User Manual
Resistivity Meter
R12-800SC
TROUBLESHOOTING INSTRUMENT
THE SYMPTOMS
POSSIBLE CAUSE
1. Low Current Inject
UNREADABLE
POTENTIAL
2. Low Power Battery Input
VOLTAGE INPUT
3. Bad Connection Cable from
terminal or elekrtodes
ALARM OVER CURRENT
High Current Inject
ACTIVATED
ALARM OVER VOLTAGE
Out of Range Setting Measurement Chek Manual Range Setting
ACTIVATED
NO OUTPUT POWER
Converter Module Fail
INSTRUMENT
MCB DC POWER TRIP
Short Circuit Connection
December, 2019
2019 Gatrik Innotech Indonesia, PT. All rights reserved. Specifications are subject to change without notice.
C
All product names are trademarks of their respective companies.
PROBLEM SOLVING
1. Switch On some additional
converters
2. Change Battery Connection /
Replace New Battery
3. Check cable & Electrodes
connection
Switch Off some converters
1.
If one of the converter switches
is pressed, but voltage output
display is not increase, the
converter module is fails
2.
The measurement process can
still be done, but the maximum
output voltage (800V) cannot be
achieved
Please Contact Authorized Dealer
for Repair the Instrument