN Reset System Reset the configuration system of IMAGO with factory settings. User Protocol System Access to the User Protocol menu to create user-defined protocols. The contextual menu includes: •...
Page 13
Fat : Automatic measurement of back fat (for pigs with sector or linear probes ; cattle or small ruminants with linear probe). Available in frozen mode also. Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 13 sur 20...
Page 14
CINELOOP : Replay the last seconds of scanning that have just been done (124 pictures with linear and convex probes ; 62 pictures with sectorial probe). Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 14 sur 20...
B mode menu (it unfreezes the picture). 5.1.3 Way to move from one Interface to the other. Go to SETTINGS function Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 15 sur 20...
Page 16
Having ticked "SIMPLIFY", you have access to the simplified mode. A single menu is so proposed in live mode. If you untick the compartment " SIMPLIFY " you go to the "normal" said menu. Cf 5.1.2 and 5.2.2 menu. Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 16 sur 20...
6. Connection of peripherals The Imago equipment can be connected to several peripherals like : TURN ON/OFF CHARGE LED BATTERY POWER SUPPLY COMPOSITE VIDEO USB PORT CHARGE FOR GOGGLES BLUETOOTH PORT Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx...
WARNING : The use of any other disinfecting solution than the ones indicated above is dangerous as it may damage the probe components. Check the list of specified agents carefully. Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 18 sur 20...
Never use the device if these environmental conditions are not respected. Stop operation of the device if one of these conditions is no more respected. The device can be used in any room or place respecting these environmental conditions. Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 19 sur 20...
Page 20
ECM Echo Control Medical ……..…………………… 126 bd de la République ......F-16000 Angoulême …………………………… Tel : +33 5 45 92 03 57 …………….……………… Version 1.06.1 du 29/07/2016 Réf : CD MA User Manual Imago EN 1.06.docx Page 20 sur 20...
Need help?