Advertisement

Blood Pressure
Monitor
NS-BPMW1
Before using your new product, please read these instructions to prevent any damage.
California Residents
WARNING: Reproductive harm -
www.p65warnings.ca.gov
USER GUIDE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NS-BPMW1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Insignia NS-BPMW1

  • Page 1 USER GUIDE Blood Pressure Monitor NS-BPMW1 California Residents WARNING: Reproductive harm - www.p65warnings.ca.gov Before using your new product, please read these instructions to prevent any damage.
  • Page 2: Table Of Contents

    Contents Introduction ........3 Indications for use .
  • Page 3: Introduction

    Blood Pressure Monitor Introduction Congratulations on the purchase of your new Insignia Blood Pressure Monitor. Your blood pressure monitor features the latest in design and capabilities and will provide reliable and trouble-free performance.  Indications for use This blood pressure monitor is a digital monitor intended for use in measuring blood pressure and heartbeat rate on persons with an arm circumference ranging from approximately 8.75”...
  • Page 4 NS-BPMW1 • The device is not intended for patient transport outside a healthcare facility. • The device is not intended for public use. • This device is intended for non-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the arm or for functions other than obtaining a blood pressure measurement.
  • Page 5 Blood Pressure Monitor • On the rare occasion of a fault causing the cuff to remain fully inflated during measurement, open the cuff immediately. Prolonged high pressure (cuff pressure >300 mm Hg or constant pressure >15 mm Hg for more than three minutes) applied to the arm may lead to an ecchymosis.
  • Page 6 NS-BPMW1 • The patient can measure, transmit data, and change batteries under normal circumstances and maintain the device and its accessories according to the accompanying documents. • To avoid measurement errors, please avoid strong electromagnetic fields, radiated interference signals, or electrical fast transient/burst signals.
  • Page 7 The device must only be serviced, repaired, and opened by individuals at authorized sales/service centers. • Please report to Insignia Customer Service if any unexpected operation or events occur. • Keep the unit out of reach of infants, young...
  • Page 8 NS-BPMW1 • There are no luer lock connectors used in the construction of tubing. There is a possibility that they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped into a blood vessel. • Please use the device under the environment which was provided in the accompanying documents.
  • Page 9 Blood Pressure Monitor Symbol for “ENVIRONMENT PROTECTION - Electrical waste products should not be disposed of with household waste. Please recycle where facilities exist. Check with your local authority or retailer for recycling advice.” Symbol for “RECYCLE.” www.insigniaproducts.com...
  • Page 10: Features

    NS-BPMW1 Features • Blood Pressure Monitor • Cuff • AAA batteries (4) • User Guide Front Cuff Air hose Air connector plug Swith user button DC power jack Memory button Set button Start/Stop button LCD display Back Battery compartment www.insigniaproducts.com...
  • Page 11: Display

    Blood Pressure Monitor Display ITEM DESCRIPTION Shows that the monitor is in the memory mode and Memory which group of memory it is. Recording measurements for User A. User A Recording measurements for User B. User B Average The average value of the latest three records. value Batteries are low and need to be replaced.
  • Page 12: Learning About Blood Pressure

    NS-BPMW1 ITEM DESCRIPTION Inflation/ The cuff is inflating or deflating. Deflation Shows the date (year, month, day) and time of Current measurements. time Blood Indicates the blood pressure level. pressure level Systolic Shows your heart’s systolic pressure. blood pressure Blood pressure is displayed in kilopascals (kPa).
  • Page 13 Blood Pressure Monitor What is the standard blood pressure classification? The chart on the right is the standard blood pressure classification published by American Heart Association (AHA): SYSTOLIC DIASTOLIC BLOOD PRESSURE MMHG MMHG CATEGORY (UPPER #) (LOWER #) Normal Less than 120 Less than 80 Elevated 120–129...
  • Page 14 NS-BPMW1 • You have four or more pulse intervals and the difference between each interval and the average is more than the average value of ±15% CAUTION: The appearance of the irregular heartbeat symbol indicates that a pulse irregularity consistent with an irregular heartbeat was detected during measurement.
  • Page 15: Setting Up Your Blood Pressure Monitor

