9.1 The following replacement parts are available: ............. 13 10. Transport, packaging and storage..................14 10.1 Packaging: ........................14 10.2 Storage: ......................... 14 10.3 Transport: ........................14 11. Disposal of the Discharge Unit CD 2000 ................15 12. Feedback form........................16 ® Palas GmbH, Karlsruhe, Version V002022014...
It might also be necessary to wear protective clothes. Please pay attention to the correspondent standards and safety rules. Explanation of warning signs: This sign is placed near parts where dangerous high voltage is present. ® Palas GmbH, Karlsruhe, Version V002022014...
The following parts are attached to the system: 1 power cord 1 operating manual 1 electrotronical control unit 1 Discharge unit with a cable with three connectors: Negative Ion Positiv Ion Grounding ® Palas GmbH, Karlsruhe, Version V002022014...
For safety reasons, under no circumstances may the machine be commissioned if there is visible damage. In this event, please contact Palas® (Tel.: 0721/96 213-0). The discharge system CD 2000 has been permanently adjusted by the manufacturer to the mains voltage requested in the order.
AT LAST Deliver aerosol. Thus it is assured that aerosols do not enter the discharge chamber. 3.3 Switching-off Procedure Follow the switching on procedure in reverse order: Aerosol off, discharge off, mixing air off. ® Palas GmbH, Karlsruhe, Version V002022014...
(For example in the following tubes or in the upper part of a filter holder). An estimate for the needed time for the discharge process is around 0,5 seconds. Good discharge was found in applications assuring a time longer than 0,2 seconds. ® Palas GmbH, Karlsruhe, Version V002022014...
5.3 Adjusting the external potentiometers with scale: Relation between scales and voltage or current: voltage (Upos/Uneg): 1000 scales correspond to 6000 V 1 scale corresponds to current (Ipos/Ineg): 1000 scales correspond to 1000 µA 1 scale correspond 1 µA ® Palas GmbH, Karlsruhe, Version V002022014...
Page 9
Caution! In order to keep the values given by the EMC-directive, the current should be so low that no sparking takes place. This will be correct as long as the current is kept 20 A or less. ® Palas GmbH, Karlsruhe, Version V002022014...
Palas® employee. In the event of infringement, the customer loses their warranty claims; in addition Palas® is not liable for the damage arising. In case of problems with the discharge system CD 2000 please contact Palas® GmbH (Tel.: 0721/96 213-0).
The function of the unit can be controlled by measuring the corona current at I while operating. Using the standard settings, there should be shown approx. 0.2 V, which corresponds to approx. 20 µA. ® Palas GmbH, Karlsruhe, Version V002022014...
2 - 18 m³/h 3 - 36 m³/h i 6 mm i 13 mm hose coupling for hose: a 8 mm outlet: aerosol and charged mixing air connection: da = 16 mm di = 12 mm ® Palas GmbH, Karlsruhe, Version V002022014...
OPERATING MANUAL CD 2000 9. Spare parts list / accessories list 9.1 The following replacement parts are available: Aerosol inlet tube for CD 2000 ® Palas GmbH, Karlsruhe, Version V002022014...
10.1 Packaging: Please keep the original packaging, so that if it should become necessary (for service or maintenance) you can use this packaging to send the unit back to Palas®. 10.2 Storage: If the unit will not be used for a longer period, the instrument should be cleaned. Then the unit should be stored in a dry location.
OPERATING MANUAL CD 2000 11. Disposal of the Discharge Unit CD 2000 Once the discharge system CD 2000 has reached the end of its lifetime, please contact Palas® (0721-96213-0) for proper disposal. ® Palas GmbH, Karlsruhe, Version V002022014...
Need help?
Do you have a question about the CD 2000 Type A and is the answer not in the manual?
Questions and answers