The Q-tag
®
Quad
The Q-tag
Quad is a temperature monitoring
®
indicator which measures reliably and precisely
environmental temperature in which it is situated.
It will compute up to 4 individual alarms
according to the programmed conditions.
Triggered alarms will be displayed on the LCD
with simple and easy to understand symbols.

Alarm explanation

ALM 1
 Single-event alarm "high"
ALM 2
 Cumulative alarm "high"
ALM 3
 Cumulative alarm "low" or "high"
ALM 4
 Single-event alarm "low"
Operating instructions
1) State of delivery / Sleep Mode
20160927_Release_8.2
Devices with a shorter expiry date than 2021-07
Q-tag
®
Quad is shipped in its so-called "Sleep
Mode". This will be observed by the constantly
flashing dot (running indicator) on the display.
In sleep mode, no data will be recorded.
Devices with a longer expiry date than 2021-08
Q-tag
®
Quad is shipped in its so-called "Sleep
Mode". This will be observed by the constantly
flashing dot (running indicator).
temperature excursion will already be determined
before the device has been activated (sleep mode),
the display will indicate the corresponding, flashing
alarm flag. Once the temperature returns to its
normal range, the flashing alarm flag on the display
disappears.
In sleep mode, no data will be recorded.
page 2 of 26