Summary of Contents for Cerlic MultiTracker

  • Page 1 Multi Tracker MultiTracker 2019-02-05 C84G5EN19...
  • Page 2: Table Of Contents

    8. TROUBLE SHOOTING ..........................15 9. SPARE PARTS ............................. 15 10. CONTACT INFORMATION ........................15 11. WARRANTY .............................. 16 12. TECHNICAL SPECIFICATIONS, MULTITRACKER HAND CONTROLLER ......... 16 APPENDIX 1, BLANKO (SLUDGE BLANKET SENSOR) ................ 17 Description of the function ............................17 Measurement Functions ..............................17 Sensor body ..................................18...
  • Page 3 Multi Tracker APPENDIX 2, OXYDUO (DO SENSOR) ...................... 28 Description of the function ............................28 Measurement Functions ..............................28 Sensor body ..................................28 ..................................30 AIN MENU Description of the Display ............................29 Settings ....................................30 Calibration ..................................32 Getting Started ................................33 ............................... 34 PECIFICATION XYDUO Maintenance ..................................33 ............................
  • Page 4: Ecurity Information

    It is of great importance to read all parts of this manual prior to start up of the General information instrument. If the Multitracker is not used and handled according to this manual, then the life and functionality may be jeopardized, and all warranties will be void.
  • Page 5: Emoving The Cable Hook

    Multi Tracker In order to facilitate the handling of the MultiTracker it is supplied with a SS cable hook Cable Hook that is attached to the battery case, see fig 2. It is possible to wind up all of the cable on the cable hook.
  • Page 6: Turning On The Hand Controller

    Multi Tracker 3. Turning On the Hand Controller It is important that any sensor that is attached to the hand controller is in the air when the hand controller is powered up. Sensors should not be in the liquid since this will effect the pressure sensor! To start measuring with the instrument press the key (on/off).
  • Page 7 Multi Tracker left side of the screen. Different sensors attached to the hand unit will display different information on the display. The screen shoot below is showing an attached Blanko sensor.          fig 6 ...
  • Page 8 Multi Tracker        fig 3 Profile displayed in x=mg/l and y=depth  Solids Concentration Depth – current  Fluff level  Blanket depth  Maximum depth  Sensor position during depth  measuring versus max. depth ...
  • Page 9: Main Menu

    Multi Tracker 5. Main menu Press to open the main menu. See fig 8. to pick the desired sub menu and open with fig 8 In the SETTINGS menu (fig 9) it is possible to set thresholds concentrations. In this Settings menu the span (max depth) measurement is set to fit the profile scale maximum depth.
  • Page 10: Setup

    Multi Tracker The SETUP menu (fig 11) enables selection of language, time and date formats, SETUP temperature units, concentration format, and depth format. fig 11 The following languages are available; Swedish, English, German and French. To select Language the desired language, perform the following steps; Press to open the menu and step down with and select ”SETUP”...
  • Page 11: Concentration Units

    (text mode or profile mode). Up to 250 measurements may be stored in the MultiTracker internal memory. Every stored measurement is saved and it contains all data available at the actual measurement. Each individual line in the log may be tagged with 10 alphanumerical characters;...
  • Page 12: Getting Started

    When all settings are done, exit with until one of the two main windows appear (text mode or profile mode). This data is stored in the MultiTracker controller and not in the sensor. 6. Getting Started...
  • Page 13: Ownloading Data From The

    You also need a PC software named Traker Talk to interface with the hand unit and move the stored data to the PC. The software can be downloaded free of charge from the Cerlic home page. fig 15 To transfer the stored data, disconnect the sensor from the hand controller and attach the communication cable to the same M12 connector, fig 16.
  • Page 14: Maintenance

    The sensor should be cleaned, if any solids or fouling of the measuring windows occurs. Inspection of the sensor When cleaning the sensor, it is preferred to use the Cerlic Sensor Cleaning (CSC) liquid. It is also possible to use a soft cloth and water. Pay attention not to scratch the quartz measuring windows or damage the built in pressure sensor.
  • Page 15: Trouble Shooting

