1600KL CALIBRATION
The calibration is performed using kilos (kg) or pounds (lb), according to the units of measure used
upon entering into the calibration path.
1.
Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
2.
Press ENTER
3.
Using
keys and/or the keypad, set the
calibration load to at least 200 lbs (for best results
and greater accuracy, use a 600 lb weight). Press
ENTER
4.
Please clear the scale platform and press ENTER
5.
Please wait without touching or shaking the scale
until the zero calibration process is finished
6.
Load the scale platform with the required weight and
press ENTER
7.
Please wait without touching or shaking the scale
until the calibration process is finished
8.
Press ENTER
9.
Please remove weight from the scale platform
10. Please wait until the scale resumes normal operation
MAINTENANCE
The following pages provide instructions for maintenance, cleaning, troubleshooting and operator
replaceable parts for the ProPlus
this manual should be performed by qualified service personnel.
MAINTENANCE
Before first use and after periods of non-use, check the scale for proper operation and function. If
the scale does not operate correctly, refer to qualified service personnel.
1.
Check overall appearance of the total scale for any obvious damage, wear and tear.
2.
Inspect AC adapter for cord cracking or fraying or for broken or bent prongs.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation.
Disconnect the scale from the AC power source.
1.
Clean all external surfaces with a clean damp cloth or tissue. Mild antimicrobial or antibacterial
soap and water solution may be used. Dry with a clean soft cloth.
2.
Do not immerse the scale into cleaning or other liquid solution.
3.
Do not use Isopropyl Alcohol or other solutions to clean the display surface.
4.
Do not use abrasive cleaners.
Operator Action
®
1600KL. Maintenance operations other than those described in
Please clear the scale
Please clear the scale
9
Display
Calibration
Process
Enter load weight
600.0
Zero calibration
Zero calibration
Please wait
Weight calibration
Put: XXX.X
Weight calibration
Please wait
Calibration
factor: X.XXXXX
Calibration
Rebooting
Please wait
Health o meter
ProPlus