Codes

Code Alerts
E-0
Software error detected
E-1
Hardware error detected
E-2
Meter temperature is outside
of the range for testing
E-3
Defective or previously used
test strip is detected
E-4
A short sample (control
solution or blood) is identified
LO or If test results are outside of the
HI
measuring range of the meter
Battery level is low
36

Error Messages

Action
Perform the test again. If you get the
same error, call Nova Technical Support.
Perform the test again. If you get the
same error, call Nova Technical Support.
Move the meter to an area
where the temperature is acceptable
(41°-113°F), allow meter to adjust to the
temperature. Repeat testing.
Repeat test with a new
test strip.
Repeat test with a new test
strip and adequate sample.
Repeat test. If you get the
same result, consult your physician.
Replace the battery