INSTALLATION, OPERATION OR MAINTENANCE OF GMW ASSOCIATES PRODUCTS. This manual is for the model 3470 electromagnet with serial numbers 206 and above. For the model 3470 electromagnet with serial numbers 196 to 205 see manual M3470d For the model 3470 electromagnet with serial numbers 195 and below see manual M3470c File No: M3470h.407...
Page 5
Section 1 SPECIFICATIONS Table 1. Model 3470 Specifications Pole Diameter: 45mm (1.75 inch) Pole Gap: 0 - 75mm (0 to 3 inch) Standard Pole Face: 40mm (1.57 inch) cylindrical end. 20mm ( 0.79 inch) tapered end. Coils (series connection) coil resistance (20°C) 7.3 Ohm...
Page 6
Section 1 SPECIFICATIONS Table 2. Model 3470 Electrical and Water Connections DC Current (refer to Drawing 11801470) Right hand coil terminal 2 Positive Left hand coil terminal 1 Negative Ground An M4 screw (Item 20 on drawing 11801470) is provided on the magnet yoke to enable the magnet to be grounded according to local safety regulations.
Section 2 WARNINGS REFER TO WARNINGS BELOW BEFORE OPERATING ELECTROMAGNET 1 Personnel Safety In operation the magnet fringing field is in excess of 0.5mT (5G). This can cause malfunctioning of heart pacemakers and other medical implants. We recommend that the fringing field should be mapped and warning signs be placed outside the 0.5mT (5G) contour.
Section 3 INSTALLATION Caution: This is a heavy system. All movement, lifting and installation of the 3470 Electromagnet must be under the supervision of an experienced person to prevent the possibility of serious injury or damage to the Electromagnet and associated equipment.
50°C +/3°C. Cooling The Model 3470 can be operated to an average coil temperature of 70°C. Assuming an ambient laboratory temperature of 20°C and a temperature coefficient of resistivity for copper of 0.0039/°C, the hot resistance of the coil should not exceed 20% more than the ambient temperature "cold"...
Page 10
At currents of approximately 3.50A and below the Model 3470 can be operated safely without water cooling. However the coil temperature will vary with the power dissipation. This results in dimensional changes of the magnet yoke and air cooling is not suitable when high field stability is required.
1. Adjust the poles to the desired gap with the poles approximately symmetrical about the center magnet line. 2. Adjust the cooling water flow to about 1 liters/min (0.26 USgpm) for the 3470. For operation at less than maximum power the water flow may be correspondingly reduced. Note that the inlet water temperature will determine the actual flow rate required.
The necessity to use calibration curves can be avoided by using a field controller to sense the magnetic field and provide a corresponding power supply control signal through the power supply programming inputs. Contact GMW for suitable instrumentation. Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Section 5 MAINTENANCE Periodically check that the poles are clean, properly lubricated and free of grit and dirt, which may cause binding. Be very careful not to damage the relatively soft pole surface since this may degrade the magnetic field uniformity in the gap. Note that the surface treatments used provide good corrosion protection but in order to maintain the inherent mechanical precision of the magnet, heavy build-up of plating materials is deliberately avoided.
Page 19
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Toomas Rett Serial No Pole Gap 10 mm Date 20 June, 1995 Hole Dia 4mm Magnet Current3.5 Amps US Army Redstone Arsenal C7916 Plot Y = 0.0 mm, Z = 0.0 mm...
Page 20
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Toomas Rett Serial No Pole Gap 10 mm Date 20 June, 1995 Hole Dia 4mm Magnet Current3.5 Amps US Army Redstone Arsenal C7916 Plot Y = 0.0 mm, Z = 0.0 mm...
Page 23
Coil position set to minimum gap Pole Gap: As per table below Pole Spacers: None Pole Face Gap in mm Filename: 3470 Gap-Field.xls Revised: March 13, 2000 Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 24
Serial No: None Notes: Coil position set to minimum gap Pole Gap: As per table below Pole Spacers: None Gap mm Current in Amps Filename: 3470 Ex 40-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 25
Serial No: None Notes: Coil position set to minimum gap Pole Gap: As per table below Pole Spacers: None Gap mm Current in Amps Filename: 3470 Ex 20-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 26
Serial No: None Notes: Coil position set to minimum gap Pole Gap: As per table below Pole Spacers: None Gap mm Current in Amps Filename: 3470 Ex 45-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 27
Pole Face: Position: X=0, Y=0, Z=0 Serial No: None Notes: Pole Gap: 10mm Pole Spacers: None Coil Position Current in Amps Filename: 3470 Ex Coil Position 40-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 28
Pole Face: Position: X=0, Y=0, Z=0 Serial No: None Notes: Pole Gap: 10mm Pole Spacers: None Coil Position Current in Amps Filename: 3470 Ex Coil Position 20-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 29
45 Square Position: X=0, Y=0, Z=0 Serial No: None Notes: Pole Gap: 10mm Pole Spacers: None Coil Position Current in Amps Filename: 3470 Ex Coil Position 45-05-40.xls Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com...
Page 32
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm Magnet Field...
Page 33
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm Magnet Field...
Page 34
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm Magnet Field...
Page 35
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 20 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Oct 4, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm Magnet Field...
Page 36
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 40 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm Magnet Field...
Page 37
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 40 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm Magnet Field...
Page 38
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 40 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm Magnet Field...
Page 39
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face 40 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Sept 25, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm Magnet Field...
Page 40
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face Square 45 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 22, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm...
Page 41
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face Square 45 mm Engr Greg Douglas Serial No 52 Pole Gap 10 mm Date Sept 22, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm...
Page 42
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face Square 45 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Sept 22, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, Z = 0.0 mm...
Page 43
GMW ASSOCIATES LABORATORY ELECTROMAGNET UNIFORMITY PLOT Model 3470 Pole Face Square 45 mm Engr Greg Douglas Serial No 52 Pole Gap 20 mm Date Sept 22, 1995 Magnet Current 5.0 Amps Plot Y = 0.0 mm, X = 0.0 mm...