NOVION®
Vacuum sensor
Instruction manual
Version 2019:1.1, Last update: 05.03.2020
Nichts ist unerreichbar.

Advertisement

Table of Contents
loading

Summary of Contents for Vacom NOVION

  • Page 1 NOVION® Vacuum sensor Instruction manual Version 2019:1.1, Last update: 05.03.2020 Nichts ist unerreichbar.
  • Page 3: Table Of Contents

    Establishing the data connection (cable) ..............19 3.3.6 Establishing the data connection (Bluetooth) ............... 20 3.4 Software Installation ......................21 ® Quick start (Software NOVION Viewer) ................. 22 Operation ..........................25 Operating modes ....................... 25 Measurement modes ......................25 ®...
  • Page 4 Table of Contents Troubleshooting ........................27 Error messages ........................27 Service / Maintenance ......................28 Operating information....................... 28 Calibration Mass spectrum ....................29 Disassembly .......................... 30 Uninstall ..........................30 Return to VACOM ......................31 NOVION Instruction manual...
  • Page 5: Safety

     Before using the device, consider possible contamination of the device by the process media used. Safety symbols used The safety and warning symbols explained below are used in this operating manual: Warning ! Danger of personal injury Attention Risk of damage to device or vacuum system NOVION Instruction manual...
  • Page 6: Warranty / Guarantee

    (Note: Section "Intended Use") VACOM does not assume any warranty if the user or third parties make changes to the device which goes beyond the work listed in this document. The return is at the expense of the customer in the original packaging.
  • Page 7: Ce - Declaration Of Conformity

    Safety and Security CE - Declaration of Conformity NOVION Instruction manual...
  • Page 8: Device Description

    NOVION® features a wide range of functions and intuitive software. Traditional vacuum sensors for total pressure measurement and residual gas analysis are replaced by a compact and cost-effective NOVION®.
  • Page 9: Technical Data

    ± 15 % of measured value (for p = mbar) 1 × 10 … 1 × 10 −8 −2 repeatability ± 10 % of measured value (for p = mbar) adjustable repetition rate up to 250ms Gas type dependence NOVION Instruction manual...
  • Page 10 (*depending on total pressure and mass number) gases 45 ... 300 amu Operating unit RJ45 socket: - digital In / Out: RS232/RS485 (VACOM protocol) interfaces - Analog Out: 0...10 V, max. 20mA Bluetooth 2.0 (VACOM protocol) power supply 24 VDC ±...
  • Page 11 Device description Software NOVION® Viewer Compatible Software (free download from vacom.de/downloads/software) Display of time curves of different gases and gas compositions Examination of all recorded spectra Analysis Possibilities Comparison of recorded spectra various export possibilities Helium leak detection Simple helium leak detection with acoustic signal...
  • Page 12: Comissioning

    Keep the packaging material, as warranty claims require that the unit be returned in the manufacturer's original packaging. Check the delivery for completeness. Check the components for visual damage. If damage is found, send a damage report to the carrier and inform the manufacturer. NOVION Instruction manual...
  • Page 13: Vacuum Connection

    Operation Vacuum connection To install the NOVION® on a vacuum system, the following steps must be executed: Warning ! Overpressure in the vacuum system can lead to injury from flying parts. Open the vacuum system only at atmospheric pressure of 1 bar.
  • Page 14 Do not switch on the control gear until it is connected to the sensor head, otherwise dangerous voltages may be accessible. Attention Do not switch on the control gear until it is connected to the sensor head, otherwise the operating unit may be damaged. NOVION Instruction manual...
  • Page 15: Electrical Connection

    Do not switch on the control gear until it is connected to the sensor head, otherwise dangerous voltages may be accessible. Attention The supplied RJ45 connection cable must not be connected to an Ethernet socket on the PC side. Incorrect connection and inadmissible voltages can damage the NOVION® operating unit. NOVION Instruction manual...
  • Page 16 3. connect the SUB-D9 connector to your PC (alternatively use RS232 to USB adapter - see accessories) 4. connect the provided power supply unit with a power cord 1. Connect the power cable to the mains supply. The device starts ... NOVION Instruction manual...
  • Page 17 The device goes through different operating states, which are displayed on the LEDs (power and status). The states are also displayed in the NOVION®-VIEWER software (if connected). The process of a normal start sequence is as follows:...
  • Page 18: Led Status Codes

