Page 2
TABLE OF CONTENTS Page Safety ..........................2 Product Information ....................2 Assembly ........................3 Unpacking ....................... 3 Installation ........................ 3 Operation ........................3 Controls and Indicator Lamp (See Figure 1) ..........3 Setting the Temperature ..................4 Fault Diagnosis ......................5 Technical Specifications ..................
Page 3
2. Product Information The Boekel Paraffin Wax Dispenser is designed to help meet the needs of today’s histology labs by having 5 liters of molten paraffin wax available when required by user. The unit has a set temperature range of 40.0⁰C to 75.0⁰C.
Page 4
3. Assembly Unpacking Remove packing materials carefully, and retain for future shipment or storage of the unit. Inspect for damage. Report all shipping damage to the carrier immediately. Shipping damage is covered by the carrier and repair/replacement for shipping damages must be coordinated through the carrier.
Page 5
FIGURE 1 FIGURE 2 Caution: Assure the reservoir is filled with a minimum of 2.5 cm of wax and a maximum of 2.5 cm below the top of the reservoir and the dispenser valve is positioned for “On Demand” dispensing before applying power. Use caution when filling the reservoir with wax to avoid overfilling the unit.
Page 6
Setting the Temperature 1. When setting the temperature to melt the Paraffin Wax, refer to Paraffin Wax manufacturers recommended melting temperature prior to adjusting the heater control knob. Generally, it will take approximately 2 hours to fully melt a reservoir full of Paraffin pellets. 2.
Page 7
Technical Specifications Operating Temperature Range 20⁰C to 30⁰C Settable Temperature Range 40.0⁰C to 75⁰C Temperature Accuracy ±2.0⁰C Reservoir Capacity 5.0 Liters Electrical Model # 145600 115VAC 50/60hz 250W Model # 145600-2 230VAC 50/60hz 250W Approvals Model # 145600 Certified to UL 61010-1 and CSA C22.2...
Page 8
Decontamination Certificate must be completed, signed by the user, and returned to Boekel Scientific prior to receiving the RMA number. Please be sure to mark the outside of the returned goods package with this RMA number to ensure prompt handling.
Need help?
Do you have a question about the 145600 and is the answer not in the manual?
Questions and answers