Larson Electronics XLE-MD-TM-IR-R1 Instruction Manual

Medical grade infrared non-touch thermometer

Advertisement

Quick Links

XLE-MD-TM-IR-R1
Medical Grade Infrared Non-touch
Thermometer
Instruction Manual
Do not attempt use until you are familiar with
all warnings, precautions, and procedures
outlined within this instruction sheet.

Advertisement

Table of Contents
loading

Summary of Contents for Larson Electronics XLE-MD-TM-IR-R1

  • Page 1 XLE-MD-TM-IR-R1 Medical Grade Infrared Non-touch Thermometer Instruction Manual Do not attempt use until you are familiar with all warnings, precautions, and procedures outlined within this instruction sheet.
  • Page 2: General Description

    Allow device to acclimate to ambient temperature for 15 to 20 minutes before using. Temperature may also be taken behind the ear lobe. (See Part 10-5) Clean the device surface with a cotton pad lightly moistened with 70% Isopropyl alcohol. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 3: General Information

    'beep' sound. Press the SET button to select F4, and use the MEMO button switches between ℃ mode and ℉ mode. After setting, press the POW button to power off. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 4: Intended Use

    Temp °F 0-2 years 36.4 to 38.0 97.5 to 100.4 3-10 years 36.1 to 37.8 97.0 to 100.0 11-65 years 35.9 to 37.6 96.6 to 99.7 > 65 years 35.8 to 37.5 96.4 to 99.5 Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 5 XLE-MD-TM-IR-R1 5. Configuration 1. IR Sensor 2. LCD Display 3. Mode Selection 4. Memory Button 5. Power Button 6. Measurement Trigger 7. Battery Cover Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 6: Description Of Symbols

    0197 FDA Approved (2) 1.5v AAA Batteries Type B equipment In order to protect the environment, please recycle the battery according to the local regulations Body Indication of Mode Object Attention! Consult accompanying documents. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 7: Technical Specifications

    Measuring range (32℉~212℉)/(0℃~100℃) object mode Measuring distance 1~10cm (50℉~104℉)/(10℃~40℃) Operating ambient temperature Operating ambient humidity ≤80%RH Storage temperature range (-13℉~131℉)/(-25℃~55℃) DC3V Power Battery 2 AAA batteries IP rating IP20 116x55x170mm(LxWxH) Dimension Weight 150g (without battery) Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 8 Ensure there is no hair, perspiration, cosmetics or clothing covering the area. 3 to 5cm 1.18 to 1.96in Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 9: Measuring Operation

    F1 mode. Once in F1 mode, press the "SET" key again to enter F2 mode. In F2 mode, press "SET" again to enter F3 mode. Press "MEMO" to set temperature compensation. Press "POW" to exit. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 10 To obtain the internal temperature always use the “Body” mode. Please make sure to select “Body” mode for an internal temperature reading and “Object” mode for an external area reading, such as bottle, bath, room, etc. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com 10...
  • Page 11: Maintenance And Cleaning

    ● Measurement results are not a substitute for physician diagnosis. ● No special maintenance is required during use, please contact the seller or manufacturer if there is any malfunction appears. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...
  • Page 12: Frequently Asked Questions

    It is recommended to set the value to 0. Q3: The measurement temperature is obviously too high A3: Enter the parameter setting mode and check whether the temperature compensation setting is positive. It is recommended to set the value to 0. Larson Electronics, LLC P:(800) 369-6671 www.larsonelectronics.com...