Instructions for use These safety instructions are intended to be used alongside safety any local safety guidelines, rules and regulations. If in doubt please consult the Health and Safety Officer of your organisation. legal considerations Only skilled electricians or engineers authorised by MMM It is important that you read this manual can perform repairs or maintenance on this incubator.
Instructions for use The sensor is not mounted within the chamber; instead a construction small pump passes the chamber atmosphere through the The inner chamber is made from 316 grade stainless steel. sensor. It is deep drawn from one piece with no welds and is The sensor does not have to be removed during high electro-polished, meaning that the inner surfaces are temperature sterilisation.
Instructions for use delivery storage transportation If the incubator has been in a cold location for a period then it needs to acclimatise to room and storage temperature. Move the incubator to its final location and wait for 4 hours before start up.
Instructions for use When it is level, secure the feet by tightening the lock nuts gas connection with a 10 mm spanner. Risk of injury. Compressed gas. shelves and usable space Risk of damage to unit. Each incubator is supplied with 4 perforated stainless steel Observe relevant regulations for shelves.
Instructions for use Plug the supplied power lead into the IEC connector on the main display rear of the incubator. The main screen shows chamber set points (SP) and Door open alarm timer 30 seconds Plug the other end of the IEC lead into the electrical supply. Door open alarm timer 30 seconds process values (PV).
Instructions for use Use the keypad to Use the keypad to change the low alarm change the low alarm threshold. threshold. This must be negative This must be negative so remember to first so remember to first press press Press to confirm Press to confirm...
Instructions for use As an additional warning the colours of the process values The CO graph is viewed in the same manner. change during alarm condition. utility menu The temperature process On the main screen press the utility menu symbol: value has changed colour from black to red.
Instructions for use door alarm menu Press the door icon: The door has a magnetic switch that senses when the door is opened. user security The door icon turns red as soon as the door is opened. After a set time the alarm will sound. This can be muted in the This incubator allows for a user name and password alarm status menu.
Instructions for use 6.6.3 start logging logging out On the utility menu screen, press and hold the log out icon: If the card is already inserted when the incubator is powered up then logging will start automatically. You are logged out and the padlock icon is now locked: If the SD symbol is grey then the incubator is No changes to the incubator can be made without logging...
Instructions for use 7.3. Removing the SD card Always perform the following steps before changing filename removing the sd card these steps will results in corrupted data. You can easily change the filename to start a new log. Always perform the following steps before removing the SD Card.
Instructions for use This shows that the door locks have activated and that high temperature Operator 6 is logged on. sterilisation Risk of damage to unit. The Comfort 190 is not intended to be used as an autoclave and must not be used for sterilising Press any other equipment.
Instructions for use This sterilise cycle graph shows a typical start to finish aborting the sterilise cycle duration of 8 hours: Should you need to stop the sterilise cycle for any reason then press the abort icon: cooling phase After the 2 hour dwell is completed the incubator goes into the cooling phase of the sterilisation cycle.
Instructions for use The incubator will be supplied with a pre-wired 3 metre SD card critical error cable socket. SD has not been safely removed. Press and hold Switch power off, insert card and switch power on again. If problem persists contact MMM Service SD card removal error The temperature signal is scaled as: 4 mA = 0 °C;...
Instructions for use 12.3 oxidation at high temperature In line CO reducing valve with gauge (0.1 to 2 bar) comfort s only PNEU Automatic change over unit for two CO When operated at 200 °C the action of atmospheric oxygen cylinders may cause discolouration of the stainless steel surfaces.
Instructions for use disposal Volume Litres (cu. ft.) 190 (6.7) Number of shelves 4 shelves. Maximum of 8 SHIPPING DIMENSIONS The incubator must not be disposed of at public waste collection points. It must only be disposed of by licensed Width mm (inches) 925 (36.4)
Page 20
Instructions for use CO2CELL Comfort 190_np_en 1301_mmm_V1.02...
Page 21
Date Europ pean Directiv ve 2006/95/EC C, 2004/108/EC 2-06-09 Manu ufacturer Medcenter E inrichtungen GmbH, Sem melweisstras sse 6, D-8215 52 Planegg / M München declare e on our own responsibility that product All culture s safe touch an d precision CO O2 incubator incub...
Page 24
Manufactured in the EU MMM Medcenter Einrichtungen GmbH Semmelweisstrasse 6 D-82152 Planegg Germany T. +49 89 89 92 26 20 F. +49 89 89 92 26 30 e-mail: medcenter@mmmgroup.com...