    Blood Pressure Monitor Setting up your blood pressure monitor Powering your blood pressure monitor Your blood pressure monitor can be powered with either batteries (included) or the AC adapter (not included). CAUTION: For the best results and to protect your monitor, use the the correct batteries or a power adapter (not included).
  • Page 16: Attaching The Cuff To The Monitor

    NS-BPMW1 Connecting an AC adapter (not included) 1 Make sure that your adapter meets the following requirements: • Input: AC 100–240 V, 50/60 Hz, 0.2 A max Output: 6 V, 1 A 2 Connect the AC adapter from the monitor’s DC power jack to a power outlet.
  • Page 17 Blood Pressure Monitor 3 Press MEM repeatedly to change the year, then press SET to save your selection. Each press increases the year by one, cycling from 2018–2058. 4 When you get the right year, press SET to save the year. The display shows the date (month/day) settings.
  • Page 18 NS-BPMW1 8 Press MEM repeatedly to change the hour, then press SET to save your selection. Each press increases the hour by one, cycling from 1–12 (if you selected 12h) or 0–23 (if you selected 24h). 9 Press MEM repeatedly to change the minutes, then press SET to save your selection.
  • Page 19: Using Your Blood Pressure Monitor

    Blood Pressure Monitor Using your blood pressure monitor Getting accurate measurements Follow these tips for the most accurate readings (especially for people with hypertension): • Rest for five minutes before the first measurement. Relax as much as possible and do not move and talk during the measurement. •...
  • Page 20 NS-BPMW1 4 Position the tube off-center toward the inner side of your arm (in-line with the little finger). Position the artery mark ( ) over the main artery on the inside of your arm. Note: To locate the main artery, press two fingers approximately .8 in.
  • Page 21: Taking Your Measurements

    Blood Pressure Monitor Taking your measurements WARNING: • This monitor is not suitable for measuring children. • If you need to stop the cuff from inflating, press START/STOP to immediately release the air. Loosen the cuff and remove it from your arm.
  • Page 22 NS-BPMW1 1 When the monitor is off, press the START/STOP to turn on the monitor. The monitor takes your measurements and saves the data for the selected user. Note: You can select user A or B by pressing the user button during measurement or when in the memory mode.
  • Page 23: Recalling Records

    Blood Pressure Monitor Recalling records 1 When the monitor is off, press MEM. The monitor cycles through the last three records. If there are less than three records, it will show the latest record instead. Note: The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record.
  • Page 24: Deleting All Records

    NS-BPMW1 Deleting all records If you did not get the correct measurement, you can delete all records for the selected user. You cannot delete a single record. 1 When the monitor is off, press MEM to enter memory recall mode.
  • Page 25: Maintaining Your Blood Pressure Monitor

    Blood Pressure Monitor 4 When all records are deleted, the screen shows “- - -. ” Press MEM to return to the correct screen. Maintaining your blood pressure monitor In order to get the best performance: • Store in a dry place and avoid the sunshine. •...
  • Page 26 NS-BPMW1 PROBLEM POSSIBLE SOLUTIONS The display says “out.” • Your results are out of measurement range. Take a moment to relax. Refasten the cuff and measure again. If the problem persists, contact your physician. Error messages ERROR CODE POSSIBLE PROBLEM/SOLUTION “E 01”...
  • Page 27: Specifications

    Blood Pressure Monitor Specifications Dimensions Blood pressure monitor: 2.3 × 5.1 × 5.5 in. (5.9 × 13 × (H × W × D) 14 cm) Weight .6 lbs (260 g) Power supply Batteries: 6VDC 4×AAA batteries AC adapter: Input: AC 100–240 V, 50/60 Hz, 0.2 A Max Output: 6 V, 1 A Display mode Digital LCD...
  • Page 28: Legal Notices

