Table of Contents APPLICABLE MODEL(S) ................................V REVISION HISTORY TABLE ................................. V GENERAL INFORMATION ................................1 Purpose of this Manual ..............................1 Intended Use ..................................1 Training Options ................................. 1 Symbols and Warnings ............................... 2 Safety Instructions ................................4 Electromagnetic Interference ............................5 Electromagnetic Compatibility (EMC) ..........................
Page 4
LIMITED WARRANTY ................................28 TROUBLESHOOTING GUIDE ..............................29 GUIDE TO INDICATOR LIGHTS AND SOUNDS ........................31 SPECIFICATIONS ..................................33 IN-SERVICE FORM ..................................35...
APPLICABLE MODEL(S) This manual is for the following product numbers and serial numbers: PRODUCT NUMBER DESCRIPTION EFFECTIVE S/N 840-003-22 840 Battery Powered Stretcher Bed, Mobile Base, 100-240V 200101-XXXXXX to Current REAR Steer (rear swivel casters, front LH steering caster) 840-003-23 840 Battery Powered Stretcher Bed, Mobile Base, 100-240V 200101-XXXXXX to Current FRONT Steer (front swivel casters, rear RH steering caster)
Training is accomplished either through reading the installation and operation sections of the user manual or through remote training provided by MTI Customer Service. To request training, contact MTI. PRODUCT INFORMATION Date of Purchase:...
User Manual | MTI 840 Symbols and Warnings Type B applied part This manual contains Notices, Cautions, and Warnings, which Type B, partie appliquée are meant to call attention to particular parts or sections Proper shipping orientation for the product of this manual.
Page 9
Upper and lower limits of relative humidity Steering Mode (green) Limites haute et basse de humidité relative Mode de Direction (vert) Free Mode (yellow) Upper and lower limits of atmospheric pressure Mode Libre (jaune) Limites haute et basse de la pression atmosphérique Locked Mode (red) Mode verrouillé...
Assurer l’appareil, patient, et d’autre objets sont dans un état sûr Backrest, seat, footrest, armrest, headrest, and other parts of the avant d’utiliser l’appareil. equipment should only be used for their intended purpose. MTI is not responsible for use of components outside of their intended WARNING/AVERTISSEMENT use.
California Proposition 65. par le fabricant, pour éviter tout risque de dégradation des Ce produit peut vous exposer à des agents chimiques, y compris performances de cet équipement.
Page 12
EQUIPMENT AND THE MTI STRETCHER BEDS The MTI Stretcher Bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MTI Stretcher Bed can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MTI Stretcher Bed as recommended below, according to the maximum output power of the communications equipment.
Page 13
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MTI Stretcher Bed is used exceeds the applicable RF compliance level above, the MTI Stretcher Bed should be observed to verify normal operation.
GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY The MTI Stretcher Bed is intended for use in the electromagnetic environment specified below. The customer or the user of the MTI Stretcher Bed should assure that it is used in such an environment.
Lors de l’utilisation de cathéters endocavitaires ou dispositifs NOTICE/AVIS chirurgicale / procédure haute fréquence, utiliser un matériau When the equipment is powered solely by AC power, the power non conducteur pour isoler le patient des parties métalliques de cord may be used to simultaneously isolate both poles of the l’équipement.
User Manual | MTI 840 EQUIPMENT FEATURES Description of 840 Mobile Battery Powered Stretcher Bed The 840 Stretcher Bed has power lift, back, tilt, foot, and foot extension functions. It has a 725 lb. (328.8 kg) patient capacity. The 840 Stretcher Bed is 25” (63.5 cm) wide with an entry height of 18 3/4” (47.6 cm). It features a mobile base and collapsible steering handles for convenient maneuverability.
After equipment is installed, verify programmable positions immediately, and contact MTI Customer Service. Request an (Home/P1-P4) are safe and known positions. inspection by the carrier if any concealed damage is noted.
User Manual | MTI 840 Unpacking and Set-up 1. Remove shipping straps, 2. Remove the shipping bolts 3. Slide the stretcher bed off the 4. Temporarily install the Hand shipping box, and other boxes and mounting straps with a pallet. To move equipment, lift Control in the arm spline following from the pallet.
Maximum surface temperature of applied part stretcher bed surfaces will not exceed 42°C. SmartTech™ SmartTech™ is the combination of MTI’s Smart Controller™ and Smart Safety™ technology using a centralized computer controller that enhances safety, reduces energy consumption, and provides smoother movement, additional memory options, and more user preferences.
User Manual | MTI 840 User Program Mode The User Program Mode allows users to set how the Home, Program and Swivel Buttons function. For user preference these button settings can be changed by entering the User Program Mode to make the setting changes described below.
Safety Lockout The Safety Lockout disables all commands to prevent unwanted or inadvertent movements. To enable the Safety Lockout, press the following sequence of buttons (within 1 second between presses): Stop, Stop, Back Up, Stop. To disable the Safety Lockout, press the following sequence of buttons: Stop, Stop, Back Down, Stop. Controls ITEM DESCRIPTION FUNCTION...
Page 22
User Manual | MTI 840 On/Off Button Turns the stretcher bed on and off (Conserves power and prevents accidental actuation of the stretcher bed). LED Indicators Used to provide user feedback on conditions, inputs, and errors. Left to right are green, blue, and orange LEDs.
