SUPMEA SUP-C703S User Manual

SUPMEA SUP-C703S User Manual

Signal generator

Advertisement

Headquarters
5th floor,Building 4,Singapore Hangzhou Science Technology Park,
No. 6 street,Hangzhou Economic Development Area,
Hangzhou 310018,China
Singapore / Philippines
info@supmea.com
www.supmea.com
Supmea Automation Co.,Ltd.
User's Manual
Signal generator
U-S4.1-S4.2-MYEN1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SUP-C703S and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for SUPMEA SUP-C703S

  • Page 1 User's Manual Signal generator Headquarters 5th floor,Building 4,Singapore Hangzhou Science Technology Park, No. 6 street,Hangzhou Economic Development Area, Hangzhou 310018,China Singapore / Philippines info@supmea.com www.supmea.com U-S4.1-S4.2-MYEN1 Supmea Automation Co.,Ltd.
  • Page 2 Preface Thank you for purchasing Signal calibrator. Please read this manual carefully before operating and using it correctly to avoid unnecessary losses caused by false operation. Note  Modification of this manual’s contents will not be notified as a result of some factors, such as function upgrading.
  • Page 3 Safety Precautions In order to use this product safely, be sure to follow the safety precautions described. About this manual  Please submit this manual to the operator for reading.  Please read the operation manual carefully before applying the instrument. On the precondition of full understanding.
  • Page 4 due to violation of precautions shall not be borne by our company.  When installing lightning protection devices for this product and its control system, or designing and installing separate safety protection circuits for this product and its control system, it needs to be implemented by other devices.
  • Page 5 Warning:Pay special attention to the important information linked to product or particular part in the operation manual.  Confirm if the supply voltage is in consistent with the rated voltage before operation.  Don’t use the instrument in a flammable and combustible or steam area.
  • Page 6 If it is loose, please tighten it before use.  It is not allowed to disassemble, process, modify or repair the product without authorization, otherwise it may cause abnormal operation, electric shock or fire accident.  Wipe the product with a dry cotton cloth. Do not use alcohol, benzine or other organic solvents.
  • Page 7  Don’t use the instrument if it is found damaged or deformed at opening of package.  Prevent dust, wire end, iron fines or other objects from entering the instrument during installation, otherwise, it will cause abnormal movement or failure. ...
  • Page 8 Disclaimer  The company does not make any guarantees for the terms outside the scope of this product warranty.  This company is not responsible for damage to the instrument or loss of parts or unpredictable damage caused directly or indirectly by improper operation of the user.
  • Page 9: Table Of Contents

    Contents Chapter 1 Overview................ 1 1.1 Introduction..............1 1.2 Functions..............1 1.3 Parameter..............2 1.4 Specifications..............4 Chapter 2 Component and functions..........5 2.1 Terminal............... 6 2.2 Navigation button............6 2.3 LCD display..............8 Chapter 3 Signal output..............9 3.1 Voltage and Active Current........9 3.2 4-20 mA...............
  • Page 10 Contents Chapter 4 Signal Measurement...........14 4.1 Voltage and Active Current........14 4.2 Passive Current............15 4.3 Resistance/Thermal Resistance......17 4.4 Thermocouple............17 4.5 Regulating Valve............19 Chapter 5 Programmable Output..........20 5.1 Split Output Function (n/m)........20 5.2 Linear Output Function..........21 5.3 Automatic Stepping Function........22 Chapter 6 Troubleshooting and maintenance......
  • Page 12: Chapter 1 Overview

    Thermocouple: output and measurement of K, E, J, T, R, B, S and N  Resistance/thermal resistance: output and measurement of 0-400 Ω and -199°C-650°C (the minimum resistance output is 15Ω) * Note: Resistance/thermal resistance is only for SUP-C703S www.supmea.com...
  • Page 13: Parameter

    MΩ Output: 0.00- 24.00 ±0.2% 0.01 mA maximum load 750 Ω DC current Measurement: 4-20 4/8/12/16/20 Output ±0.2% 0.01 mA impedance 100 Ω Output: Passive 0.00- 24.00 ±0.2% 0.01 mA external power current supply 16-30 V - 2 - www.supmea.com...
  • Page 14 0-1300°C ±1% 1°C Note: The Output minimum ±0.2% 0.1 Ω 15.0-400.0Ω excitation Resistance Ω Measuremen current is 0.5 ±0.2% 0.1 Ω t 0.0-400.0 Ω mA and the maximum excitation Thermal -199.0°C- PT100 ±0.2% 0.1°C current is 3 resistance 650.0°C www.supmea.com...
  • Page 15: Specifications

    About 200 mA; with power supplied by 4*AAA batteries ( each nominal capacity of 1100 Power consumption mAh ), can be used for 4 hours with a full load and 17 hours standing by. Reverse connection and 30 V Over current protection - 4 - www.supmea.com...
  • Page 16: Chapter 2 Component And Functions

    Chapter 2 Component Names and Functions Chapter 2 Component and functions Figure 1 Front view of the signal calibrator www.supmea.com...
  • Page 17: Terminal

