Guardian Monitor™
Conventional Manifold/Alarm System
PRESSURE
PRESSURE
N
O
O
2
2
Line Pressure
Line Pressure
Use No Oil
Installation Instructions
Guardian Monitor™
M A N I F O L D
P.S.I.
PRESSURE
P.S.I.
PRESSURE

Advertisement

Table of Contents
loading

Summary of Contents for Accutron Guardian Monitor 38003 Guardian Monitor 38005

  • Page 1 Guardian Monitor™ Conventional Manifold/Alarm System Installation Instructions Guardian Monitor™ M A N I F O L D PRESSURE PRESSURE Line Pressure Line Pressure Use No Oil P.S.I. P.S.I. PRESSURE PRESSURE...
  • Page 2 GETTING STARTED To ensure proper installation and operation of your Guardian Monitor™ Conventional Manifold/Alarm System, please read these instructions thoroughly before beginning installation. PACKAGE CONTENTS Main Manifold with Cover 100 ft. cable Alarm Unit (Desk or Wall) Installation Screws Power Transformer Tank Restraints Pressure Regulators (O and N...
  • Page 3: Table Of Contents

    Table of Contents Page INSTALLATION SECTION Main Manifold ..............4 Tank Cylinder Restraints .
  • Page 4: Main Manifold

    INSTALLATION I. MAIN MANIFOLD 1. Take main manifold from the shipping carton and separate manifold cover from the main mani- fold block. 2. Remove the brass caps from both the nitrous oxide gas line (left) and the oxygen gas line (right) that are located in the tank room.
  • Page 5: Tank Cylinder Restraints

    II. TANK CYLINDER RESTRAINTS Center the tank restraint's mounting brackets directly beneath the main manifold and attach them to the wall 40 inches up from the floor. When attaching the tank restraint brackets, locate wall studs and secure the brackets to the wall using the 3 1/2-inch black screws. Note: When studs are not available to secure the brackets directly to the wall, it is recommended that a mounting board 48 to 60 inches wide, 4 inches tall and 1 inch thick be secured to the wall studs.
  • Page 6: Diss Hoses/Cables

    IV. DISS HOSES 1. Attach one end of the blue hose with DISS connector to the DISS fitting at the far left side of the bottom of the main manifold. Attach the other end of the hose to the fitting on the nitrous oxide regulator.
  • Page 7: Wall Mount Alarm - Models #38002, #38004 & # 34064

    8. Remove the red cap from the gray cable that was left coiled inside the receptacle box (in step 4). Connect the gray cable to the cable port on the back of the receptacle box faceplate. Push the cable connector into the port until there is a click and the cable connector cannot be pulled out of the port. 9.
  • Page 8: Power Supply

    9. Push the cable back into the receptacle box and carefully slip the circuit board inside the box. Fasten the aluminum sub plate to the outlet box using the enclosed four flat-head Phillips screws. 10. Position the faceplate over the aluminum sub plate and insert the flared portion of the two-part rivet through the faceplate and aluminum subplate.
  • Page 9: System Test

    IX. SYSTEM TEST (CONDUCTED BY INSTALLER PRIOR TO USE) 1. Turn system on by depressing the on/off button on the wall or desk alarm. 2. Press the test button - all four lights should blink and the audible alarm should sound. To turn off audible alarm press cancel button.
  • Page 10: Warranty

    XI. WARRANTY ACCUTRON 1-YEAR MANIFOLD LIMITED WARRANTY IF AN ACCUTRON MANIFOLD NEEDS TO HAVE REPAIR WORK OR REPLACEMENT PARTS DUR- ING THE 1-YEAR WARRANTY PERIOD DUE TO MANUFACTURING DEFECTS, ACCUTRON WILL PROVIDE THE PARTS AND LABOR AT NO CHARGE. THE MANIFOLD OWNER IS RESPONSIBLE ONLY FOR A $35.OO SHIPPING AND HANDLING FEE, WHICH WILL BE ASSESSED EACH TIME A MANIFOLD IS RETURNED TO ACCUTRON FOR WARRANTY WORK.
  • Page 11: Warranty & Returned Goods Policy

    XII. WARRANTY AND RETURNED GOODS POLICY All warranty resolution issues and merchandize returns will be handled through the local authorized Accutron distributor. Contact distributor where unit was purchased. XIII. REPAIR SERVICE POLICY All service issues will be handled through the local authorized Accutron distributor. Contact distributor where unit was purchased.
  • Page 12 1733 W. Parkside Lane • Phoenix, Arizona 85027 Toll-free: 800-531-2221 • Local Phone: 623-780-2020 • Fax: 623-780-0444 www.accutron-inc.com Part No. 25339 - Rev. 06 - 2/2010...

Table of Contents