Non-Contact Forehead IR Thermometer User Manual Contents 1.General Description..............2.Safety Information..............3.Features..................4.Intended Use................5.Configuration................6.Indicator..................7.Description of Symbols.............. 8.Technical Specifications............. 9. Use..................10. Measuring Operation............... 11. Advice..................12. Maintenance and Cleaning............13. Troubleshooting..............14. This appliance conforms to the following standards....15.
Non-Contact Forehead IR Thermometer User Manual 1.General Description The device is intended for the intermittent determination of human body temperature in a no touch mode using the center of the forehead as the measurement site on people of all ages. The measured temperature correlates to axillary temperature.
Non-Contact Forehead IR Thermometer User Manual Importance: • Before taking of the temperature make sure to remove hair and perspiration from the forehead. • Selecting “Body” mode to measure the body temperature; • Use of this thermometer is not intended as a substitute for consultation with your physician.
Non-Contact Forehead IR Thermometer User Manual 4.Intended Use The device is intended for the intermittent determination of human body temperature in a no touch mode using the center of the forehead as the measurement site on people of all ages. The measured temperature correlates to axillary temperature.
Non-Contact Forehead IR Thermometer User Manual 7.Description of Symbols The device is in accordance with Medical Device Directive 93/42/EEC The device is in accordance with FCC Part 15 Subpart B:2007/Radio Frequency Devices IC Regulation ICEC-003: 2004 Interference-causing Equipment Standard-Digitial Apparatus 3V DC power supply, [2x1.5V AA (R6P/LR6)] Application part of Type B In order to protect the environment, please recycle the...
Non-Contact Forehead IR Thermometer User Manual 8.Technical Specifications Normal Conditions of Use Display Resolution 0.1°C (0.1°F) Operating Temperature 10 to 40°C (50 to 104°F) Storage Temperature -20 to 50°C (-4 to 122°F) Humidity Rate <85% DC 3V [2x1.5V AA (R6P/LR6)] Power Size 149x77x43mm/5.9x3x1.7in(LxWxH)
Non-Contact Forehead IR Thermometer User Manual 9. Use Selecting "Body" mode to measure the body temperature This mode displays the armpit-equivalent value of the measured forehead temperature. 9-1.Install battery 9-2.For the first use or when inserting new batteries, wait for 10 minutes for the warm-up of the apparatus.
Non-Contact Forehead IR Thermometer User Manual 10. Measuring Operation 10-1.Choosing The Temperature Unit – F1 Function Press “ MODE ” button for 2 seconds, the screen displays: F1. Select “ ” for degrees Celsius, “ ” for degrees Fahrenheit. 10-2.Alarme Setup – F2 Menu Press “...
Page 12
Non-Contact Forehead IR Thermometer User Manual 10-6.The Non-contact Body Infrared Thermometer is specially designed to take the body temperature of a human being. For this, use the Body mode. Measurement range for Body mode: 32 to 42.5°C (86 to 108°F) You can also use the Non-contact Body Infrared Thermometer to measure the temperature of an area or an object, a food, a liquid or a room temperature.
Non-Contact Forehead IR Thermometer User Manual 10-9.Longevity Use The Non-contact Body Infrared Thermometer was conceived for an intense and professional use, its longevity is guaranteed for 40000 takings. 11. Advice •The protective glass over the lens is the most important and fragile part of the thermometer, please take great care of it.
Non-Contact Forehead IR Thermometer User Manual 13. Troubleshooting If you happen to have one of the following problems while using your the non- ontact orehead IR hermometer please refer to this breakdown service guide to help resolve the problem. If the problem persists please contact our customer service.
Page 15
Non-Contact Forehead IR Thermometer User Manual This message displays in various cases – please find below a list of the main cases Reasons for “ ” or “ ” Advice message display Temperature reading hampered Make sure that there is no by hair, perspiration…...
Non-Contact Forehead IR Thermometer User Manual 14. This appliance conforms to the following standards: • ISO80601-2-56 and ASTM E1965-98 • EN 60601-1: Medical electrical equipment Part 1: General requirements for safety (IEC:60601-1) • EN60601-1-11: Medical electrical equipment-Part 1-11:General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment(IEC60601-1-11) 15.
Page 17
Table 201 Declaration - Electromagnetic Emissions Guidance and manufacturer’s declaration - electromagnetic emissions The DT-8806H is intended for use in the electromagnetic environment specified below. The customer or the user of the DT-8806H should assure that it is used in such an environment. Emissions test...
Page 18
Test Level Level Environment-Guidance Portable and mobile RF communications equipment should be used no closer to any part of the DT-8806H, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended...
Page 19
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the DT-8806H is used exceeds the applicable RF compliance level above, the Model Number or Type Number - by manufacturer should be observed to verify normal operation.
Page 20
Recommended separation distances between Portable and mobile RF communications equipment and the DT-8806H The DT-8806H is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the DT-8806H can help prevent electromagnetic interference...
Aethod(s) of sterilization or 70% isopropyl alcohol disinfection: Degree of safety of application DT-8806H is not suitable for use in the presence of a flammable in the presence of a flammable Anaesthetic mixture with ai or anaesthetic mixture with air or...
Page 22
Non-Contact Forehead IR Thermometer User Manual SHENZHEN EVERBEST MACHINERY INDUSTRY CO.,LTD 19 th Building, 5 th Region, Baiwangxin Industry Park, Songbai Road, Baimang, Xili, Nanshan, Shenzhen TEL: 0755-27353188 Fax: 0755-27653699 The manufacturer reserves the right to alter the specifications of the product without prior notification.
Page 23
Non-Contact Forehead IR Thermometer User Manual Doncument No:IFU-DT8806H-A1 Rev.090710...