    Cerlic or a local Cerlic representative. In the event of returning the MultiTracker to Cerlic for checkup or repair, please make sure to use the form for Return of Material (RMA) prior to shipping . The RMA document can be downloaded from the Cerlic web page www.cerlic.com.
  • Page 16: Warranty

    Instruments delivered from Cerlic Controls AB, are carefully checked and tested prior to 11. Warranty the shipment. Cerlic will repair or replace the product if a problem related to manufacturing or design occur during the warranty period. Number of Detectable Levels 12.
  • Page 17: Appendix 1, Blanko (Sludge Blanket Sensor)

    Multi Tracker Appendix 1, Blanko (Sludge Blanket Sensor) Blanko Sludge Blanket Sensor Blanko is a portable optical suspended solids meter designed to measure sludge blanket Description of the function depth and suspended solids in clarifiers, thickeners, aeration basins, etc. in water and wastewater plants, as well in other facilities.
  • Page 18: Sensor Body

    Multi Tracker The sensor body contains optics and electronics that should not be exposed to Sensor body mechanical abuse or high temperatures. If the sensor body has mechanical damages, water may penetrate into the sensor and destroy the electronics and optics. Please see section Maintenance for more information.
  • Page 19: Description Of The Display In Profile Mode

    Multi Tracker When pressing the Profile button marked the display will show the sludge profile. Description of the Display in Profile Mode See fig 3 below. At any time during a measurement it is possible to switch between text mode and profile mode by pressing the button without affecting the measuring value.
  • Page 20: Main Menu

    Multi Tracker Press to open the main menu. See fig 5. Main menu to pick the desired sub menu and open with fig 5 In the SETTINGS menu (fig 6) it is possible to set the following functions: Settings Threshold 1 – preset Fluff concentration Threshold 2 –...
  • Page 21 Multi Tracker fig 7 Press to open the menu and select ”SETTINGS” confirm with . Step down with and pick ”Threshold 1” with . Step up with or down with until desired concentration is obtained. Confirm and step to next digit with .
  • Page 22 . When all settings are done, exit with until one of the two main windows appear (text mode or profile mode). See Concentration Units in section for MultiTracker hand controller. Concentration Units See Temperature units in section for MultiTracker hand controller.
  • Page 23: Calibration

    It is also possible to perform a manual calibration of zero and one meter (3.3 ft) depth. This is done by starting the MultiTracker and perform a zero and 1 meter calibration of the depth. fig 9...
  • Page 24: Getting Started

    ON after you have completed zero & 1 m Must turn Tracker OFF/ON after zero & 1 m depth calibrations depth calibrations. This resets the MultiTracker to accept the new depth values. Getting Started Start the MultiTracker by pressing the button marked .
  • Page 25 Multi Tracker In order to switch between text mode and profile mode, press the button marked . It Profile mode is possible to step back and forth with this button without affecting the actual measurement. To save a measurement profile, or to perform a new measurement, press , when Save a profile doing this, an option to save...
  • Page 26 In case of any malfunction that is not possible to correct with a new calibration, please contact Cerlic or a local Cerlic representative. In case of sending the Blanko to Cerlic for check up or repair, please make sure to use the form for Return of Material (RMA) prior to dispatch.
  • Page 27: Specification, Blanko

    Multi Tracker Function Sludge Concentration and Depth Specification, Blanko Measuring principle Optical light transmission Wave length NIR 850 nm Measuring range Max – 20,000 mg/l (ppm) Accuracy suspended solids 1% FS (full scale) Repeatability < 2% of Measured Value Type of Measurement Continuous with Profile Number of Detectable Levels Two (fluff and sludge blanket)
  • Page 28: Appendix 2, Oxyduo (Do Sensor)

    Multi Tracker Appendix 2, Oxyduo (DO Sensor) Oxyduo DO Sensor Oxyduo is a portable oxygen meter designed to measure oxygen levels in aeration Description of the function basins, etc. in water and wastewater plants, as well in other facilities. It is possible to display measured value as text on the screen and set two different alarm levels for high and low levels.
  • Page 29: Description Of The Display