    Table 4 Button and switch description Description Button Move the decimal point/switch between positive ④ sign and negative sign Turn on or off the measurement function. ⑤ Display/change cold conjunction compensation ⑥ (only works for measuring the thermocouple). - 6 - www.supmea.com...
  • Page 18 16 V voltage to the transmitter to lower the power dissipation and prolong the service life. Switch the type of measurement/output signals. ⑩ Switch the measurement/output range. Turn on the programming function. Switch the programmable output wave form (linear output/stepping output)) Increase or decrease the value. www.supmea.com...
  • Page 19: Lcd Display

    RH is the upper limit. SL is the lower limit of the signal, while SH is the upper limit.  G. Battery level: the real-time battery level is displayed. Three bars are the full level. - 8 - www.supmea.com...
  • Page 20: Chapter 3 Signal Output

    Chapter 3 Signal output The signal calibrator can source voltage, active current, passive current, thermocouple signal, and PT100 signal. * Note: PT100 signal is available only for SUP-C703S 3.1 Voltage and Active Current  Connect the black signal cable to the public end and the yellow signal cable to the output end.
  • Page 21: Thermocouple

    3.3 Thermocouple Outputs a voltage equal to the thermocouple temperature minus cold conjunction compensation temperature.  Press the yellow signal button to switch the signal type to K/E/J/T/R/B/S/N.  Press to adjust the output temperature value. - 10 - www.supmea.com...
  • Page 22: Resistance/Thermal Resistance

    * Note: Resistance/Thermal Resistance is only for SUP-703 3.5 Passive Current  The passive current output can be used as a 2- wire transmitter simulator for loopback testing.  Press the yellow signal button to switch the signal type to mA. www.supmea.com -11-...
  • Page 23: Output Or Measurement Of Voltage And Current

    Valid when the signal type is voltage and current.  Press the yellow signal button to switch the display and the switching order is: range lower limit (RL)→range upper limit (RH)→signal lower limit (SL)→signal upper limit (SH)→none. - 12 - www.supmea.com...
  • Page 24 There is no unit display for display based on the range.  The actual measurement value or conversion value based on the range can be displayed in the measurement area. www.supmea.com -13-...
  • Page 25: Chapter 4 Signal Measurement

    Press the blue measurement button to enable the measurement function.  Press the blue signal button to switch the signal type.  The actual measurement value is displayed in the measurement value display area on the LCD. - 14 - www.supmea.com...
  • Page 26: Passive Current

    (16 V if the dial switch is on the ON side of low-load mode).  The actual measurement value is displayed in the measurement value display area on the LCD, and the output display area steadily displays 25 mA. www.supmea.com -15-...
  • Page 27 Chapter 4 Signal Measurement Figure 5 Measurement of 2-wire transmitter Figure 6 Measurement of 3-wire transmitter - 16 - www.supmea.com...
  • Page 28: Resistance/Thermal Resistance

    Press the blue signal button to switch the signal type to K/E/J/T/R/B/S/N.  The actual measurement value is shown in the measurement value display area on the LCD. If you need to check or adjust cold conjunction compensation temperature, do as follows: www.supmea.com -17-...
  • Page 29 If you flip the dial to the manual cold conjunction position, the LCD will display M. Now, you can manually set cold conjunction compensation value with  Press cold conjunction compensation button to close cold conjunction compensation display. - 18 - www.supmea.com...
  • Page 30: Regulating Valve

    Press the yellow source button and the value of on the screen will turn from and the output is started  The feedback measurement value of the valve is shown on the LCD measurement value display area. Figure 7 Regulating valve www.supmea.com -19-...
  • Page 31: Chapter 5 Programmable Output

    , and the value of on the screen will turn from and the output is started.  Press the yellow source button to stop the output.  Press the programming button to exit the split output mode. - 20 - www.supmea.com...
  • Page 32: Linear Output Function

    Press the yellow source button , and the value of on the screen will turn from and the output is started.  The LCD displays the current output step.  Press the yellow source button to stop the output. www.supmea.com -21-...
  • Page 33: Automatic Stepping Function

     Press to set the main set value for output.  Press the wave form button , and the LCD will display STEP and the stepping output mode is enabled. - 22 - www.supmea.com...
  • Page 34  The LCD displays the current output step.  Press the yellow source button to stop the output.  Press the programming button to exit the stepping output function. Figure 9 stepping output www.supmea.com -23-...
  • Page 35: Chapter 6 Troubleshooting And Maintenance

    Symptom: there is no measurement value display.  Check if the measurement function is enabled.  Symptom: the output value or measurement value is abnormal. a. Check if the signal line is correctly connected. b. Check if the signal type is correct. - 24 - www.supmea.com...
  • Page 36: Maintenance

     Avoid placing this instrument in an unstable position or where it may be affected by violent shocks.  Avoid placing this instrument outdoors or in a place with high moisture, dust, direct sunlight, or any heat sources. www.supmea.com -25-...