Life Science Suite ACQ-7700 Installation Manual Model: PNM-P3P-7001 Manual: MU00225-001 Revision 51 Data Sciences International 119 14 Street NW St. Paul, MN 55112 P: +1 (651) 481.7400 • F: +1 (651) 481.7404 www.datasci.com Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 3
DSI/Ponemah Life Science products are not “medical devices” intended to be used for the purposes of diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, or used as a life support device. Use of DSI/Ponemah products are solely for the purposes of conducting life science research.
Contents System Overview ACQ-7700 Acquisition Interface Unit..................4 ACQ-7700 Signal Conditioners....................4 System Requirements ........................ 5 Configuring the System Assistance ..........................6 Initial Inspection ........................6 Parts List ............................ 6 EPP Interface Card Installation....................7 Software Installation ........................7 Installing ACQ-7700 Signal Conditioners................. 8 Removing ACQ-7700 Signal Conditioners ................
Page 5
Product Issue Report Form ...................... 25 Feature Request Feature Request Form......................26 ii • System Overview ACQ-7700 / Type 3 Installation Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 6
ON (connection to AC mains) DSI/Ponemah Life Science products are not “medical devices” intended to be used for the purposes of diagnosis of disease or other conditions, or in the cure, mitigation, treatment, or prevention of disease, or used as a life support device.
Page 7
POWER SOURCE This instrument is intended to operate indoors from a power source that does not apply more than 250 volts RMS between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding conductor in the power cord is required. If the unit is mounted in a permanent installation (i.e.
Page 8
EMI/RFI The amplifier has been tested to meet EMI/RFI susceptibility and radiation standards. However, the user should be aware of possible electromagnetic interactions between this device and other devices in the same area and correct the situation as needed. WARNINGS FOR AUTHORIZED SERVICE PERSONNEL Dangerous voltages exist at several points in this instrument.
ACQ-7700 Acquisition Interface Unit The ACQ-7700 Acquisition Interface Unit is one of the available acquisition interfaces for the DSI/Ponemah Life Science Suite (LSS). An acquisition interface unit interfaces signals into the Ponemah Physiology Platform (P3 Plus). The ACQ-7700 has a modular design which allows the user to easily expand the number of available acquisition inputs to a maximum of 128, through the installation of various signal conditioners.
Your choice of instrumentation signal conditioning modules and analysis modules. Please view the P3 Plus Manual for system requirements. For information on DSI/Ponemah products and services, check out our website at www.ponemah.com. System Overview • 5 ACQ-7700 / Type 3 Installation Manual...
Configuring the System Assistance If you require assistance in configuring your system, contact our Technical Support department at (216) 328-7000. You can also request assistance from our support department by contacting us through our web site at www.ponemah.com or email us at lifescience@datasci.com.
EPP Interface Card Installation If the system was delivered with a configured computer or if the on board computer LPT port is going to be used, these steps can be skipped. It is highly recommended that the EPP parallel interface card be used as the interface to the ACQ-7700. Make sure that the personal computer power cord is unplugged.
Installing ACQ-7700 Signal Conditioners Turn OFF the power to the ACQ-7700 (The signal conditioners may be damaged if inserted or removed under power and the ACQ-7700 must be powered with the signal conditioners inserted for the controlling software to recognize their presence). The power switch is located on the front of the ACQ-7700.
ACQ-7700 Connections The standard ACQ-7700 has two input/output connectors on the front panel of the EPP interface board: The EPP Port connector and the AUX connector. The EPP acquisition board is installed in the first slot from the left on the front of the ACQ- 7700.
Auxiliary Connector (Preview Feature) This is a 15 pin female D-shell connector. This 15 pin D-SUB provides connection for the auxiliary functions. Signal Signal Reserved Trigger In Reserved Reserved Reserved Event Out Reserved Alarm Out Reserved Timer Out Reserved Reserved Reserved Ground Reserved...
Getting Started Introduction The power switch for the ACQ-7700 is located on the front of the unit. A green light located on the front of the Acquisition Module illuminates when power is ON. There are no operating controls for the ACQ-7700. Signal conditioner setup is performed through the graphical user interface.
After the license has been loaded, the main P3 Plus dialog will appear as displayed below. Main P3 Plus Dialog The User Name that is entered will be recalled automatically the next time the system is started. The name entered here will be used on all printouts and audit logs for identification of the data collected.
This dialog will appear every time P3 Plus is started and also when an acquisition is started or stopped (when amplifiers are near or past their calibration). There are four types of icons for easy reference. This information will be written to the Application Log. - Yellow Warning This icon will appear if the amplifier is getting close to its calibration date.
Page 19
This tab lists each module, the DSP Version, and the Last Cal Date of each amplifier. If a DSP does not exist on an amplifier it will be listed as (NA). If the calibration date cannot be stored electronically on the amplifier it will be listed as (NA). 14 •...
Maintenance Introduction This section describes routine maintenance procedures for the ACQ-7700. If the unit needs repair, consult your local Service Representative. Service/repair parts are available from the factory. A detailed parts list may be obtained from the factory upon request. Cleaning General To maintain proper operation, clean the ACQ-7700 when necessary as described...
Page 21
annually by the factory. For further information please contact Technical Support at 216-328-7000 or e-mail lifescience@datasci.com. 16 • Maintenance ACQ-7700 / Type 3 Installation Manual Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Appendix Specifications Electrical Line Voltage Auto-switching 120/240VAC (±10%) Line Frequency 50/60Hz Fuse T/250V 3.5A (Internal, not serviceable) Analog Inputs Maximum Input Number 128 (dependent on ACQ-7700 signal conditioners) Acquisition System Maximum Aggregate Sample Rate 300k samples/second Resolution 16 bit, signal conditioner dependent Auxiliary Output Outputs 3 TTL level I/O lines...
Troubleshooting If the system displays the following message on startup, ACQ-7700 Unit Not Found, then verify the following: The ACQ-7700 is powered up and the parallel port cable is connected to the personal computer. The correct parallel port is being used (LPT1/LPT2/LPT3). Select Application Configuration from the Options menu and select the Acquisition Interface tab.
Declaration Electromagnetic Emissions/Immunity Electromagnetic Emissions/Immunity Tables Guidance and manufacturer’s declaration – electromagnetic emissions The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. Emissions Test Compliance Electromagnetic environment - guidance...
Page 25
Guidance and manufacturer’s declaration - electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. Immunity test IEC 60601 test Compliance Electromagnetic environment - guidance level...
Page 26
Guidance and manufacturer’s declaration - electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified below. The customer or the user of the equipment should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance Electromagnetic environment - guidance level...
Page 27
être employés comme dispositif de soutien de la vie. L'utilisation des produits de DSI/Ponemah sont seulement pour les buts de conduire la recherche de la science de vie.
Page 28
SOURCE D'ÉNERGIE Cet instrument est prévu pour fonctionner à l'intérieur à partir d'une source d'énergie qui n'applique pas plus de 250 volts de RMS entre les conducteurs d'approvisionnement ou entre l'un ou l'autre conducteur d'approvisionnement et la terre. Une prise de terre protectrice par le conducteur de terre dans le cable électrique est exigée.
Page 29
EMI/RFI L'amplificateur a été examiné pour répondre à des normes de susceptibilité et de rayonnement d'EMI/RFI. Cependant, l'utilisateur devrait se rendre compte des interactions électromagnétiques possibles entre ce dispositif et d'autres dispositifs dans le même secteur et corriger la situation si nécessaire. AVERTISSEMENTS POUR LE PERSONNEL DE SERVICE AUTORISÉ...