If a problem persists, call your supplier or Tamson Instruments b.v. WARRANTY Tamson Instruments b.v. warrants that all their manufactured equipment is free from defects in material and workmanship, preventing the machine from normal operation. Tamson Instruments b.v does not warranty that the equipment is fit for any other use than stated in this manual.
Conducting pollution must be prevented INSTALLATION 4.1 Important Tamson Instruments bv is not responsible for any consequential damage or harm caused by using this bath. Repairs on the electrical system of the bath may only be carried out by well trained and authorized persons.
304, 316, glass or brass and silicon sealing. 4.4 Drain bath fluid The TV12 can be emptied via the drain tap located at the backside of the apparatus. For safety reasons the tap can USE SCREWDRIVER only be opened when using a screwdriver.
1 to 2l/min. is sufficient. Instead of using tap water a Tamson cooling circulator can be connected to the TV12 to provide cooling. The TLC10-3 is our smallest circulator and well suited for this purpose.
INTRODUCTION The TAMSON model TV12 bath is designed to perform as ultra stable visability bath. The intende use is kinematic viscosity measurement and sensor calibration. 6.1 General The heat input is controlled by a microprocessor system. A special optimized electronic temperature measurement circuit ensures reproducibility of operation conditions.
Use well grounded mains. Before plugging the TV12 into the mains socket, make sure the voltage of the bath corresponds to the local voltage. The front panel layout shows the turn-push button: Next / increase: Turn right Previous / decrease:...
't Hoffstraat 12 NL28 INGB 0007 350 370 2665 JL Bleiswijk, The Netherlands NL95 RABO 0160100046 T. 31 (0) 10 522 43 73 Chamber of commerce 27 16 95 41 www.tamson.com tv12-man Rev 2.0 UK 1017 Page 22/29 info@tamson.com...
Page 23
The graph shows the temperature of a conventional bath(blue) and the temperature measured in the glass capilary of a viscometer(pink). The green curve shows the TV12 behaviour. Tamson Instruments bv ISO 9001 : 2015...
When aligning viscometers or sensors in horizontal position they will all have the same temperature. TV12 The homogeneity in the TV12 lies in the order of the delta T fluctuation ( ± 0.005°C). T No T...
High viscosity will activate the motor fuse. Electrical defect. Motor capacitor defective, replace capacitor or contact local dealer or Tamson instruments bv 10.3 Problems with set point Heater LED is not burning, motor is turning and temperature rises above set point.
The equipment conforms with all the specifications and norms in this regard. The equipment conforms without any further notice. Entity responsible for marking this declaration : Manufacturer, Tamson Instruments bv, van 't Hoffstraat 12, Bleiswijk The Netherlands, Name R.C. van Hall...