Breast Pump Instructions For Use
Model NB603
www.manamed.com
888.508.0712
IFUNB603 Rev A 10132020

Advertisement

Table of Contents
loading

Summary of Contents for Manamed Nature's Bond NB603

  • Page 1 Breast Pump Instructions For Use Model NB603 www.manamed.com 888.508.0712 IFUNB603 Rev A 10132020...
  • Page 2: Symbol List

    Symbol List Warning/Caution: See instructions for use Dangerous Voltage Electrical shock hazard. Disconnect LINE CORD before servicing; refer servicing to qualified service personnel. Date Of Manufacture Type BF Applied Part Keep Dry Serial Number Manufacture Lot Number Authorized Representative in the European Community Class II Equipment Use By Manufacturer...
  • Page 3: Specification

    Specification Dimensions For Main Unit: 185mm X 126mm X 57mm Flow Rate; No Filter; ≥7 L/min Under Maximum Pressure: Highest Negative Pressure ≤200mmHg At Stimulation Stage: Highest Negative Pressure ≤300mmHg At Expression Stage: Operating Cycle At Stimulation Stage: Approx.120 times/min Operating Cycle At Expression Stage: Approx.
  • Page 4 Always inspect power adaptor and batteries prior for use for damage and exposed wires. If damage is found, immediately discontinue use of power adaptor or batteries and contact ManaMed’s Customer Service. • Do not use outdoors or operate during transportation.
  • Page 5 Warnings (continued) WARNING: To avoid health risk and reduce the risk of injury: • Wash all parts that come in contact with your breast and breast milk after every use. • Inspect all appropriate pump components before each use. • Do NOT continue pumping for more than 2 consecutive pumping sessions if no results are achieved.
  • Page 6: Intended Use

    Intended Use Nature’s Bond breast pump is intended for lactating women to express and collect milk from their breasts to complement breast feeding. Description Nature’s Bond is a personal use electric breast pump that has two pumping modes and is capable of single and double pumping. Nature’s Bond is to be used when the baby would normally be breast feeding, which is usually 2 - 3 times a day.
  • Page 7 Cleaning Before using for the first time: • Clean and sanitize the following parts before using your pump for the first time. Use drinking quality tap or bottled water for cleaning. Take apart and wash all parts that come in contact with the breast and breast milk immediately after use to avoid dry up of milk residues and to prevent growth of bacteria.
  • Page 8: Device Operation

    Assembly of Pump Kit Wash hands before touching breast pump, kit and breasts, and avoid touching the inside of containers or lids. Place the breast pump on a clean dry surface. Securely attach breast shield to breast shield connector. Push valves onto bottom of breast shield connectors. Screw bottles onto breast shield connectors.
  • Page 9: Storing Breast Milk

    Storing Breast Milk Breast Milk Storage Guidelines Room Temp Cooler with Ice Pack Refrigerator Freezer Thawed Breast Milk 4–6 hours 24 hours 3–8 days 6–12 use within 24 hrs at 66–78 °F at 59 °F at 39 °F months (19–26 °C) (15 °C) or lower 0 –...
  • Page 10: Troubleshooting

    In the event the Nature’s Bond device requires repairs, it should be returned to the medical equipment company or to Manamed directly. Expected lifetime of the device is two years under normal use. Warranty Information ManaMed warrants its Nature’s Bond Breast...
  • Page 11 PARTICULAR PURPOSE AS TO ANY PRODUCT OR DEVICE, WHETHER OR NOT THAT PRODUCT OR DEVICE IS COVERED BY ANY EXPRESS WARRANTY CONTAINED HEREIN. IN NO EVENT SHALL MANAMED BE LIABLE FOR ANY SPECIAL, INCIDENTAL, CONSEQUENTIAL, OR INDIRECT DAMAGES (INCLUDING, WITHOUT LIMITATION, DAMAGES FOR LOSS OF PROFITS, USE OR TIME INCURRED BY PURCHASER OR END USER).
  • Page 12 Appendix 2 A. Instructions for use MODEL NB603 needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS; 2. Portable and mobile RF communications equipment can affect MODEL NB603.
  • Page 13 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity MODEL NB603 is intended for use in the electromagnetic environment specified below. The customer or the user of the MODEL NB603 should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment Guidance...
  • Page 14 Guidance and Manufacturer’s Declaration - Electromagnetic Immunity MODEL NB603 is intended for use in the electromagnetic environment specified below. The customer or the user of the MODEL NB603 should assure that it is used in such an environment. Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment...
  • Page 15 Recommended separation distances between portable and mobile RF communications equipment and the MODEL NB603 MODEL 603 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MODEL NB603 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the...
  • Page 16 888.508.0712...