Download Print this page

Advertisement

Quick Links

®

Advertisement

loading
Need help?

Need help?

Do you have a question about the Contour next EZ and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for ASCENSIA Contour next EZ

  • Page 1 ®...
  • Page 2: Intended Use

    Intended Use Information For Safety The C EZ blood glucose monitoring system ® • Read your C EZ user guide, the lancing ontour ® ontour (meter, test strips and control solution) is intended for device insert and all instructional materials provided in your use by persons with diabetes for self-testing, and by meter kit before testing.
  • Page 3 ®...
  • Page 4 ®...
  • Page 5 ®...
  • Page 6 Information For Safety • Do not press the test strip against your finger • Have all the materials you will need ready when testing. This may block the sample tip. before you begin testing. This includes your EZ meter, the C test ®...
  • Page 8 ®...
  • Page 9 ®...
  • Page 10 ®...
  • Page 11 ®...
  • Page 12 Disinfecting Your Meter Cleaning and Disinfecting Your Lancing Device Supplies Needed for Disinfecting: Refer to the lancing device insert provided with • Germicidal wipes containing 0.55% sodium your kit for detailed instructions for cleaning and hypochlorite (bleach) disinfecting the lancing device. •...
  • Page 13 ®...
  • Page 14 ®...
  • Page 15 ®...
  • Page 16 ®...
  • Page 17 ®...
  • Page 18 ®...
  • Page 19 ®...
  • Page 20 ®...
  • Page 21 ®...
  • Page 22 ®...
  • Page 23 ®...
  • Page 24 If you do not resolve the problem, contact If you do not resolve the problem, contact Customer Service. See the back cover for Customer Service. See the back cover for contact information. contact information. What What It What What It What You Should Do What You Should Do You See...
  • Page 25 If you do not resolve the problem, contact If you do not resolve the problem, contact Customer Service. See the back cover for Customer Service. See the back cover for contact information. contact information. What What It What What It What You Should Do What You Should Do You See...
  • Page 26 ®...
  • Page 27: Caring For Your Meter

    Caring for Your Meter Symptoms of High or Low Blood Glucose • Store meter in the carrying case provided whenever possible. You can better understand your test results by being • Wash and dry hands well before handling to keep aware of the symptoms of high or low blood glucose.
  • Page 28: System Specifications

    System Specifications The C EZ meter meets the requirements ontour ® of IEC 61000-4-2 for immunity to electrostatic Test Sample: Venous or capillary whole blood discharge. Avoid use of electronic devices in very Test Result: Referenced to plasma/serum glucose dry environments, especially if synthetic materials Sample Volume: 0.6 µL are present.
  • Page 29: Technical Information

    Technical Information Limitations Accuracy Preservatives: Blood may be collected by health care professionals into test tubes containing heparin. Do not The C EZ blood glucose monitoring ® ontour use other anticoagulants or preservatives. system was tested in 100 capillary blood samples using 600 C test strips.
  • Page 30 Technical Information Technical Information Table 3 – System accuracy results for glucose Table 4 – System repeatability results for concentrations between 1.0 mmol/L and  eZ meter using C ® ® ontour ontour 25.6 mmol/L test strips Pooled 95% Confidence Within ± 0.83 mmol/L or ±15% 600 of 600 (100%) Mean, Coefficient of...
  • Page 32 3. Ascensia Diabetes Care reserves the right to make ASCENSIA DIABETES CARE MAKES NO changes in the design of this instrument without OTHER EXPRESS OR IMPLIED WARRANTY FOR obligation to incorporate such changes into THIS PRODUCT. THE OPTION OF REPLACEMENT, previously manufactured instruments. DESCRIBED ABOVE, IS THE ONLY OBLIGATION OF ASCENSIA DIABETES CARE UNDER THIS...
  • Page 33: Replacement Parts

    Supplies When calling or writing for supplies be sure to include the number with the name of the replacement part or accessory item. Replacement Parts Part Number* Item 40030030 Two 3-volt lithium batteries DL2032 or CR2032 84310760 EZ user guide ®...
  • Page 35 4052 Basel, Switzerland Customer Service: 1-800-268-7200 www.diabetes.ascensia.com For information on patents, and related licenses, see: www.patents.ascensia.com Ascensia, the Ascensia Diabetes Care logo, Clinilog, Contour, Glucofacts and the No Coding logo are trademarks and/or registered trademarks of Ascensia Diabetes Care Holdings AG. All other trademarks are the property of their respective owners.