    24 Volt DC voltage sources to guarantee the power supply. An adapter is required for this. The power supply socket (1) has the following wiring: Contact Parameters 23 Volt to 25 VDC 30 Watt Must be grounded NOVION Instruction manual...
  • Page 19: Establishing The Data Connection (Cable)

    An input / output connector in the form of an RJ45 socket is located on the rear of the NOVION® control gear for communication and data exchange with the device. The function of the contacts of this socket is explained in the following figure. The socket is connected to a PC via the supplied RJ45 connection cable and the supplied RJ45 / Sub-D9 adapter.
  • Page 20: Establishing The Data Connection (Bluetooth)

    The terminal device used must search for Bluetooth devices. As soon as the NOVION® operating device is visible in the list of found devices, pairing can take place. Afterwards, a "virtual" serial interface (COM port) is available. The remaining communication is identical to the wired serial connections.
  • Page 21: Software Installation

    3.4 Software Installation Reference If you have any questions or problems installing the software, contact your system administrator. Download the current version of the installation files of the NOVION® Viewer software from the VACOM homepage (https://www.vacom.de/downloads/software) System requirements: 2 GHz Processor...
  • Page 22: Quick Start (Software Novion Viewer)

    Quick start (Software NOVION Viewer) ® Start NOVION® Viewer via executable file or desktop shortcut to get to the start window of the NOVION® Viewer. Establishing a connection to the NOVION® operating unit Select COM port...
  • Page 23 Operation Make connection Connection established Connection has been established and the NOVION® operating unit is in the "Ready" state. Data is being received from the operating unit. NOVION Instruction manual...
  • Page 24 Comissioning Start data recording: Activate the desired measurement mode via the corresponding button. Initialize data recording via the "Start measurement" button. Change measuring mode Start measurement NOVION Instruction manual...
  • Page 25: Operation

    For pressures below mbar the ionization mode is used. The total pressure is always measured by the NOVION® even if the device is in a different measuring mode. The helium leak detection mode can be activated in addition to the total Helium Leak pressure measurement.
  • Page 26: Operating The Novion Viewer

    Operating the NOVION Viewer ® The free "NOVION Viewer®" software allows easy and intuitive access to your NOVION®. All important functions of the NOVION® are graphically displayed. Generated data can be saved and exported. Further information about the NOVION Viewer® software can be found in the NOVION®...
  • Page 27: Troubleshooting

    LED lights up permanently, green LED is off in software: Solution: 1. First, make sure if both filaments are burnt out: In NOVION®-Viewer Software: Go to SETTINGS -> Advanced SETTINGS If both filaments are OK Contact support If exactly one filament burnt out and the other is intact...
  • Page 28: Service / Maintenance

    Operating information Degas of filaments VACOM recommends carrying out a regular DEGAS of the active filament. In particular, a DEGAS can reduce the outgassing rate of the sensor and increase the life of the filament! The recommended DEGAS time is 10 minutes.
  • Page 29: Calibration Mass Spectrum

    Service / Maintenance Calibration Mass spectrum It is possible to calibrate the mass resolution of NOVION® quickly and easily. For this purpose you have to let a known gas composition into the vacuum chamber and start the supplied calibration tool: (Download at: https://www.vacom.de/downloads/software, is installed together with the...
  • Page 30: Disassembly

    1. end all processes running in the vacuum system 2. ventilate the vacuum system 3. take the device out of operation 4. remove electrical cables / connections from the device 5. open the vacuum connection and disconnect the device from the vacuum system NOVION Instruction manual...
  • Page 31: Return To Vacom

    If the device has been used in atmospheres containing substances hazardous to health or the environment, these substances must be declared to the disposal company. If the sensor is returned to VACOM® Vakuum Komponenten & Messtechnik GmbH, a legally binding declaration on the contamination of the product must be made to be completed and signed by a competent person.