• This manual reflects the device specifications and appli‑ cable safety standards valid at the time of printing. All rights are reserved for devices, circuits, techniques, software programs, and names appearing in this manual. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Attention is drawn to the fact that local laws take priority over the above mentioned requirements. If in doubt, please consult your local dealer or ergoline GmbH. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
If you have questions in this matter, please contact ergoline GmbH or an ergoline representative. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
NRTL label for the USA and Canada. Do not lean against device: tipping hazard. Manufacturer’s identification. Date of manufacture. The number found under this symbol is the date of manufacture in the YYYY‑MM‑DD format. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales office: Via F.lli Bronzetti, 8...
Figure 4 – 2: Tightening the clamping lever NOT with maximum force. • Lubricate the threads of the clamping levers every three months at minimum with a suitable lubricant (e.g., OKS470). User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
It is recommended to install the terminal with the display and control keys towards the operator and the speed read‑ out towards the patient. Figure 5 – 2: Different orientations of the control terminal User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Figure 5 – 4: Leveling foot of the xr50+ ergometer damage by the feet. The xr50+ has 2 castors at the front for transport. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Removing the power cord results in a complete disconnection of the device from the power supply (all poles). Ensure that the power plug is readily accessible at all times. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
• Connection Cables • Use only connection cables approved by ergoline. A special PC driver software, which can be obtained from ergoline, is required for operation via the USB port. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
If this option is selected, the start screen of the selected operating mode (e.g., Ergometry) will be displayed instead of the main menu. With the key, you can display the main menu. Figure 6 – 3: Main menu User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales office: Via F.lli Bronzetti, 8...
(= patient should pedal slower) Note • If, during an exercise test, the speed drops below 30 RPM, the load readout starts blinking on the display. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com...
Use the softkeys on the right and left (↑ ↓) to position the bar cursor on Ergometry and confirm the selection with Select. Figure 6 – 9: Main menu User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com...
It is recommended that the patient continue to pedal in the recovery phase. The End key in the middle will terminate the test. Figure 6 – 13: Recovery phase User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com...
End key located in the middle. The load will immediately drop to 0 watt. There is no recovery phase in the manual mode. Figure 6 – 16: Display during the exercise test User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Use the softkeys on the right and left (↑ ↓) to position the bar cursor on the protocol to change (No. 6 to 15) and Figure 6 – 20: Selecting the exercise test protocol to edit confirm the selection with Select. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Depending on your choice, each key press will change the load by +/– 1, 5, 10 or 25 watts. Figure 6 – 24: Selecting the increment for manual load changes User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
6.9.8 Pulse Display The pulse readout on the display can be turned off. Figure 6 – 28: Enable/disable the pulse readout User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales office: Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470...
Please contact your authorized manufacturer ergoline GmbH for information concerning the disposal of your equipment. There is no waste approval. Proper disposal is documented by ergoline GmbH. Consult Operator's Manual! User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com...
Protection class / degree of protection / B (ergometer) MDD classifi cation class IIa to 93/42 EEC RF emission class B to DIN EN 55011 / 5.0 DIN EN 60601-1-2 User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales offi ce: Via F.lli Bronzetti, 8...
8.3 Family of characteristics of the braking torque control range Figure 8 – 1: black: speed-independent range to DIN VDE 0750-0238 black + gray: speed-independent range of the xr50+ ergometer User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Harmonic emissions to EN 61000‑3‑2 Class A to the public low‑voltage power supply network that supplies buildings used for domestic purposes. Voltage fluctuations / flicker emissions Complies to EN 61000‑3‑3 User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales office: Via F.lli Bronzetti, 8...
Page 31
The xr50+ ergometer has no compo- nents susceptible to magnetic fi elds. Note: UT is the a.c. mains voltage prior to application of the test level. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com...
Page 32
Over the frequency range from 150 kHz to 80 MHz, field strengths should be less than 3 V/m. User Manual: xr50+_eng Cardioline S.p.A.
Page 33
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it...
Page 34
User Manual: xr50+_eng Cardioline S.p.A. Via Linz,151 38121 Trento (TN) info@cardioline.it www.cardioline.com Sales office: Via F.lli Bronzetti, 8 20129 Milano (MI) Italy tel.+39 02 94750470 fax +39 02 94750471...