    NS-BPMW1 Mode of Continuous operation operation Degree of Type BF applied part protection Protection IP21 against ingress (The device is protected against solid foreign objects of of water .5 in. (12.5 mm) and greater and is protected against vertically falling water drops.)
  • Page 29 Blood Pressure Monitor General • EN 60601-1:2006+A1:2013/ IEC requirements 60601-1:2005+A1:2012 Medical electrical for safety equipment - Part 1: General requirements for basic safety and essential performance • EN 60601-1-11:2015/ IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment...
  • Page 30 NS-BPMW1 Usability • EN 60601-1-6:2010+A1:2015/IEC 60601-1-6:2010+A1:2013 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability • IEC 62366-1:2015 Medical devices - Part 1: Application of usability engineering to medical devices Software •...
  • Page 31 12 in. (30 cm) to any part of the equipment NS-BPMW1, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
  • Page 32 NS-BPMW1 Table 2 GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY IEC 60601-1-2 IMMUNITY TEST COMPLIANCE LEVEL TEST LEVEL Electrostatic discharge ±8 kV contact ±8 kV contact (ESD) ±2 kV, ±4kV, ±8 kV, ±15 ±2 kV, ±4kV, ±8 kV, IEC 61000-4-2 kV air ±15 kV air...
  • Page 33 Blood Pressure Monitor GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC IMMUNITY IEC 60601-1-2 IMMUNITY TEST COMPLIANCE LEVEL TEST LEVEL Conduced RF IEC61000-4-6 0.15 MHz – 80 MHz 0.15 MHz – 80 MHz 6 V in ISM and amateur 6 V in ISM and amateur radio bands between radio bands between 0.15 MHz and 80 MHz...
  • Page 34 NS-BPMW1 Table 3 GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY Radiated RF IEC61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment). IMMU- TEST MODU- DIS- NITY FRE- BAND SERVICE MODULATION LATION TANCE TEST QUENCY (MHZ) LEVEL (MHZ) (V/M) 380–...
  • Page 35 Blood Pressure Monitor GUIDANCE AND MANUFACTURER’S DECLARATION - ELECTROMAGNETIC IMMUNITY Radiated RF IEC61000-4-3 (Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment). IMMU- TEST MODU- DIS- NITY FRE- BAND SERVICE MODULATION LATION TANCE TEST QUENCY (MHZ) LEVEL (MHZ) (V/M) 2450 2400–...
  • Page 36 NS-BPMW1 FCC Caution Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate this equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.
  • Page 37: One-Year Limited Warranty

    Product is determined to be defective by an authorized Insignia repair center or store personnel, Insignia will (at its sole option): (1) repair the Product with new or rebuilt parts; or (2) replace the Product at no charge with new or rebuilt comparable products or parts.
  • Page 38 • Damage due to incorrect operation or maintenance • Connection to an incorrect voltage or power supply • Attempted repair by any person not authorized by Insignia to service the Product • Products sold “as is” or “with all faults”...
  • Page 39 PROVINCE TO PROVINCE. Contact Insignia: 1-877-467-4289 www.insigniaproducts.com INSIGNIA is a trademark of Best Buy and its affiliated companies. *Distributed by Best Buy Purchasing, LLC 7601 Penn Ave South, Richfield, MN 55423 U.S.A. ©2020 Best Buy. All rights reserved. Guangdong Transtek Medical Electronics Co., Ltd.
  • Page 40 1-877-467-4289 (U.S. and Canada) or 01-800-926-3000 (Mexico) INSIGNIA is a trademark of Best Buy and its affiliated companies. Distributed by Best Buy Purchasing, LLC 7601 Penn Ave South, Richfield, MN 55423 U.S.A. V1 ENGLISH ©2020 Best Buy. All rights reserved.

Table of Contents

Save PDF