Fax: +44 1480 396296 Web site: www.picotech.com Overview The USB DrDAQ features 14 channels, comprising of sensors, inputs and outputs. These are covered in more detail in Specifications . These connections and sensors cover measurement of sound, light, voltage, resistance, pH levels, and redox potential.
AC or DC voltage. For computers that do not have an earth connection (for example, laptops), it must be assumed that USB DrDAQ is not protected by an earth (in the same way a battery multimeter is not protected by an earth).
Page 8
Introduction Fitness for purpose No two applications are the same: Pico Technology cannot guarantee that its equipment or software is suitable for a given application. It is your responsibility, therefore, to ensure that the product is suitable for your application.
"Install Software" link and follow the on-screen instructions. 2. Plug in the USB DrDAQ device. Windows will recognise it and prepare your computer to work with it. Wait until Windows tells you that the device is ready to use.
3.3.5 Adding spectrum views So far we have looked at the oscilloscope functions. PicoScope allows USB DrDAQ to be used as a spectrum analyser as well. The buttons on the top left-hand side control the oscilloscope, and spectrum views.
When Enable LED C ontrol is active, the LED will not flash when sampling. The USB DrDAQ units include an RGB LED. The colour of this LED can be changed, to do this click the LED icon above the oscilloscope window. The following drop down will...
The USB DrDAQ device includes several digital connections. To activate the digital outs, ensure your accessory is connected both to the G N D of the USB DrDAQ and the digital connection you wish to utilise. To operate the connections click the digital out icon above the oscilloscope window.
3.6.1 Overview As well as the built in sensors, USB DrDAQ has sockets for optional external sensors. When a sensor is plugged in to the external sensor sockets, the software detects it and automatically scales readings. For example, if a temperature sensor is plugged in, readings are displayed in °C, or if a humidity sensor is plugged in, readings are...
Page 21
USB DrDAQ User's Guide Using USB DrDAQ with pH Electrodes The pH input on USB DrDAQ is a very high impedance input that is suitable for use with any standard pH electrode. For most applications no calibration is required: just plug in an electrode and measure.
Page 22
1. Go to Settings and click on Input Channels on the pull-down menu. 2. The converter input type will then be displayed. Check settings and click OK. You should now see the USB DrDAQ measurements box. 3. Select pH and click Edit.
You may use the EL029 to connect a simple, single-pole switch (such as a micro- switch) to USB DrDAQ. You will need to obtain a switch and some insulated connecting wire. Remember to keep magnets or magnetic fields away from the EL029 when you use it with an external switch.
Do not allow the DD101 sensor to become wet. The DD101 is a non- condensing sensor and liquid entering the case (including condensation) may damage The DD101 may be connected to either Ext 1, Ext 2 or Ext 3 on USB DrDAQ. Tips for use of DD101 Humidity Sensors The sensor responds to humidity changes more slowly in still air.
Page 25
Calibration of the sensor involves: Preparation of the calibration equipment. Measurement of the standard saturated solutions. Creation of calibration data for the USB DrDAQ software. A check that the calibrated sensor is accurate. Note: You must measure at least two different standard solutions to provide two or more fixed points for calibration.
Page 27
Connect the DD101 and DD100 sensors to USB DrDAQ. Take your measurements Start the PicoLog Recorder software for USB DrDAQ. Make sure USB DrDAQ is receiving readings from the sensors on Ext 1 or Ext 2. Seal the test container and close the insulated box (if you are using a fan, start the fan).
Page 28
When the two are in agreement use the values given by the calibrated reference as the correct figures for your data table.] Once you have your calibration data you must enter it into the Pico software to calibrate the sensor:...
You can check whether a sensor is working properly or not by running through the calibration procedure. If you find that the sensor values differ from the standard values by more than 10% you may have a damaged sensor. In this case contact Pico for assistance.
Page 30
(calibration is necessary to use the sensor with ±3% accuracy). If you have not yet set up USB DrDAQ with PicoLog, do so using the PicoLog help file before continuing below: 1. Connect the Oxygen Sensor to the socket labelled Ext1 on the USB DrDAQ unit 2.
Page 31
USB DrDAQ User's Guide PicoLog's monitor view should now display 20.9% (or very close) for the oxygen content in air. If so, you've successfully calibrated your USB DrDAQ oxygen sensor. Storing the Oxygen Sensor Ensure that the storage temperature and humidity ranges in the specification are not exceeded.
This is a hardware limitation and any signals outside this range will not be read by the unit. Any voltages greater than ± 30 volts fed directly into the device are likely to damage the unit, and can cause errors with all other USB DrDAQ readings. usbdrdaq.en...
The Signal Input channel has 100 kΩ pull-up resistors to 2.5 V, so that the input can be either a resistance or a voltage. Here is a block diagram of a typical USB DrDAQ sensor: 1 = Signal input 2 = Ground...
There are many factors to take into account when designing scaling circuitry: Sensitivity The USB DrDAQ has 12-bit resolution over the 0 to 2.5 V input range. This means the sensitivity is: 2500 mV / 2^12 = 0.611 mV To make the most of the resolution, the signal output from the sensor should use as much of the input range of the USB DrDAQ as possible.
Page 35
It is quite usual to see offsets in output signals from sensors. Sensor output The USB DrDAQ requires an input signal in DC volts or resistance. The sensor in question could give an output in other units such as resistance, current or AC volts. Drift It is possible that the sensor output drifts over time.
Page 37
Example 3: Measuring 10 volts Notes: This is a simple circuit that uses a potential divider to reduce the voltage going into the USB DrDAQ by four. R1 = 3k & R2 = 1k The resistors are calculated using Ohm's Law, V=IR. Auto-Detect Auto-detect resistor is placed between Ground (Pin 2) and Signal Detect (Pin 3).
USB DrDAQ using the M ask and A larm functions of PicoScope. With USB DrDAQ connected and PicoScope open: Connect the LED to the G N D connection on the USB DrDAQ and digital connection 2 (refer to Connections). Note: Any digital connection can be used but we are using 2 in this example).
The meanings of the terms in the file are as follows: .DDS [Scale1] A unique number, from 1 to 99, to identify this entry. (Pico-created numbers are from 100 upwards.) Resistor=1 The ID resistor value in kiloohms. In this example "...
Page 40
1100 Light 1000 LongName=Temperature Used in PicoLog ShortName=TempC This field is not used by USB DrDAQ running PicoScope or PicoLog. Units=C Displayed on graphs MinValue=-40 MaxValue=120 Note: For PicoScope these values will determine the maximum and minimum values displayed in Oscilloscope View. For PicoLog these values determine what Maximum range is displayed in the Graph View (set in the Graph Options dialog).
Single-Point Calibration. This only occurs in one circumstance: O n e p air o f valu es is k n o w n - for example on the USB DrDAQ oxygen sensor. The sensor is known to output 0mV at 0% oxygen content. Therefore we only need to find one other pair of values.
The latest versions of PicoScope can check for updates automatically and will advise you if an update is available. The latest version of all our software can be downloaded free of charge from the Pico Technology web site at http://www.picotech.
Need help?
Do you have a question about the USB DrDAQ and is the answer not in the manual?
Questions and answers