LI-COR 7200-101 Replacement Instructions

Blower motor
Hide thumbs Also See for 7200-101:

Advertisement

Quick Links

7200-101 Blower Motor Replacement
This installation guide describes how to replace the 7200-
101 Flow Module blower motor (part number 9972-039).
The procedure takes about 30 to 60 minutes.
Tools Required:
Hex keys: 1/16", 7/64", and 5/32" Wrenches: 3/16" and 1/4"
Disassembly
1
Power off the flow module and disconnect the power
cable.
2
Place the flow module on a level work surface.
Remove the adjust knob using a 1/16" hex key.
Remove the four cap screws. Save the screws for re-
3
use.
4
Lift the panel straight up. Disconnect the power plug
and set the panel aside.
5
Use the 1/4" wrench to remove the three screws that
secure the blower plate. Separate the blower plate
from the blower assembly. Save the screws for re-
use.
Installation Guide
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the 7200-101 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for LI-COR 7200-101

  • Page 1 7200-101 Blower Motor Replacement Installation Guide This installation guide describes how to replace the 7200- Lift the panel straight up. Disconnect the power plug and set the panel aside. 101 Flow Module blower motor (part number 9972-039). The procedure takes about 30 to 60 minutes.
  • Page 2 Use a 5/32” hex key to remove the motor clamp. Use Clean any residue from the thermal pad off of the the 3/16” wrench to remove the remaining standoff. motor mount and motor clamp. Save the cap screws and standoff for re-use.  Reassembly Remove the backing material from both sides of the thermal pad and position it in the motor mount, as...
  • Page 3 Reinstall the bottom standoff using 3/16” wrench. Plug in the motor cable to connector P4. Plug the power cable from the panel in to connector P9. Loosely install the motor clamp using the previously- removed cap screws. Carefully position the panel and install four cap screws (previously removed) to secure the panel.
  • Page 4 Global Headquarters Regional Offices LI-COR Biosciences LI-COR GmbH, Germany LI-COR Ltd., United Kingdom 4647 Superior Street Serving Andorra, Albania, Cyprus, Estonia, France, Germany, Serving Denmark, Finland, Ireland, Norway, Sweden, and UK. Lincoln, Nebraska 68504 LI-COR Biosciences UK Ltd. Iceland, Latvia, Lithuania, Liechtenstein, Malta, Moldova, Phone: +1-402-467-3576 St.