GEX Corporation Thermo Scientific GENESYS 30 Product Specifications And Usage Information

Advertisement

DESCRIPTION
The P4400 Thermo Scientific GENESYS 30 Spectrophotometer (G30) is a device used to measure the optical absorbance of
transparent samples in the spectrum of visible light. It is designed to be simple and fast with the highest performance and
quality.
The spectrophotometer is built for the fast-paced, high-volume processing environment, and can withstand the rigors of daily
usage for many years.
APPLICATION
The G30 is used for the measurement of optical absorbance of dosimeters using a select wavelength in the visible spectrum,
such as GEX B3 radiochromic film dosimeters, and FWT-60 radiochromic film dosimeters. The instrument is combined with
specially designed dosimeter holders which position the dosimeters appropriately to obtain accurate and repeatable
measurements.
SPECIFICATIONS
Physical Specifications:
GEX Part
Description
No.
P4400-US
Spectrophotometer – Type B plug
P4400-EU
Spectrophotometer – Type F Plug
P4400-UK
Spectrophotometer – Type G Plug
**For a complete list of specifications see the brochures and information from Thermo Scientific.
Material
Color
Printing
Electrical Supply
Optical Design
Spectral Bandwidth(s)
Light Source
Detector
Wavelength
Photometric
Stray Light
Baseline Flatness
Keypad
Display
100-167 Rev. B
Product Dimensions
34.3 cm L x 38.1 cm W x 20.3
(13.5" L x 15" W x 8" H)
Molded plastic
Grey and White
Thermo Scientific logo plate on front
100–240 V, 50–60 Hz, selected automatically; 150 W maximum
Single-beam
5 nm
Tungsten-Halogen lamp
Silicon Photodiode
Range:
325 –1100 nm
Accuracy:
±2 nm (full range 325 to 1100 nm)
Repeatability:
≤ 1 nm
Range:
Accuracy – Instrument:
Repeatability:
Drift (Stability)
<0.1 %T at 340nm and 400nm
<0.003 A
Tactile rubber 23 keys with numeric keypad
Touchscreen LCD panel; 800 x 480; 12.7 cm (5 in) diagonal
Release Date: 10/29/18
Product Specifications and Usage Information
Packaging Dimensions
58.4 cm L x 50.8 cm W x 33.0
cm H
(23" L x 20" W x 13" H)
-3A to +3A
±0.002 A (0 – 0.3A)
0.5% of ABS reading (0.301A – 2.5A)
±0.002 A
<0.002 A/hr
GENESYS 30 Spectrophotometer
Product
Weight
10.0 kg
cm H
(22 lbs.)
Page 1 of 9
P4400

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Thermo Scientific GENESYS 30 and is the answer not in the manual?

Questions and answers

Summary of Contents for GEX Corporation Thermo Scientific GENESYS 30

  • Page 1 DESCRIPTION The P4400 Thermo Scientific GENESYS 30 Spectrophotometer (G30) is a device used to measure the optical absorbance of transparent samples in the spectrum of visible light. It is designed to be simple and fast with the highest performance and quality.
  • Page 2: Product Photo

    Store the instrument in a cool and dry location. INSTALLATION Please refer to the Thermo Scientific GENESYS 30 Spectrophotometer User Guide for explicit detail. Information supplied herein should be considered supplemental or redundant and is specific to usage for dosimetry in radiation processing applications.
  • Page 3 Product Specifications and Usage Information P4400 GENESYS 30 Spectrophotometer Instrument performance should be verified when operating outside of these ranges. GEX has verified instrument performance at an altitude of approximately 1610m. The instrument should be positioned on a surface that is level and free of vibration. Consider all GLP’s (Good Laboratory Practices) concerning cleanliness of the area.
  • Page 4 The instrument can be used when disconnected from a PC to make measurements. For most users, this will never be a necessity, but some users may find it useful despite the limitations from using the instrument in this manner. For more information on making fixed or scanning measurements, see the Thermo Scientific GENESYS 30 User Guide. Fixed Wavelength Measurements:...
  • Page 5 Product Specifications and Usage Information P4400 GENESYS 30 Spectrophotometer FIGURE 4 Ensure the dosimeter holder is empty, then zero the instrument by pressing the yellow “0.00” button. After the instrument is zeroed, insert a dosimeter to be measured and select the green diamond-shaped button. The measurement will be captured on the screen.
  • Page 6 QUALIFICATION The information in this protocol is provided for reference only. GEX Corporation shall not be liable for errors contained herein nor incidental or consequential damages in connection with the provision, performance, or use of this protocol. The information in this protocol is subject to change without notice. Users are ultimately responsible for validation of their systems.
  • Page 7: Care And Maintenance

    Product Specifications and Usage Information P4400 GENESYS 30 Spectrophotometer Notes: • GEX has completed the necessary testing on the instrument (described above) as part of the DoseControl System validation. The validation data is available for audit. • It is suggested to conduct thirty (30) repeated measurements of single dosimeters irradiated at low, middle, and high doses in the user’s range, and to reproduce this test for a minimum of three (3) days using 3 or more operators.
  • Page 8 Product Specifications and Usage Information P4400 GENESYS 30 Spectrophotometer ACCESSORIES GEX Part No. Description Purpose P4220 Spectronic Standards Set Performance Verification G30 Tungsten-Halogen Lamp P4401 Lamp replacement part Assembly Dosimeter holder system baseplate Platform for interchangeable dosimeter holders allowing the sample P4405 with beam tubes compartment lid to remain open during measurement.
  • Page 9: Revision History

    WARRANTY/GUARANTEE Warranty: Please refer to the Thermo Scientific GENESYS 30 Warranty Information document for details. Guarantee: 1-year GEX satisfaction guarantee. Product may be returned to GEX within one year from the date of delivery for any customer dissatisfaction.

This manual is also suitable for:

P4400-usP4400-euP4400-uk

Table of Contents