    Multi Tracker At start up, the display is shown in fig 1 and description in fig 2. Value for oxygen Description of the Display concentration are displayed on the left and alarm values to the right. Date, time and temperature are shown across the top. ...
  • Page 30: Main Menu

    Multi Tracker Press to open the main menu. See fig 3. Main menu to pick the desired sub menu and open with fig 3 In the SETTINGS menu, below, it is possible to set alarm values for high and low oxygen Settings level concentrations.
  • Page 31 See preferences in section for MultiTracker hand controller. Preferences See language in section for MultiTracker hand controller. Language See Concentration Units in section for MultiTracker hand controller. Concentration Units See Temperature units in section for MultiTracker hand controller. Temperature units...
  • Page 32: Calibration

    Multi Tracker Calibration Calibration of sensor with Optical cell Remove the black protective cover around the new cell's connector. End the circuit with the bracket as shown. Connect the cell. Zero calibration of the sensor: • Open the main menu with •...
  • Page 33: Getting Started

    By entering the log position by tag name, all actual information for the stored measurement may be retrieved. At delivery the MultiTracker is preset with the following values; Threshold 1, 9mg/l Alarm values (High oxygen level) and Threshold 2, 2mg/l (Low oxygen level). Using these values, it is possible to instantly start to measure.
  • Page 34: Oxyduo - Chart

    In case of any malfunction that is not possible to correct with a new calibration, please Trouble shooting contact Cerlic or a local Cerlic representative. In case of sending the Oxyduo to Cerlic for checkup or repair, please make sure to use the form for Return of Material (RMA) prior to dispatch.
  • Page 35: Solubility

    Multi Tracker Oxygen Solubility - chart...
  • Page 36: Appendix 3, Solido (Suspended Solids Sensor)

    Multi Tracker Solido Appendix 3, Solido (Suspended Solids Sensor) Suspended Solids Sensor Solido is a portable optical suspended solids meter designed to measure suspended Description of the function solids in aeration basins, RAS, aerobic digesters, etc. in water and wastewater plants, as well in other facilities.
  • Page 37: Description Of The Display

    Multi Tracker At start up, the display is shown in fig 1 and description in fig 2. Value for suspended Description of the Display solids concentration is displayed on the left and alarm values to the right. Date, time and temperature are shown across the top.
  • Page 38: Main Menu

    Multi Tracker Press to open the main menu. See fig 3. Use to pick the Main menu desired sub menu and open with . Exit the menu with until one of the main windows appear. fig 3 In the SETTINGS menu (fig 4) it is possible to set alarm values for high and low Settings suspended solids level.
  • Page 39 See preferences in section for MultiTracker hand controller. Preferences See language in section for MultiTracker hand controller. Language See Concentration Units in section for MultiTracker hand controller. Concentration Units See Temperature units in section for MultiTracker hand controller. Temperature units...
  • Page 40: Calibration

    Multi Tracker Calibration can be done for both zero and lab sample concentration. Calibration fig 7 Press to open the menu and step down with . Select ”Calibration” Zero Sample with . Step down with and select ”zero”. Confirm with .
  • Page 41: Getting Started

    *This function is supported with software version MT2.3 and above. Getting Started Start the MultiTracker by pressing the button marked . To switch OFF/ON the unit, Getting started with the Solido press the same button.
  • Page 42 In case of any malfunction that is not possible to correct with a new calibration, please Trouble shooting contact Cerlic or a local Cerlic representative. In case of sending the Solido to Cerlic for checkup or repair, please make sure to use the form for Return of Material (RMA) prior to dispatch.
  • Page 43: Specification, Solido

    Multi Tracker Function Suspended solids Specification, Solido Measuring principle Optical light transmission Wave length NIR 850 nm Measuring range Up to 35,000 mg/l (ppm) dependent on solids type Accuracy suspended solids 1% FS (full scale) Repeatability < 2% of Measured Value Type of Measurement Continuous Number of Detectable Levels...