IMPORTANT INFORMATION: File in your maintenance records
CastVac
TM
Maintenance Manual & Operating Instructions
Stryker Instruments
European Authorized Rep:
4100 E. Milham
RA/QA Manager
Kalamazoo, Michigan
Stryker France
(USA) 49001
ZAC Satolas Green Pusignan
269-323-7700
Av. de Satolas Green
1-800-253-3210
69881 MEYZIEU Cedex
www.stryker.com
France
3/06
Syste m Control
REF 986
REF 987
REF 996
986-1-11 Rev-L

Advertisement

Table of Contents
loading

Summary of Contents for Stryker CastVac REF 986

  • Page 1 IMPORTANT INFORMATION: File in your maintenance records CastVac Maintenance Manual & Operating Instructions Syste m Control REF 986 Stryker Instruments European Authorized Rep: REF 987 4100 E. Milham RA/QA Manager Kalamazoo, Michigan Stryker France REF 996 (USA) 49001 ZAC Satolas Green Pusignan 269-323-7700 Av.
  • Page 2 • Use only Stryker approved accessories. Other accessories may result in increased emissions or decreased immunity of the system. Contact your Stryker sales representative for a complete list of accessories. DO NOT modify any accessory. Failure to comply may result in patient and/or operating room...
  • Page 3: Symbol Definition

    Stryker Instruments Representative or call Stryker Customer Service at (800) 253-3210. (Outside the USA, contact your nearest Stryker Subsidiary.) The Stryker CastVac simultaneously suctions dust and chips as you cut casting material. The CastVac is designed to be used in combination with the Stryker Cast Cutter.
  • Page 4: Hose Attachment

    Hose Attachment Hose/Cast Cutter Attachment The vacuum hose must be securely connected to both the can- ister and the Cast Cutter to provide optimum power. 1. Depress the button located on the hose cuff and align it with the hole in the canister connector. Push hose into canister connector.
  • Page 5: Operation

    2. Plug the Stryker Cast Cutter into the receptacle located on the side of the canister. See illustration. Wrap excess cord length around the handles of the CastVac Stand.
  • Page 6: Filter Replacement

    DO NOT use if damage is apparent. Activity/Item Interval Tools and Equipment Required Instructions Models REF 986 & 987 Instructions Model REF 996 Source cord Annually Check for cracks. Return to Stryker Check for cracks. Replace for for repair if required. cord.
  • Page 7: Troubleshooting

    Make sure filter cartridge is clean. Ensure adequate ventilation is provided for exhaust. Refer all other problems to your Stryker Instruments representative or call Stryker Instruments Customer Service at 1-800-253-3210. Outside the US, contact your nearest Stryker subsidiary. To ensure the longevity, performance, and safety of this equipment, package in original package materials when storing or...
  • Page 8: Specifications

    Outlet: 220-240V~ Maximum load 0.75A Use with Stryker Cast Outlet: NEMA 5-15R " Hospital Grade" 100-120V~ Cutter REF 941 Use with Stryker Cast Cutter models CLASS I REF 840, 848, 940 only. CLASS I IPX0 Ordinary Equipment Duty Cycle: Continuous Operation...
  • Page 9 Guidance and manufacturer's declaration - electromagnetic emissions The CastVac REF 996 is intended for use in the electromagnetic environment specified below. The customer or the user of the CastVac REF 996 should assure that it is used in such an environment. Compliance Emissions test Electromagnetic environment - guidance...
  • Page 10 Guidance and manufacturer's declaration - electromagnetic immunity The CastVac REF 996 is intended for use in the electromagnetic environment specified below. The customer or the user of the CastVac REF 996 should assure that it is used in such an environment. Immunity test Compliance level IEC 60601...
  • Page 11 Guidance and manufacturer's declaration - electromagnetic immunity The CastVac REF 996 is intended for use in the electromagnetic environment specified below. The customer or the user of the CastVac REF 996 should assure that it is used in such an environment. Immunity test IEC 60601 Compliance level...
  • Page 12 Recommended separation distances between portable and mobile RF communications equipment and the CastVac REF 996 The CastVac REF 996 is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the CastVac REF 996 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CastVac REF 996 as recommended below, according to the maximum output power of the communications equipment.
  • Page 13: Reorder Information

    This warranty does not cover damages caused by misuse or by failure to follow procedures outlined in this manual or as demonstrat- ed by the Stryker Instruments representatives. This instrument is not warranted if it is used for purposes other than that of a cast vacuum.
  • Page 14 Euro Equiv. 0986-001-715 N. Euro Equiv. 0986-001-730 Japanese Equiv. 0986-001-725 Stryker Instruments European Authorized Rep: 4100 E. Milham RA/QA Manager Kalamazoo, Michigan Stryker France (USA) 49001 ZAC Satolas Green Pusignan 269-323-7700 Av. de Satolas Green 1-800-253-3210 69881 MEYZIEU Cedex www.stryker.com...