ThermoWorks WAND Blue Operating Instructions

ThermoWorks WAND Blue Operating Instructions

Non-contact, digital forehead thermometer bluetooth wireless technology

Advertisement

Quick Links

Manufacturer's declaration – electromagnetic immunity
WAND Blue is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of
WAND Blue should assure that it is used in such an environment.
IEC 60601
Compliance
Immunity test
Electromagnetic environment – guidance (for home healthcare environment)
test level
level
Recommended separation distance:
d = 1,2
d = 1,2
80MHz to 800 MHz
10 V/m
10 V/m
d = 2,3
800MHz to 2,7 GHz
80 MHz – 2,7
80 MHz – 2,7
Where P is the maximum output power rating of the transmitter in watts (W)
Radiated RF
GHz
GHz
according to the transmitter manufacturer and d is the recommended separation
IEC 61000-4-3
80 % AM at
80 % AM at 1
distance in metres (m).
1 kHz
kHz
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey,
a
should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with
the following symbol:
NOTE 1 - At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
WAND Blue is used exceeds the applicable RF compliance level above, WAND Blue should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating WAND Blue.
Recommended separation distances between portable and mobile RF communications equipment and WAND Blue
WAND Blue is intended for use in an electromagnetic environment (for home healthcare) in which radiated RF disturbances are controlled.
The customer or the user of WAND Blue can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and WAND Blue as recommended below, according to the maximum
output power of the communications equipment.
Separation distance according to frequency of transmitter - m
Rated maximum
output power of
150 kHz to 80 MHz
80 MHz to 800 MHz
transmitter- W
d = 1,2
d = 1,2
0,01
N/A
0,12
0,1
N/A
0,38
1
N/A
1,2
10
N/A
3,8
100
N/A
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 - At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 - These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired
operation.
This equipment complies with the FCC RF radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with a
minimum distance of 20 centimeters between the radiator and your body.
This device complies with Industry Canada Licence-exempt RSSs. Operation is subject to the following two conditions: (1) this device may not cause harmful
interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions
suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est
susceptible d'en compromettre le fonctionnement.
Contain FCC ID: SH6MDBT40 / Contain IC: 8017A-MDBT40
Manufacturer's declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
WAND Blue is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of
WAND Blue should assure that it is used in such an environment.
Test
Band
a)
frequency
Service
a)
(MHz)
(MHz)
385
380–390
TETRA 400
GMRS 460,
b
450
430–470
FRS 460
710
704–787
LTE Band 13,17
745
780
810
GSM 800/900,
TETRA 800,
870
800–960
iDEN 820,
CDMA 850,
930
LTE Band 5
GSM 1800;
1720
CDMA 1900;
1700–
GSM 1900;
1845
1990
DECT;
LTE Band 1, 3,
1970
4, 25; UMTS
Bluetooth,
WLAN,
2400–
2450
802.11 b/g/n,
2570
800 MHz to 2,7 GHz
RFID 2450,
d = 2,3
LTE Band 7
0,23
5240
5100–
WLAN 802.11
0,73
5500
5800
a/n
2,3
5785
7,3
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME
SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
23
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation,
it would be worst case.
This is not as AP or APG product.
There is no gender or age limitation for using
the infrared thermometer.
Special Notice
Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee
that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined
by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:\
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/ TV technician for help.
Compliance
Maximum
IMMUNITY TEST
LEVEL
Distance
Modulation
b)
power
LEVEL
(V/m)
(m)
(W)
(V/m)
(for home
healthcare)
Pulse
modulation b)
1,8
0,3
27
27
18 Hz
FM c)
±5 kHz deviation
2
0,3
28
28
1 kHz sine
Pulse
modulation b)
0,2
0,3
9
9
217 Hz
Pulse
modulation b)
2
0,3
28
28
18 Hz
Pulse
modulation b)
2
0,3
28
28
18 Hz
Pulse
modulation b)
2
0,3
28
28
217 Hz
Pulse
modulation b)
0,2
0,3
9
9
217 Hz
This device should not submerge into any liquids and expose it to direct moisture.
Choking from swallowing small parts and batteries by children or pets is possible, please keep small
parts and batteries at places where children and pets can't reach.
WAND Blue
Non-Contact,
Digital Forehead Thermometer
Bluetooth
Wireless Technology
®
THD2FE-B
Operating Instructions
For tips on using infrared thermometers visit the ThermoWorks Help Center at help.thermoworks.com.

Advertisement

loading

Summary of Contents for ThermoWorks WAND Blue

  • Page 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment WAND Blue is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of WAND Blue should assure that it is used in such an environment.
  • Page 2 Forehead Accuracy ±0.4°F (0.2°C) within 95 to 107.6°F (35 to 42°C) Otherwise ±0.5°F (0.3°C) WAND Blue is intended for use in the electromagnetic environment (for home healthcare) specified below. The customer or the user of Surface Accuracy ±0.5°F (±0.3°C) with 17.6 to 108°F (22 to 42.2°C) Otherwise ±4% or ±4°F (±2°C), whichever is greater...

This manual is also suitable for:

Thd2fe-b