Headrest Operation Articulating Headrests 1. To attach the headrest. 2. To adjust the height of the 3. To adjust the angle of the 4. To remove a headrest: headrest: headrest: A. Loosen the knob on the A. Loosen the knob on the articulating headrest lock A.
User Manual | MTI 840 Mobile Base Operation Total Lock Caster Steering Caster The 840 Stretcher Bed can be ordered in one of three different Steering Options. Each is indicated by a two digit number at the end of the stretcher bed’s item number. E.g. XXX-XXX-22. See the Serial Number Label to determine 1.
Page 25
MODE INDICATOR: Each locking bar has a mode indicator line on the center of the pedal that points to the adjacent indicator label on the base of the stretcher bed. Note that the (3) Total Lock Casters operate the same whether they are in FREE or STEER Mode.
3. Align the battery connector and charger connector and push The charging status of the battery is indicated by three LEDs on them together. The MTI Logo on the battery and label pull tab should face out (Figure 2). the front of the charger: 4.
The runtime of the battery will be reduced if it is operated below WARNING/AVERTISSEMENT room temperature. Only use MTI supplied batteries in the 840 Stretcher Bed. Calibration Utilisez uniquement les batteries fournies par MTI dans la civière Frequent partial discharges will reduce the accuracy of the fuel mobile 840.
However, the batteries do La batterie doit être chargée avec le chargeur de MTI. Si un contain recyclable materials and are accepted for recycling by a autre chargeur est utilisé, les dommages de la batterie ou autres...
LIST OF ACCESSORIES DESCRIPTION RESTRICTIONS Headrests Deep Oval, Horseshoe, Magnetic, Provides an articulating surface with various Ophthalmic, Prosthetic, and Rectangular pillow shapes/sizes to support patient’s Articulating Headrest head. Arms Arm, IV/Surgery, 18”Deluxe or Standard Provides an articulating surface for patient’s Requires Spline Assy, A Size (set of 2) 8-way Seat Rail Mtd arms.
Page 30
User Manual | MTI 840 LIST OF ACCESSORIES DESCRIPTION RESTRICTIONS Foot Mounted Steering Handles Provides (2) handles mounted at the foot of the stretcher bed for positioning or transport Forearm Support f/Deep Oval Headrest Assy Provides a forearm support during chinning Requires Deep Oval Headrest.
Page 31
LIST OF ACCESSORIES DESCRIPTION RESTRICTIONS Transfer Support/Bed Rail, Flip-up, Provides a support for entry and exit into Cannot be inserted while an arm is in the Ambidextrous the stretcher bed. Also used as a Bed Rail to Floating Arm Spline. secure a patient.
Policy” for more details on disinfecting solutions which have Contact with sharp or pointed objects can damage the been tested on MTI’s products. It is available for download at upholstery. Do not place hard surface objects on the upholstery www.mti.net/DisinfectantsPolicy. In the U.S. & Canada, call...
Equipment’s model and serial number information are required when calling for service. Serial number is indicated by the symbol located on 30 Coils the serial number tag. If service is required, contact MTI. U.S. & Canada (888) 520-4998 International (801) 875-4998 Fax (801) 952-0548 customer.service@mti.net www.mti.net Power Cord Removal Instructions 1.
This warranty does not apply to any part or product which upon examination by MTI or its authorized agents indicates the product has been repaired or altered in any way by persons other than MTI or its authorized agents, has been misused, abused or used in a manner contrary to any instructions issued by MTI.
Voir «GUIDE TO INDICATOR LIGHTS AND SOUNDS» à la page 31 pour des solutions supplémentaires. Pour plus d’informations sur cet équipement, y compris des schémas de circuits, les manuels de service et pièces, et autres descriptions, s’il vous plaît contacter MTI.
Page 36
Voir «GUIDE TO INDICATOR LIGHTS AND SOUNDS» à la page 31 pour des solutions supplémentaires. Pour plus d’informations sur cet équipement, y compris des schémas de circuits, les manuels de service et pièces, et autres descriptions, s’il vous plaît contacter MTI.
GUIDE TO INDICATOR LIGHTS AND SOUNDS LIGHTS SOUND STATUS OR CONDITION GREEN BLUE ORANGE Start Up Tune Stretcher bed has been plugged in. Solid None Stretcher bed is in hibernation state to conserve Slow Flash power. To exit hibernation press any button. None Stretcher bed is moving to the Programmed Home Rapid...
Page 38
User Manual | MTI 840 GUIDE TO INDICATOR LIGHTS AND SOUNDS LIGHTS SOUND STATUS OR CONDITION GREEN BLUE ORANGE Shut Down Tune Stretcher bed has been powered off. Press power button on hand control to resume operation. Low Beep Stop Button has been pressed for 5 seconds...
SPECIFICATIONS DESCRIPTION SPECIFICATION Load Capacity Lift 725 lb (328.9 kg) Back 400 lb (181.4 kg) Tilt 725 lb (328.9 kg) Foot (footrest fully extended) 150 lb (68 kg) Foot (footrest retracted) 200 lb (90.7 kg) Weight of stretcher bed 294 lb (133.4 kg) Shipping Weight of stretcher bed 361 lb (163.8 kg) Shipping Weight of Upholstery Kit...
Need help?
Do you have a question about the 840 Series and is the answer not in the manual?
Questions and answers